Forma de onda óptima de desfibrilación

Autores
Paredes, Mauro Nehuén; Rojas, Yasmin Madelen
Año de publicación
2023
Idioma
español castellano
Tipo de recurso
tesis de grado
Estado
versión publicada
Colaborador/a o director/a de tesis
Carranza, Gustavo Ernesto
Nieva, Eduardo Gabriel
Descripción
Proyecto Integrador (I.Biom.)--FCEFN-UNC, 2023
Fil: Paredes, Mauro Nehuén. Universidad Nacional de Córdoba. Facultad de Ciencias Exactas, Físicas y Naturales; Argentina.
Fil: Rojas, Yasmin Madelen. Universidad Nacional de Córdoba. Facultad de Ciencias Exactas, Físicas y Naturales; Argentina.
La desfibrilación eléctrica es un proceso fundamental para una adecuada reanimación cardíaca. Sin embargo, este procedimiento conlleva efectos secundarios graves para los pacientes. Este trabajo tuvo como objetivo investigar y analizar la forma de onda utilizada en los desfibriladores comúnmente empleados en el sistema de salud, evaluando sus características eléctricas y su impacto en el tejido cardíaco. Se investigaron en profundidad las características de la membrana celular, las corrientes y los canales iónicos con el fin de desarrollar un modelo mejorado de membrana celular. Debido a limitaciones técnicas, se optó por utilizar modelos preexistentes según conveniencia. A partir de estos fundamentos, se formuló una propuesta metodológica actualizada para obtener la forma de onda óptima de desfibrilación transtorácica, aquella que restablece la función cardíaca con la mínima energía aplicada al paciente, generando el menor daño colateral posible. Ante la falta de una resolución concreta, se propuso el uso de un dispositivo de carga controlada, el cual inyecta una carga ajustable en función del tiempo durante la desfibrilación. Se sugiere una metodología para la evaluación y validación clínica de dicho dispositivo. Se espera que los resultados demuestren una reducción significativa de la energía suministrada, reduciendo así los posibles efectos secundarios y aumentando la efectividad y tasa de éxito del dispositivo para terminar la fibrilación ventricular. Además, se propone realizar investigaciones adicionales para optimizar la forma de onda utilizada en los desfibriladores mediante estudios clínicos prospectivos que comparen directamente diferentes formas de onda y evalúen su eficacia en distintas situaciones clínicas. Se plantea la posibilidad de desarrollar enfoques personalizados de desfibrilación basados en la respuesta individual del paciente. El objetivo final es contribuir al avance del conocimiento en este campo y mejorar la atención médica en situaciones críticas.
Electrical defibrillation is a fundamental process for proper cardiac resuscitation. However, it carries serious side effects for patients undergoing the procedure. The objective of this work was to investigate and analyze the waveform used in defibrillators commonly employed by the healthcare system, assessing their electrical characteristics and their impact on cardiac tissue. In-depth research was conducted on the specific characteristics of the cell membrane, currents, and ion channels in an attempt to create an improved model of the cell membrane. Due to technical limitations, pre-existing models were chosen for use according to convenience. Building on these foundations, an updated methodological proposal was developed for obtaining the optimal waveform for transthoracic defibrillation, which refers to the waveform that restores cardiac function with the minimum energy applied to the patient while generating the least collateral damage possible. Upon realizing that this approach did not lead to a tangible and conclusive resolution, the suggestion was made to employ a controlled charging source that injects an adjustable charge over time during defibrillation. A methodology is proposed for the real clinical evaluation and validation of said device. It is expected that the results will demonstrate a significant reduction in the energy required to achieve defibrillation, thereby potentially minimizing the possible side effects while also increasing the effectiveness and success rate of the controlled charging device in terminating ventricular fibrillation. Furthermore, additional investigations are proposed to optimize the waveform used in defibrillators through prospective clinical studies directly comparing different waveforms and evaluating their efficacy in various clinical situations. The possibility of developing personalized defibrillation approaches based on the individual patient's response is raised. The ultimate goal is to contribute to the advancement of knowledge in this field and enhance medical care in critical situations.
Fil: Paredes, Mauro Nehuén. Universidad Nacional de Córdoba. Facultad de Ciencias Exactas, Físicas y Naturales; Argentina.
Fil: Rojas, Yasmin Madelen. Universidad Nacional de Córdoba. Facultad de Ciencias Exactas, Físicas y Naturales; Argentina.
Materia
Proyecto Integrador (I.Biom.)
