Surface characterization of reprocessed single-use medical catheters by fractal mass dimension
- Autores
- Scandurra, Adriana Gabriela; Arizmendi, Constancio Miguel; Cuadrado, Teresita Raquel; Granados, Dolly Lucía
- Año de publicación
- 2004
- Idioma
- inglés
- Tipo de recurso
- artículo
- Estado
- versión publicada
- Descripción
- Reprocessing of single-use medical devices is an issue of concern and discussion due to infection risk and operation failure. Cleaning procedures and sterilization processes can produce structural changes and relevant topographical alteration of the biomedical device surfaces which may activate mechanisms able to cause unwanted biological response. Atomic Force Microscopy analysis (AFM) to measure Poly(vinyl chloride) (PVC) catheters surface roughness variation along successive sterilization cycles was applied. In this work, the relation between the number of reprocessing cycles and the fractal mass dimension of reprocessed catheter surface microscopic images is studied.
Fil: Scandurra, Adriana Gabriela. Universidad Nacional de Mar del Plata; Argentina
Fil: Arizmendi, Constancio Miguel. Universidad Nacional de Mar del Plata; Argentina
Fil: Cuadrado, Teresita Raquel. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - Mar del Plata. Instituto de Investigaciones en Ciencia y Tecnología de Materiales. Universidad Nacional de Mar del Plata. Facultad de Ingeniería. Instituto de Investigaciones en Ciencia y Tecnología de Materiales; Argentina
Fil: Granados, Dolly Lucía. Universidad Nacional de San Juan; Argentina - Materia
-
CATHETER REPROCESSING
FRACTAL MASS DIMENSION
IMAGE PROCESSING
SURFACE ROUGHNESS - Nivel de accesibilidad
- acceso abierto
- Condiciones de uso
- https://creativecommons.org/licenses/by-nc-sa/2.5/ar/
- Repositorio
- Institución
- Consejo Nacional de Investigaciones Científicas y Técnicas
- OAI Identificador
- oai:ri.conicet.gov.ar:11336/98898
Ver los metadatos del registro completo
id |
CONICETDig_9e0a25d8c29321427b6e62bb977417f6 |
---|---|
oai_identifier_str |
oai:ri.conicet.gov.ar:11336/98898 |
network_acronym_str |
CONICETDig |
repository_id_str |
3498 |
network_name_str |
CONICET Digital (CONICET) |
spelling |
Surface characterization of reprocessed single-use medical catheters by fractal mass dimensionScandurra, Adriana GabrielaArizmendi, Constancio MiguelCuadrado, Teresita RaquelGranados, Dolly LucíaCATHETER REPROCESSINGFRACTAL MASS DIMENSIONIMAGE PROCESSINGSURFACE ROUGHNESShttps://purl.org/becyt/ford/2.9https://purl.org/becyt/ford/2https://purl.org/becyt/ford/2.5https://purl.org/becyt/ford/2Reprocessing of single-use medical devices is an issue of concern and discussion due to infection risk and operation failure. Cleaning procedures and sterilization processes can produce structural changes and relevant topographical alteration of the biomedical device surfaces which may activate mechanisms able to cause unwanted biological response. Atomic Force Microscopy analysis (AFM) to measure Poly(vinyl chloride) (PVC) catheters surface roughness variation along successive sterilization cycles was applied. In this work, the relation between the number of reprocessing cycles and the fractal mass dimension of reprocessed catheter surface microscopic images is studied.Fil: Scandurra, Adriana Gabriela. Universidad Nacional de Mar del Plata; ArgentinaFil: Arizmendi, Constancio Miguel. Universidad Nacional de Mar del Plata; ArgentinaFil: Cuadrado, Teresita Raquel. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - Mar del Plata. Instituto de Investigaciones en Ciencia y Tecnología de Materiales. Universidad Nacional de Mar del Plata. Facultad de Ingeniería. Instituto de Investigaciones en Ciencia y Tecnología de Materiales; ArgentinaFil: Granados, Dolly Lucía. Universidad Nacional de San Juan; ArgentinaWorld Scientific2004-03info:eu-repo/semantics/articleinfo:eu-repo/semantics/publishedVersionhttp://purl.org/coar/resource_type/c_6501info:ar-repo/semantics/articuloapplication/pdfapplication/pdfhttp://hdl.handle.net/11336/98898Scandurra, Adriana Gabriela; Arizmendi, Constancio Miguel; Cuadrado, Teresita Raquel; Granados, Dolly Lucía; Surface characterization of reprocessed single-use medical catheters by fractal mass dimension; World Scientific; Fractals; 12; 1; 3-2004; 33-390218-348XCONICET DigitalCONICETenginfo:eu-repo/semantics/altIdentifier/url/https://www.worldscientific.com/doi/abs/10.1142/S0218348X04002331info:eu-repo/semantics/altIdentifier/doi/10.1142/S0218348X04002331info:eu-repo/semantics/openAccesshttps://creativecommons.org/licenses/by-nc-sa/2.5/ar/reponame:CONICET Digital (CONICET)instname:Consejo Nacional de Investigaciones Científicas y Técnicas2025-09-29T09:59:44Zoai:ri.conicet.gov.ar:11336/98898instacron:CONICETInstitucionalhttp://ri.conicet.gov.ar/Organismo científico-tecnológicoNo correspondehttp://ri.conicet.gov.ar/oai/requestdasensio@conicet.gov.ar; lcarlino@conicet.gov.arArgentinaNo correspondeNo correspondeNo correspondeopendoar:34982025-09-29 09:59:45.089CONICET Digital (CONICET) - Consejo Nacional de Investigaciones Científicas y Técnicasfalse |
dc.title.none.fl_str_mv |
Surface characterization of reprocessed single-use medical catheters by fractal mass dimension |
title |
Surface characterization of reprocessed single-use medical catheters by fractal mass dimension |
spellingShingle |
Surface characterization of reprocessed single-use medical catheters by fractal mass dimension Scandurra, Adriana Gabriela CATHETER REPROCESSING FRACTAL MASS DIMENSION IMAGE PROCESSING SURFACE ROUGHNESS |
title_short |
Surface characterization of reprocessed single-use medical catheters by fractal mass dimension |
title_full |
Surface characterization of reprocessed single-use medical catheters by fractal mass dimension |
title_fullStr |
Surface characterization of reprocessed single-use medical catheters by fractal mass dimension |
title_full_unstemmed |
Surface characterization of reprocessed single-use medical catheters by fractal mass dimension |
title_sort |
Surface characterization of reprocessed single-use medical catheters by fractal mass dimension |
dc.creator.none.fl_str_mv |
Scandurra, Adriana Gabriela Arizmendi, Constancio Miguel Cuadrado, Teresita Raquel Granados, Dolly Lucía |
author |
Scandurra, Adriana Gabriela |
author_facet |
Scandurra, Adriana Gabriela Arizmendi, Constancio Miguel Cuadrado, Teresita Raquel Granados, Dolly Lucía |
author_role |
author |
author2 |
Arizmendi, Constancio Miguel Cuadrado, Teresita Raquel Granados, Dolly Lucía |
author2_role |
author author author |
dc.subject.none.fl_str_mv |
CATHETER REPROCESSING FRACTAL MASS DIMENSION IMAGE PROCESSING SURFACE ROUGHNESS |
topic |
CATHETER REPROCESSING FRACTAL MASS DIMENSION IMAGE PROCESSING SURFACE ROUGHNESS |
purl_subject.fl_str_mv |
https://purl.org/becyt/ford/2.9 https://purl.org/becyt/ford/2 https://purl.org/becyt/ford/2.5 https://purl.org/becyt/ford/2 |
dc.description.none.fl_txt_mv |
