Automated multiresidue analysis of pesticides in olive oil by on-line reversed-phase liquid chromatography-gas chromatography using the through oven transfer adsorption-desorption...

Autores
Sanchez, R.; Vazquez, A.; Andini, Juan Carlos; Villén, J.
Año de publicación
2004
Idioma
inglés
Tipo de recurso
artículo
Estado
versión publicada
Descripción
A multiresidue, automated and rapid method for the determination of pesticide residues in olive oil is presented. The method employs the through oven transfer adsorption-desorption interface for the on-line coupling of reversed-phase liquid chromatography and gas chromatography. In this fully automated system, olive oil is directly injected with no sample pre-treatment step other than filtration. Methanol-water is used as eluent in the liquid chromatography pre-separation step. The selected liquid chromatography fraction containing the pesticides is automatically transferred to the gas chromatography. The liquid chromatography column flow during elution is different from the flow during the transfer. Using a flame ionisation detector, pesticide detection limits varied from 0.1 to 0.3 mg/l.
Fil: Sanchez, R.. Universidad de Castilla-La Mancha; España
Fil: Vazquez, A.. Universidad de Castilla-La Mancha; España
Fil: Andini, Juan Carlos. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - Santa Fe. Instituto de Desarrollo Tecnológico para la Industria Química. Universidad Nacional del Litoral. Instituto de Desarrollo Tecnológico para la Industria Química; Argentina
Fil: Villén, J.. Universidad de Castilla-La Mancha; España
Materia
Liquid Chromatographygas Chromatography
Pesticides
Olive Oil
Interfaces
Nivel de accesibilidad
acceso abierto
Condiciones de uso
https://creativecommons.org/licenses/by-nc-sa/2.5/ar/
Repositorio
CONICET Digital (CONICET)
Institución
Consejo Nacional de Investigaciones Científicas y Técnicas
OAI Identificador
oai:ri.conicet.gov.ar:11336/28540

id CONICETDig_fff2c3dd9c42a1c07a617a6cd5af7ec0
oai_identifier_str oai:ri.conicet.gov.ar:11336/28540
network_acronym_str CONICETDig
repository_id_str 3498
network_name_str CONICET Digital (CONICET)
spelling Automated multiresidue analysis of pesticides in olive oil by on-line reversed-phase liquid chromatography-gas chromatography using the through oven transfer adsorption-desorption interfaceSanchez, R.Vazquez, A.Andini, Juan CarlosVillén, J.Liquid Chromatographygas ChromatographyPesticidesOlive OilInterfaceshttps://purl.org/becyt/ford/1.4https://purl.org/becyt/ford/1A multiresidue, automated and rapid method for the determination of pesticide residues in olive oil is presented. The method employs the through oven transfer adsorption-desorption interface for the on-line coupling of reversed-phase liquid chromatography and gas chromatography. In this fully automated system, olive oil is directly injected with no sample pre-treatment step other than filtration. Methanol-water is used as eluent in the liquid chromatography pre-separation step. The selected liquid chromatography fraction containing the pesticides is automatically transferred to the gas chromatography. The liquid chromatography column flow during elution is different from the flow during the transfer. Using a flame ionisation detector, pesticide detection limits varied from 0.1 to 0.3 mg/l.Fil: Sanchez, R.. Universidad de Castilla-La Mancha; EspañaFil: Vazquez, A.. Universidad de Castilla-La Mancha; EspañaFil: Andini, Juan Carlos. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - Santa Fe. Instituto de Desarrollo Tecnológico para la Industria Química. Universidad Nacional del Litoral. Instituto de Desarrollo Tecnológico para la Industria Química; ArgentinaFil: Villén, J.. Universidad de Castilla-La Mancha; EspañaElsevier Science2004-01info:eu-repo/semantics/articleinfo:eu-repo/semantics/publishedVersionhttp://purl.org/coar/resource_type/c_6501info:ar-repo/semantics/articuloapplication/pdfapplication/pdfhttp://hdl.handle.net/11336/28540Sanchez, R.; Vazquez, A.; Andini, Juan Carlos; Villén, J.; Automated multiresidue analysis of pesticides in olive oil by on-line reversed-phase liquid chromatography-gas chromatography using the through oven transfer adsorption-desorption interface; Elsevier Science; Journal Of Chromatography - A; 1029; 1-2004; 167-1720021-9673CONICET DigitalCONICETenginfo:eu-repo/semantics/altIdentifier/doi/10.1016/j.chroma.2003.12.018info:eu-repo/semantics/openAccesshttps://creativecommons.org/licenses/by-nc-sa/2.5/ar/reponame:CONICET Digital (CONICET)instname:Consejo Nacional de Investigaciones Científicas y Técnicas2025-09-29T10:16:49Zoai:ri.conicet.gov.ar:11336/28540instacron:CONICETInstitucionalhttp://ri.conicet.gov.ar/Organismo científico-tecnológicoNo correspondehttp://ri.conicet.gov.ar/oai/requestdasensio@conicet.gov.ar; lcarlino@conicet.gov.arArgentinaNo correspondeNo correspondeNo correspondeopendoar:34982025-09-29 10:16:50.123CONICET Digital (CONICET) - Consejo Nacional de Investigaciones Científicas y Técnicasfalse
dc.title.none.fl_str_mv Automated multiresidue analysis of pesticides in olive oil by on-line reversed-phase liquid chromatography-gas chromatography using the through oven transfer adsorption-desorption interface
title Automated multiresidue analysis of pesticides in olive oil by on-line reversed-phase liquid chromatography-gas chromatography using the through oven transfer adsorption-desorption interface
spellingShingle Automated multiresidue analysis of pesticides in olive oil by on-line reversed-phase liquid chromatography-gas chromatography using the through oven transfer adsorption-desorption interface
Sanchez, R.
Liquid Chromatographygas Chromatography
Pesticides
Olive Oil
Interfaces
title_short Automated multiresidue analysis of pesticides in olive oil by on-line reversed-phase liquid chromatography-gas chromatography using the through oven transfer adsorption-desorption interface
title_full Automated multiresidue analysis of pesticides in olive oil by on-line reversed-phase liquid chromatography-gas chromatography using the through oven transfer adsorption-desorption interface
title_fullStr Automated multiresidue analysis of pesticides in olive oil by on-line reversed-phase liquid chromatography-gas chromatography using the through oven transfer adsorption-desorption interface
title_full_unstemmed Automated multiresidue analysis of pesticides in olive oil by on-line reversed-phase liquid chromatography-gas chromatography using the through oven transfer adsorption-desorption interface
title_sort Automated multiresidue analysis of pesticides in olive oil by on-line reversed-phase liquid chromatography-gas chromatography using the through oven transfer adsorption-desorption interface
dc.creator.none.fl_str_mv Sanchez, R.
Vazquez, A.
Andini, Juan Carlos
Villén, J.
author Sanchez, R.
author_facet Sanchez, R.
Vazquez, A.
Andini, Juan Carlos
Villén, J.
author_role author
author2 Vazquez, A.
Andini, Juan Carlos
Villén, J.
author2_role author
author
author
dc.subject.none.fl_str_mv Liquid Chromatographygas Chromatography
Pesticides
Olive Oil
Interfaces
topic Liquid Chromatographygas Chromatography
Pesticides
Olive Oil
Interfaces
purl_subject.fl_str_mv https://purl.org/becyt/ford/1.4
https://purl.org/becyt/ford/1
dc.description.none.fl_txt_mv A multiresidue, automated and rapid method for the determination of pesticide residues in olive oil is presented. The method employs the through oven transfer adsorption-desorption interface for the on-line coupling of reversed-phase liquid chromatography and gas chromatography. In this fully automated system, olive oil is directly injected with no sample pre-treatment step other than filtration. Methanol-water is used as eluent in the liquid chromatography pre-separation step. The selected liquid chromatography fraction containing the pesticides is automatically transferred to the gas chromatography. The liquid chromatography column flow during elution is different from the flow during the transfer. Using a flame ionisation detector, pesticide detection limits varied from 0.1 to 0.3 mg/l.