Desfibrilación
Forma De Onda
Característica Tensión - Carga
Función Tensión - Corrient
Fuente de carga controlada
Defibrillation
Waveforms
Tension - Charge Characteristic
Tension - Current Relationship
Controlled Current Sources
Nivel de accesibilidad
acceso abierto
Condiciones de uso
Repositorio
Repositorio Digital Universitario (UNC)
Institución
Universidad Nacional de Córdoba
OAI Identificador
oai:rdu.unc.edu.ar:11086/552544

id RDUUNC_0926c1fd0b62c36e0bd4c62cc7ae626a
oai_identifier_str oai:rdu.unc.edu.ar:11086/552544
network_acronym_str RDUUNC
repository_id_str 2572
network_name_str Repositorio Digital Universitario (UNC)
spelling Forma de onda óptima de desfibrilaciónParedes, Mauro NehuénRojas, Yasmin MadelenProyecto Integrador (I.Biom.)DesfibrilaciónForma De OndaCaracterística Tensión - CargaFunción Tensión - CorrientFuente de carga controladaDefibrillationWaveformsTension - Charge CharacteristicTension - Current RelationshipControlled Current SourcesProyecto Integrador (I.Biom.)--FCEFN-UNC, 2023Fil: Paredes, Mauro Nehuén. Universidad Nacional de Córdoba. Facultad de Ciencias Exactas, Físicas y Naturales; Argentina.Fil: Rojas, Yasmin Madelen. Universidad Nacional de Córdoba. Facultad de Ciencias Exactas, Físicas y Naturales; Argentina.La desfibrilación eléctrica es un proceso fundamental para una adecuada reanimación cardíaca. Sin embargo, este procedimiento conlleva efectos secundarios graves para los pacientes. Este trabajo tuvo como objetivo investigar y analizar la forma de onda utilizada en los desfibriladores comúnmente empleados en el sistema de salud, evaluando sus características eléctricas y su impacto en el tejido cardíaco. Se investigaron en profundidad las características de la membrana celular, las corrientes y los canales iónicos con el fin de desarrollar un modelo mejorado de membrana celular. Debido a limitaciones técnicas, se optó por utilizar modelos preexistentes según conveniencia. A partir de estos fundamentos, se formuló una propuesta metodológica actualizada para obtener la forma de onda óptima de desfibrilación transtorácica, aquella que restablece la función cardíaca con la mínima energía aplicada al paciente, generando el menor daño colateral posible. Ante la falta de una resolución concreta, se propuso el uso de un dispositivo de carga controlada, el cual inyecta una carga ajustable en función del tiempo durante la desfibrilación. Se sugiere una metodología para la evaluación y validación clínica de dicho dispositivo. Se espera que los resultados demuestren una reducción significativa de la energía suministrada, reduciendo así los posibles efectos secundarios y aumentando la efectividad y tasa de éxito del dispositivo para terminar la fibrilación ventricular. Además, se propone realizar investigaciones adicionales para optimizar la forma de onda utilizada en los desfibriladores mediante estudios clínicos prospectivos que comparen directamente diferentes formas de onda y evalúen su eficacia en distintas situaciones clínicas. Se plantea la posibilidad de desarrollar enfoques personalizados de desfibrilación basados en la respuesta individual del paciente. El objetivo final es contribuir al avance del conocimiento en este campo y mejorar la atención médica en situaciones críticas.Electrical defibrillation is a fundamental process for proper cardiac resuscitation. However, it carries serious side effects for patients undergoing the procedure. The objective of this work was to investigate and analyze the waveform used in defibrillators commonly employed by the healthcare system, assessing their electrical characteristics and their impact on cardiac tissue. In-depth research was conducted on the specific characteristics of the cell membrane, currents, and ion channels in an attempt to create an improved model of the cell membrane. Due to technical limitations, pre-existing models were chosen for use according to convenience. Building on these foundations, an updated methodological proposal was developed for obtaining the optimal waveform for transthoracic defibrillation, which refers to the