Reprocessing of single-use medical devices is an issue of concern and discussion due to infection risk and operation failure. Cleaning procedures and sterilization processes can produce structural changes and relevant topographical alteration of the biomedical device surfaces which may activate mechanisms able to cause unwanted biological response. Atomic Force Microscopy analysis (AFM) to measure Poly(vinyl chloride) (PVC) catheters surface roughness variation along successive sterilization cycles was applied. In this work, the relation between the number of reprocessing cycles and the fractal mass dimension of reprocessed catheter surface microscopic images is studied. Fil: Scandurra, Adriana Gabriela. Universidad Nacional de Mar del Plata; Argentina Fil: Arizmendi, Constancio Miguel. Universidad Nacional de Mar del Plata; Argentina Fil: Cuadrado, Teresita Raquel. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - Mar del Plata. Instituto de Investigaciones en Ciencia y Tecnología de Materiales. Universidad Nacional de Mar del Plata. Facultad de Ingeniería. Instituto de Investigaciones en Ciencia y Tecnología de Materiales; Argentina Fil: Granados, Dolly Lucía. Universidad Nacional de San Juan; Argentina |
description |
Reprocessing of single-use medical devices is an issue of concern and discussion due to infection risk and operation failure. Cleaning procedures and sterilization processes can produce structural changes and relevant topographical alteration of the biomedical device surfaces which may activate mechanisms able to cause unwanted biological response. Atomic Force Microscopy analysis (AFM) to measure Poly(vinyl chloride) (PVC) catheters surface roughness variation along successive sterilization cycles was applied. In this work, the relation between the number of reprocessing cycles and the fractal mass dimension of reprocessed catheter surface microscopic images is studied. |
publishDate |
2004 |
dc.date.none.fl_str_mv |
2004-03 |
dc.type.none.fl_str_mv |
info:eu-repo/semantics/article info:eu-repo/semantics/publishedVersion http://purl.org/coar/resource_type/c_6501 info:ar-repo/semantics/articulo |
format |
article |
status_str |
publishedVersion |
dc.identifier.none.fl_str_mv |
http://hdl.handle.net/11336/98898 Scandurra, Adriana Gabriela; Arizmendi, Constancio Miguel; Cuadrado, Teresita Raquel; Granados, Dolly Lucía; Surface characterization of reprocessed single-use medical catheters by fractal mass dimension; World Scientific; Fractals; 12; 1; 3-2004; 33-39 0218-348X CONICET Digital CONICET |
url |
http://hdl.handle.net/11336/98898 |
identifier_str_mv |
Scandurra, Adriana Gabriela; Arizmendi, Constancio Miguel; Cuadrado, Teresita Raquel; Granados, Dolly Lucía; Surface characterization of reprocessed single-use medical catheters by fractal mass dimension; World Scientific; Fractals; 12; 1; 3-2004; 33-39 0218-348X CONICET Digital CONICET |
dc.language.none.fl_str_mv |
eng |
language |
eng |
dc.relation.none.fl_str_mv |
info:eu-repo/semantics/altIdentifier/url/https://www.worldscientific.com/doi/abs/10.1142/S0218348X04002331 info:eu-repo/semantics/altIdentifier/doi/10.1142/S0218348X04002331 |
dc.rights.none.fl_str_mv |
info:eu-repo/semantics/openAccess https://creativecommons.org/licenses/by-nc-sa/2.5/ar/ |
eu_rights_str_mv |
openAccess |
rights_invalid_str_mv |
https://creativecommons.org/licenses/by-nc-sa/2.5/ar/ |
dc.format.none.fl_str_mv |
application/pdf application/pdf |
dc.publisher.none.fl_str_mv |
World Scientific |
publisher.none.fl_str_mv |
World Scientific |
dc.source.none.fl_str_mv |
reponame:CONICET Digital (CONICET) instname:Consejo Nacional de Investigaciones Científicas y Técnicas |
reponame_str |
CONICET Digital (CONICET) |
collection |
CONICET Digital (CONICET) |
instname_str |
Consejo Nacional de Investigaciones Científicas y Técnicas |
repository.name.fl_str_mv |
CONICET Digital (CONICET) - Consejo Nacional de Investigaciones Científicas y Técnicas |
repository.mail.fl_str_mv |
dasensio@conicet.gov.ar; lcarlino@conicet.gov.ar |
_version_ |
1844613770172694528 |
score |
13.070432 |