Fil: Sanchez, R.. Universidad de Castilla-La Mancha; España
Fil: Vazquez, A.. Universidad de Castilla-La Mancha; España
Fil: Andini, Juan Carlos. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - Santa Fe. Instituto de Desarrollo Tecnológico para la Industria Química. Universidad Nacional del Litoral. Instituto de Desarrollo Tecnológico para la Industria Química; Argentina
Fil: Villén, J.. Universidad de Castilla-La Mancha; España
description A multiresidue, automated and rapid method for the determination of pesticide residues in olive oil is presented. The method employs the through oven transfer adsorption-desorption interface for the on-line coupling of reversed-phase liquid chromatography and gas chromatography. In this fully automated system, olive oil is directly injected with no sample pre-treatment step other than filtration. Methanol-water is used as eluent in the liquid chromatography pre-separation step. The selected liquid chromatography fraction containing the pesticides is automatically transferred to the gas chromatography. The liquid chromatography column flow during elution is different from the flow during the transfer. Using a flame ionisation detector, pesticide detection limits varied from 0.1 to 0.3 mg/l.
publishDate 2004
dc.date.none.fl_str_mv 2004-01
dc.type.none.fl_str_mv info:eu-repo/semantics/article
info:eu-repo/semantics/publishedVersion
http://purl.org/coar/resource_type/c_6501
info:ar-repo/semantics/articulo
format article
status_str publishedVersion
dc.identifier.none.fl_str_mv http://hdl.handle.net/11336/28540
Sanchez, R.; Vazquez, A.; Andini, Juan Carlos; Villén, J.; Automated multiresidue analysis of pesticides in olive oil by on-line reversed-phase liquid chromatography-gas chromatography using the through oven transfer adsorption-desorption interface; Elsevier Science; Journal Of Chromatography - A; 1029; 1-2004; 167-172
0021-9673
CONICET Digital
CONICET
url http://hdl.handle.net/11336/28540
identifier_str_mv Sanchez, R.; Vazquez, A.; Andini, Juan Carlos; Villén, J.; Automated multiresidue analysis of pesticides in olive oil by on-line reversed-phase liquid chromatography-gas chromatography using the through oven transfer adsorption-desorption interface; Elsevier Science; Journal Of Chromatography - A; 1029; 1-2004; 167-172
0021-9673
CONICET Digital
CONICET
dc.language.none.fl_str_mv eng
language eng
dc.relation.none.fl_str_mv info:eu-repo/semantics/altIdentifier/doi/10.1016/j.chroma.2003.12.018
dc.rights.none.fl_str_mv info:eu-repo/semantics/openAccess
https://creativecommons.org/licenses/by-nc-sa/2.5/ar/
eu_rights_str_mv openAccess
rights_invalid_str_mv https://creativecommons.org/licenses/by-nc-sa/2.5/ar/
dc.format.none.fl_str_mv application/pdf
application/pdf
dc.publisher.none.fl_str_mv Elsevier Science
publisher.none.fl_str_mv Elsevier Science
dc.source.none.fl_str_mv reponame:CONICET Digital (CONICET)
instname:Consejo Nacional de Investigaciones Científicas y Técnicas
reponame_str CONICET Digital (CONICET)
collection CONICET Digital (CONICET)
instname_str Consejo Nacional de Investigaciones Científicas y Técnicas
repository.name.fl_str_mv CONICET Digital (CONICET) - Consejo Nacional de Investigaciones Científicas y Técnicas
repository.mail.fl_str_mv dasensio@conicet.gov.ar; lcarlino@conicet.gov.ar
_version_ 1844614116405149696