Evaluation and additional recommendations for preparing a whole blood control material

Autores
Fink, Nilda Esther; Fernández Alberti, María Alejandra; Crispini, Isabel; Cabutti, Norberto V.; Mazziotta, Daniel
Año de publicación
1998
Idioma
inglés
Tipo de recurso
artículo
Estado
versión publicada
Descripción
Objective: The assessment of an easy to prepare and low cost control material for Hematology, available for manual and automated methods. Material and Method: Aliquots of stabilized whole blood were prepared by partial fixation with aldehydes; the stability at different temperatures (4. 20 and 37 °C) during periods of up to 8-9 weeks and aliquot variability with both methods were controlled. Results: Aliquot variability with automated methods at day 1, expressed as CV% (coefficient of variation) was: white blood cells (WBC) 2.7, red blood cells (RBC) 0.7, hemoglobin (Hb) 0.6, hematocrit (Hct) 0.7, mean cell volume (MCV) 0.3, mean cell hemoglobin (MCH) 0.6, mean cell hemoglobin concentration (MCHC) 0.7, and platelets (PLT) 4.6. The CV (coefficient of variation) percentages obtained with manual methods in one of the batches were: WBC 23, Hct 2.8, Hb 4.5, MCHC 5.9, PLT 41. Samples stored at 4ºC and 20ºC showed good stability, only a very low initial hemolysis being observed, whereas those stored at 37ºC deteriobed a rapidly (metahemoglobin formation, aggregation of WBC and platelets, as well as alteration of erythrocyte indexes). Conclusions: It was confirmed that, as long as there is no exposure to high temperatures during distribution, this material is stable, allowing assessment, both esternal and internal, for control purposes, with acceptable reproductivity, both for manual and auttomatic methods.
Objetivo: Avaliar material de controle para hematologia, de fácil preparação e baixo custo, que poderá ser usado por métodos manuais e automatizados. Material e Método: Alíquotas de sangue estabilizado foram preparadas por fixação parcial com aldeídos. Foram estudadas a estabilidade a diferentes temperaturas (4, 20, 37ºC) durante períodos de até 8-9 semanas e a variabilidade das alíquotas para ambos os métodos. Resultados: A variabilidade entre alíquotas, expressada em CV% (coeficiente de variação) com métodos automatizados no primero dia, foi: glóbulos blancos (WBC) 2,7, glóbulos vermelhos (RBC) 0,7, hemoglobina (Hb) 0,6, hematócrito (Hct) 0,7, volume corpuscular médio (MCV) 0,3, Hb corpuscular médio (MCH) 0,3, concentração de Hb corpuscular médio (MCHC) 0,7 e plaquetas (Plt) 4,6. O CV% obtido com métodos manuais para um dos grupos foi: WBC 23, Hct 2,8, Hb 4,5, MCHC 5,9, Plt 41. As amostras conservadas a 4 e 20ºC foram estáveis, observando-se leve hemólise inicial, enquanto que as amostras conservadas a 37ºC mostraram uma rápida decomposição (formação de metaemoglobina, agregação de glóbulos brancos e plaquetas e alteração de índices). Conclusões: Confirmou-se que, se não houver exposição a altas temperaturas durante a distribuição, este material é estável, permitindo avaliação, tanto externa como interna para controle e com reprodutividade aceitável, tanto para métodos manuais como automáticos. Sangue. Controle de qualidade. Laboratórios.observed, whereas those stored at 37ºC deteriobed a rapidly (metahemoglobin formation, aggregation of WBC and platelets, as well as alteration of erythrocyte indexes). Conclusions It was confirmed that, as long as there is no exposure to high temperatures during distribution, this material is stable, allowing assessment, both esternal and internal, for control purposes, with acceptable reproductivity, both for manual and auttomatic methods.