waveform that restores cardiac function with the minimum energy applied to the patient while generating the least collateral damage possible. Upon realizing that this approach did not lead to a tangible and conclusive resolution, the suggestion was made to employ a controlled charging source that injects an adjustable charge over time during defibrillation. A methodology is proposed for the real clinical evaluation and validation of said device. It is expected that the results will demonstrate a significant reduction in the energy required to achieve defibrillation, thereby potentially minimizing the possible side effects while also increasing the effectiveness and success rate of the controlled charging device in terminating ventricular fibrillation. Furthermore, additional investigations are proposed to optimize the waveform used in defibrillators through prospective clinical studies directly comparing different waveforms and evaluating their efficacy in various clinical situations. The possibility of developing personalized defibrillation approaches based on the individual patient's response is raised. The ultimate goal is to contribute to the advancement of knowledge in this field and enhance medical care in critical situations.Fil: Paredes, Mauro Nehuén. Universidad Nacional de Córdoba. Facultad de Ciencias Exactas, Físicas y Naturales; Argentina.Fil: Rojas, Yasmin Madelen. Universidad Nacional de Córdoba. Facultad de Ciencias Exactas, Físicas y Naturales; Argentina.Carranza, Gustavo ErnestoNieva, Eduardo Gabriel2023info:eu-repo/semantics/bachelorThesisinfo:eu-repo/semantics/publishedVersionhttp://purl.org/coar/resource_type/c_7a1finfo:ar-repo/semantics/tesisDeGradoapplication/pdfhttp://hdl.handle.net/11086/552544spainfo:eu-repo/semantics/openAccessreponame:Repositorio Digital Universitario (UNC)instname:Universidad Nacional de Córdobainstacron:UNC2025-09-29T13:42:10Zoai:rdu.unc.edu.ar:11086/552544Institucionalhttps://rdu.unc.edu.ar/Universidad públicaNo correspondehttp://rdu.unc.edu.ar/oai/snrdoca.unc@gmail.comArgentinaNo correspondeNo correspondeNo correspondeopendoar:25722025-09-29 13:42:10.495Repositorio Digital Universitario (UNC) - Universidad Nacional de Córdobafalse
dc.title.none.fl_str_mv Forma de onda óptima de desfibrilación
title Forma de onda óptima de desfibrilación
spellingShingle Forma de onda óptima de desfibrilación
Paredes, Mauro Nehuén
Proyecto Integrador (I.Biom.)
Desfibrilación
Forma De Onda
Característica Tensión - Carga
Función Tensión - Corrient
Fuente de carga controlada
Defibrillation
Waveforms
Tension - Charge Characteristic
Tension - Current Relationship
Controlled Current Sources
title_short Forma de onda óptima de desfibrilación
title_full Forma de onda óptima de desfibrilación
title_fullStr Forma de onda óptima de desfibrilación
title_full_unstemmed Forma de onda óptima de desfibrilación
title_sort Forma de onda óptima de desfibrilación
dc.creator.none.fl_str_mv Paredes, Mauro Nehuén
Rojas, Yasmin Madelen
author Paredes, Mauro Nehuén
author_facet Paredes, Mauro Nehuén
Rojas, Yasmin Madelen
author_role author
author2 Rojas, Yasmin Madelen
author2_role author
dc.contributor.none.fl_str_mv Carranza, Gustavo Ernesto
Nieva, Eduardo Gabriel
dc.subject.none.fl_str_mv Proyecto Integrador (I.Biom.)
Desfibrilación
Forma De Onda
Característica Tensión - Carga
Función Tensión - Corrient
Fuente de carga controlada
Defibrillation
Waveforms
Tension - Charge Characteristic
Tension - Current Relationship
Controlled Current Sources
topic Proyecto Integrador (I.Biom.)
Desfibrilación
Forma De Onda
Característica Tensión - Carga
Función Tensión - Corrient
Fuente de carga controlada
Defibrillation
Waveforms
Tension - Charge Characteristic
Tension - Current Relationship
Controlled Current Sources
dc.description.none.fl_txt_mv Proyecto Integrador (I.Biom.)--FCEFN-UNC, 2023
Fil: Paredes, Mauro Nehuén. Universidad Nacional de Córdoba. Facultad de Ciencias Exactas, Físicas y Naturales; Argentina.
Fil: Rojas, Yasmin Madelen. Universidad Nacional de Córdoba. Facultad de Ciencias Exactas, Físicas y Naturales; Argentina.