Facultad de Ciencias Exactas
Materia
Ciencias Exactas
Equipo de Laboratorio
Sangre
Control de Calidad
Hematología
Nivel de accesibilidad
acceso abierto
Condiciones de uso
http://creativecommons.org/licenses/by-nc/3.0/
Repositorio
SEDICI (UNLP)
Institución
Universidad Nacional de La Plata
OAI Identificador
oai:sedici.unlp.edu.ar:10915/29367

id SEDICI_9f9a8eb3c678266b36ab9c731c9d51a9
oai_identifier_str oai:sedici.unlp.edu.ar:10915/29367
network_acronym_str SEDICI
repository_id_str 1329
network_name_str SEDICI (UNLP)
spelling Evaluation and additional recommendations for preparing a whole blood control materialFink, Nilda EstherFernández Alberti, María AlejandraCrispini, IsabelCabutti, Norberto V.Mazziotta, DanielCiencias ExactasEquipo de LaboratorioSangreControl de CalidadHematologíaObjective: The assessment of an easy to prepare and low cost control material for Hematology, available for manual and automated methods. Material and Method: Aliquots of stabilized whole blood were prepared by partial fixation with aldehydes; the stability at different temperatures (4. 20 and 37 °C) during periods of up to 8-9 weeks and aliquot variability with both methods were controlled. Results: Aliquot variability with automated methods at day 1, expressed as CV% (coefficient of variation) was: white blood cells (WBC) 2.7, red blood cells (RBC) 0.7, hemoglobin (Hb) 0.6, hematocrit (Hct) 0.7, mean cell volume (MCV) 0.3, mean cell hemoglobin (MCH) 0.6, mean cell hemoglobin concentration (MCHC) 0.7, and platelets (PLT) 4.6. The CV (coefficient of variation) percentages obtained with manual methods in one of the batches were: WBC 23, Hct 2.8, Hb 4.5, MCHC 5.9, PLT 41. Samples stored at 4ºC and 20ºC showed good stability, only a very low initial hemolysis being observed, whereas those stored at 37ºC deteriobed a rapidly (metahemoglobin formation, aggregation of WBC and platelets, as well as alteration of erythrocyte indexes). Conclusions: It was confirmed that, as long as there is no exposure to high temperatures during distribution, this material is stable, allowing assessment, both esternal and internal, for control purposes, with acceptable reproductivity, both for manual and auttomatic methods.Objetivo: Avaliar material de controle para hematologia, de fácil preparação e baixo custo, que poderá ser usado por métodos manuais e automatizados. Material e Método: Alíquotas de sangue estabilizado foram preparadas por fixação parcial com aldeídos. Foram estudadas a estabilidade a diferentes temperaturas (4, 20, 37ºC) durante períodos de até 8-9 semanas e a variabilidade das alíquotas para ambos os métodos. Resultados: A variabilidade entre alíquotas, expressada em CV% (coeficiente de variação) com métodos automatizados no primero dia, foi: glóbulos blancos (WBC) 2,7, glóbulos vermelhos (RBC) 0,7, hemoglobina (Hb) 0,6, hematócrito (Hct) 0,7, volume corpuscular médio (MCV) 0,3, Hb corpuscular médio (MCH) 0,3, concentração de Hb corpuscular médio (MCHC) 0,7 e plaquetas (Plt) 4,6. O CV% obtido com métodos manuais para um dos grupos foi: WBC 23, Hct 2,8, Hb 4,5, MCHC 5,9, Plt 41. As amostras conservadas a 4 e 20ºC foram estáveis, observando-se leve hemólise inicial, enquanto que as amostras conservadas a 37ºC mostraram uma rápida decomposição (formação de metaemoglobina, agregação de glóbulos brancos e plaquetas e alteração de índices). Conclusões: Confirmou-se que, se não houver exposição a altas temperaturas durante a distribuição, este material é estável, permitindo avaliação, tanto externa como interna para controle e com reprodutividade aceitável, tanto para métodos manuais como automáticos. Sangue. Controle de qualidade. Laboratórios.observed, whereas those stored at 37ºC deteriobed a rapidly (metahemoglobin formation, aggregation of WBC and platelets, as well as alteration of erythrocyte indexes). Conclusions It was confirmed that, as long as there is no exposure to high temperatures during distribution, this material is stable, allowing assessment, both esternal and internal, for control purposes, with acceptable reproductivity, both for manual and auttomatic methods.Facultad de Ciencias Exactas1998info:eu-repo/semantics/articleinfo:eu-repo/semantics/publishedVersionArticulohttp://purl.org/coar/resource_type/c_6501info:ar-repo/semantics/articuloapplication/pdf107-111http://sedici.unlp.edu.ar/handle/10915/29367enginfo:eu-repo/semantics/altIdentifier/url/http://www.scielo.br/pdf/rsp/v32n2/p107-111.pdfinfo:eu-repo/semantics/altIdentifier/issn/0034-8910info:eu-repo/semantics/openAccesshttp://creativecommons.org/licenses/by-nc/3.0/Creative Commons Attribution-NonCommercial 3.0 Unported (CC BY-NC 3.0)reponame:SEDICI (UNLP)instname:Universidad Nacional de La Platainstacron:UNLP2025-09-03T10:29:59Zoai:sedici.unlp.edu.ar:10915/29367Institucionalhttp://sedici.unlp.edu.ar/Universidad públicaNo correspondehttp://sedici.unlp.edu.ar/oai/snrdalira@sedici.unlp.edu.arArgentinaNo correspondeNo correspondeNo correspondeopendoar:13292025-09-03 10:29:59.668SEDICI (UNLP) - Universidad Nacional de La Platafalse
dc.title.none.fl_str_mv Evaluation and additional recommendations for preparing a whole blood control material
title Evaluation and additional recommendations for preparing a whole blood control material
spellingShingle Evaluation and additional recommendations for preparing a whole blood control material
Fink, Nilda Esther
Ciencias Exactas
Equipo de Laboratorio
Sangre
Control de Calidad
Hematología
title_short Evaluation and additional recommendations for preparing a whole blood control material
title_full Evaluation and additional recommendations for preparing a whole blood control material
title_fullStr Evaluation and additional recommendations for preparing a whole blood control material
title_full_unstemmed Evaluation and additional recommendations for preparing a whole blood control material
title_sort Evaluation and additional recommendations for preparing a whole blood control material
dc.creator.none.fl_str_mv Fink, Nilda Esther
Fernández Alberti, María Alejandra
Crispini, Isabel
Cabutti, Norberto V.
Mazziotta, Daniel
author Fink, Nilda Esther
author_facet Fink, Nilda Esther
Fernández Alberti, María Alejandra
Crispini, Isabel
Cabutti, Norberto V.
Mazziotta, Daniel
author_role author
author2 Fernández Alberti, María Alejandra
Crispini, Isabel
Cabutti, Norberto V.
Mazziotta, Daniel
author2_role author
author
author
author
dc.subject.none.fl_str_mv Ciencias Exactas
Equipo de Laboratorio
Sangre
Control de Calidad
Hematología
topic Ciencias Exactas
Equipo de Laboratorio
Sangre
Control de Calidad
Hematología
dc.description.none.fl_txt_mv Objective: The assessment of an easy to prepare and low cost control material for Hematology, available for manual and automated methods. Material and Method: Aliquots of stabilized whole blood were prepared by partial fixation with aldehydes; the stability at different temperatures (4. 20 and 37 °C) during periods of up to 8-9 weeks and aliquot variability with both methods were controlled. Results: Aliquot variability with automated methods at day 1, expressed as CV% (coefficient of variation) was: white blood cells (WBC) 2.7, red blood cells (RBC) 0.7, hemoglobin (Hb) 0.6, hematocrit (Hct) 0.7, mean cell volume (MCV) 0.3, mean cell hemoglobin (MCH) 0.6, mean cell hemoglobin concentration (MCHC) 0.7, and platelets (PLT) 4.6. The CV (coefficient of variation) percentages obtained with manual methods in one of the batches were: WBC 23, Hct 2.8, Hb 4.5, MCHC 5.9, PLT 41. Samples stored at 4ºC and 20ºC showed good stability, only a very low initial hemolysis being observed, whereas those stored at 37ºC deteriobed a rapidly (metahemoglobin formation, aggregation of WBC and platelets, as well as alteration of erythrocyte indexes). Conclusions: It was confirmed that, as long as there is no exposure to high temperatures during distribution, this material is stable, allowing assessment, both esternal and internal, for control purposes, with acceptable reproductivity, both for manual and auttomatic methods.