La desfibrilación eléctrica es un proceso fundamental para una adecuada reanimación cardíaca. Sin embargo, este procedimiento conlleva efectos secundarios graves para los pacientes. Este trabajo tuvo como objetivo investigar y analizar la forma de onda utilizada en los desfibriladores comúnmente empleados en el sistema de salud, evaluando sus características eléctricas y su impacto en el tejido cardíaco. Se investigaron en profundidad las características de la membrana celular, las corrientes y los canales iónicos con el fin de desarrollar un modelo mejorado de membrana celular. Debido a limitaciones técnicas, se optó por utilizar modelos preexistentes según conveniencia. A partir de estos fundamentos, se formuló una propuesta metodológica actualizada para obtener la forma de onda óptima de desfibrilación transtorácica, aquella que restablece la función cardíaca con la mínima energía aplicada al paciente, generando el menor daño colateral posible. Ante la falta de una resolución concreta, se propuso el uso de un dispositivo de carga controlada, el cual inyecta una carga ajustable en función del tiempo durante la desfibrilación. Se sugiere una metodología para la evaluación y validación clínica de dicho dispositivo. Se espera que los resultados demuestren una reducción significativa de la energía suministrada, reduciendo así los posibles efectos secundarios y aumentando la efectividad y tasa de éxito del dispositivo para terminar la fibrilación ventricular. Además, se propone realizar investigaciones adicionales para optimizar la forma de onda utilizada en los desfibriladores mediante estudios clínicos prospectivos que comparen directamente diferentes formas de onda y evalúen su eficacia en distintas situaciones clínicas. Se plantea la posibilidad de desarrollar enfoques personalizados de desfibrilación basados en la respuesta individual del paciente. El objetivo final es contribuir al avance del conocimiento en este campo y mejorar la atención médica en situaciones críticas.
Electrical defibrillation is a fundamental process for proper cardiac resuscitation. However, it carries serious side effects for patients undergoing the procedure. The objective of this work was to investigate and analyze the waveform used in defibrillators commonly employed by the healthcare system, assessing their electrical characteristics and their impact on cardiac tissue. In-depth research was conducted on the specific characteristics of the cell membrane, currents, and ion channels in an attempt to create an improved model of the cell membrane. Due to technical limitations, pre-existing models were chosen for use according to convenience. Building on these foundations, an updated methodological proposal was developed for obtaining the optimal waveform for transthoracic defibrillation, which refers to the waveform that restores cardiac function with the minimum energy applied to the patient while generating the least collateral damage possible. Upon realizing that this approach did not lead to a tangible and conclusive resolution, the suggestion was made to employ a controlled charging source that injects an adjustable charge over time during defibrillation. A methodology is proposed for the real clinical evaluation and validation of said device. It is expected that the results will demonstrate a significant reduction in the energy required to achieve defibrillation, thereby potentially minimizing the possible side effects while also increasing the effectiveness and success rate of the controlled charging device in terminating ventricular fibrillation. Furthermore, additional investigations are proposed to optimize the waveform used in defibrillators through prospective clinical studies directly comparing different waveforms and evaluating their efficacy in various clinical situations. The possibility of developing personalized defibrillation approaches based on the individual patient's response is raised. The ultimate goal is to contribute to the advancement of knowledge in this field and enhance medical care in critical situations.
Fil: Paredes, Mauro Nehuén. Universidad Nacional de Córdoba. Facultad de Ciencias Exactas, Físicas y Naturales; Argentina.
Fil: Rojas, Yasmin Madelen. Universidad Nacional de Córdoba. Facultad de Ciencias Exactas, Físicas y Naturales; Argentina.
description Proyecto Integrador (I.Biom.)--FCEFN-UNC, 2023
publishDate 2023
dc.date.none.fl_str_mv 2023
dc.type.none.fl_str_mv info:eu-repo/semantics/bachelorThesis
info:eu-repo/semantics/publishedVersion
http://purl.org/coar/resource_type/c_7a1f
info:ar-repo/semantics/tesisDeGrado
format bachelorThesis
status_str publishedVersion
dc.identifier.none.fl_str_mv http://hdl.handle.net/11086/552544
url http://hdl.handle.net/11086/552544
dc.language.none.fl_str_mv spa
language spa
dc.rights.none.fl_str_mv info:eu-repo/semantics/openAccess
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv application/pdf
dc.source.none.fl_str_mv reponame:Repositorio Digital Universitario (UNC)
instname:Universidad Nacional de Córdoba
instacron:UNC
reponame_str Repositorio Digital Universitario (UNC)
collection Repositorio Digital Universitario (UNC)
instname_str Universidad Nacional de Córdoba
instacron_str UNC
institution UNC
repository.name.fl_str_mv Repositorio Digital Universitario (UNC) - Universidad Nacional de Córdoba
repository.mail.fl_str_mv oca.unc@gmail.com
_version_ 1844618920319778816
score 13.070432