Objetivo: Avaliar material de controle para hematologia, de fácil preparação e baixo custo, que poderá ser usado por métodos manuais e automatizados. Material e Método: Alíquotas de sangue estabilizado foram preparadas por fixação parcial com aldeídos. Foram estudadas a estabilidade a diferentes temperaturas (4, 20, 37ºC) durante períodos de até 8-9 semanas e a variabilidade das alíquotas para ambos os métodos. Resultados: A variabilidade entre alíquotas, expressada em CV% (coeficiente de variação) com métodos automatizados no primero dia, foi: glóbulos blancos (WBC) 2,7, glóbulos vermelhos (RBC) 0,7, hemoglobina (Hb) 0,6, hematócrito (Hct) 0,7, volume corpuscular médio (MCV) 0,3, Hb corpuscular médio (MCH) 0,3, concentração de Hb corpuscular médio (MCHC) 0,7 e plaquetas (Plt) 4,6. O CV% obtido com métodos manuais para um dos grupos foi: WBC 23, Hct 2,8, Hb 4,5, MCHC 5,9, Plt 41. As amostras conservadas a 4 e 20ºC foram estáveis, observando-se leve hemólise inicial, enquanto que as amostras conservadas a 37ºC mostraram uma rápida decomposição (formação de metaemoglobina, agregação de glóbulos brancos e plaquetas e alteração de índices). Conclusões: Confirmou-se que, se não houver exposição a altas temperaturas durante a distribuição, este material é estável, permitindo avaliação, tanto externa como interna para controle e com reprodutividade aceitável, tanto para métodos manuais como automáticos. Sangue. Controle de qualidade. Laboratórios.observed, whereas those stored at 37ºC deteriobed a rapidly (metahemoglobin formation, aggregation of WBC and platelets, as well as alteration of erythrocyte indexes). Conclusions It was confirmed that, as long as there is no exposure to high temperatures during distribution, this material is stable, allowing assessment, both esternal and internal, for control purposes, with acceptable reproductivity, both for manual and auttomatic methods.
Facultad de Ciencias Exactas
description Objective: The assessment of an easy to prepare and low cost control material for Hematology, available for manual and automated methods. Material and Method: Aliquots of stabilized whole blood were prepared by partial fixation with aldehydes; the stability at different temperatures (4. 20 and 37 °C) during periods of up to 8-9 weeks and aliquot variability with both methods were controlled. Results: Aliquot variability with automated methods at day 1, expressed as CV% (coefficient of variation) was: white blood cells (WBC) 2.7, red blood cells (RBC) 0.7, hemoglobin (Hb) 0.6, hematocrit (Hct) 0.7, mean cell volume (MCV) 0.3, mean cell hemoglobin (MCH) 0.6, mean cell hemoglobin concentration (MCHC) 0.7, and platelets (PLT) 4.6. The CV (coefficient of variation) percentages obtained with manual methods in one of the batches were: WBC 23, Hct 2.8, Hb 4.5, MCHC 5.9, PLT 41. Samples stored at 4ºC and 20ºC showed good stability, only a very low initial hemolysis being observed, whereas those stored at 37ºC deteriobed a rapidly (metahemoglobin formation, aggregation of WBC and platelets, as well as alteration of erythrocyte indexes). Conclusions: It was confirmed that, as long as there is no exposure to high temperatures during distribution, this material is stable, allowing assessment, both esternal and internal, for control purposes, with acceptable reproductivity, both for manual and auttomatic methods.
publishDate 1998
dc.date.none.fl_str_mv 1998
dc.type.none.fl_str_mv info:eu-repo/semantics/article
info:eu-repo/semantics/publishedVersion
Articulo
http://purl.org/coar/resource_type/c_6501
info:ar-repo/semantics/articulo
format article
status_str publishedVersion
dc.identifier.none.fl_str_mv http://sedici.unlp.edu.ar/handle/10915/29367
url http://sedici.unlp.edu.ar/handle/10915/29367
dc.language.none.fl_str_mv eng
language eng
dc.relation.none.fl_str_mv info:eu-repo/semantics/altIdentifier/url/http://www.scielo.br/pdf/rsp/v32n2/p107-111.pdf
info:eu-repo/semantics/altIdentifier/issn/0034-8910
dc.rights.none.fl_str_mv info:eu-repo/semantics/openAccess
http://creativecommons.org/licenses/by-nc/3.0/
Creative Commons Attribution-NonCommercial 3.0 Unported (CC BY-NC 3.0)
eu_rights_str_mv openAccess
rights_invalid_str_mv http://creativecommons.org/licenses/by-nc/3.0/
Creative Commons Attribution-NonCommercial 3.0 Unported (CC BY-NC 3.0)
dc.format.none.fl_str_mv application/pdf
107-111
dc.source.none.fl_str_mv reponame:SEDICI (UNLP)
instname:Universidad Nacional de La Plata
instacron:UNLP
reponame_str SEDICI (UNLP)
collection SEDICI (UNLP)
instname_str Universidad Nacional de La Plata
instacron_str UNLP
institution UNLP
repository.name.fl_str_mv SEDICI (UNLP) - Universidad Nacional de La Plata
repository.mail.fl_str_mv alira@sedici.unlp.edu.ar
_version_ 1842260142320517120