Optimization of the preconcentration system of cadmium with 1(2-thiazolylazo)-p-cresol using a knotted reactor and flame atomic absorption spectrometric detection
- Autores
- Cerutti, Estela Soledad; Ferreira, Sergio Luis Costa; Gasquez, José Antonio; Olsina, Roberto Antonio; Martinez, Luis Dante
- Año de publicación
- 2004
- Idioma
- inglés
- Tipo de recurso
- artículo
- Estado
- versión publicada
- Descripción
- The present paper proposes an on-line preconcentration procedure for cadmium determination in drinking water samples. It is based on the precipitation of cadmium(II) ions on a knotted reactor (KR) using 1(2-thiazolylazo)-p-cresol (TAC) as complexing reagent. The optimization step was performed using a full factorial design involving the variables: pH, eluent concentration (nitric acid) and TAC concentration. The results of this experiment demonstrated that these variables at chosen levels are not statistically significant. Under optimized experimental established conditions, analytical parameters for the preconcentration method were: a detection limit of 40.0 ng/l, precision as relative standard deviation (RSD) of 1.2 and 1.0%, for cadmium concentration of 2.5 and 20.0 μg/l, respectively. The preconcentration factor considering the slopes of the analytical curves with and without preconcentration is 23 for a sample volume of 10 ml. This system shows a sampling frequency of 25 h-1. In order to check the accuracy, the standard reference material, NIST SRM 1643d trace elements in water was analyzed. A comparison, using t-test demonstrates that there is not significant difference among the achieved results with proposed method and the certified values. The addition/recovery experiments in the samples analyzed demonstrated the accuracy and applicability of the system developed for cadmium determination in water samples.
Fil: Cerutti, Estela Soledad. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - San Luis. Instituto de Química de San Luis. Universidad Nacional de San Luis. Facultad de Química, Bioquímica y Farmacia. Instituto de Química de San Luis; Argentina
Fil: Ferreira, Sergio Luis Costa. Universidade Federal da Bahia; Brasil
Fil: Gasquez, José Antonio. Universidad Nacional de San Luis; Argentina
Fil: Olsina, Roberto Antonio. Consejo Nacional de Investigaciones Científicas y Técnicas; Argentina. Universidad Nacional de San Luis; Argentina
Fil: Martinez, Luis Dante. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - San Luis. Instituto de Química de San Luis. Universidad Nacional de San Luis. Facultad de Química, Bioquímica y Farmacia. Instituto de Química de San Luis; Argentina - Materia
-
CADMIUM DETERMINATION
FAAS
KNOTTED REACTOR
ON-LINE PRECONCENTRATION
WATER SAMPLES, 1(2-THIAZOLYLAZO)-P-CRESOL - Nivel de accesibilidad
- acceso abierto
- Condiciones de uso
- https://creativecommons.org/licenses/by-nc-nd/2.5/ar/
- Repositorio
- Institución
- Consejo Nacional de Investigaciones Científicas y Técnicas
- OAI Identificador
- oai:ri.conicet.gov.ar:11336/97974
Ver los metadatos del registro completo
id |
CONICETDig_a0bd51f48a876a147869231a9e69d1ec |
---|---|
oai_identifier_str |
oai:ri.conicet.gov.ar:11336/97974 |
network_acronym_str |
CONICETDig |
repository_id_str |
3498 |
network_name_str |
CONICET Digital (CONICET) |
spelling |
Optimization of the preconcentration system of cadmium with 1(2-thiazolylazo)-p-cresol using a knotted reactor and flame atomic absorption spectrometric detectionCerutti, Estela SoledadFerreira, Sergio Luis CostaGasquez, José AntonioOlsina, Roberto AntonioMartinez, Luis DanteCADMIUM DETERMINATIONFAASKNOTTED REACTORON-LINE PRECONCENTRATIONWATER SAMPLES, 1(2-THIAZOLYLAZO)-P-CRESOLhttps://purl.org/becyt/ford/1.4https://purl.org/becyt/ford/1The present paper proposes an on-line preconcentration procedure for cadmium determination in drinking water samples. It is based on the precipitation of cadmium(II) ions on a knotted reactor (KR) using 1(2-thiazolylazo)-p-cresol (TAC) as complexing reagent. The optimization step was performed using a full factorial design involving the variables: pH, eluent concentration (nitric acid) and TAC concentration. The results of this experiment demonstrated that these variables at chosen levels are not statistically significant. Under optimized experimental established conditions, analytical parameters for the preconcentration method were: a detection limit of 40.0 ng/l, precision as relative standard deviation (RSD) of 1.2 and 1.0%, for cadmium concentration of 2.5 and 20.0 μg/l, respectively. The preconcentration factor considering the slopes of the analytical curves with and without preconcentration is 23 for a sample volume of 10 ml. This system shows a sampling frequency of 25 h-1. In order to check the accuracy, the standard reference material, NIST SRM 1643d trace elements in water was analyzed. A comparison, using t-test demonstrates that there is not significant difference among the achieved results with proposed method and the certified values. The addition/recovery experiments in the samples analyzed demonstrated the accuracy and applicability of the system developed for cadmium determination in water samples.Fil: Cerutti, Estela Soledad. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - San Luis. Instituto de Química de San Luis. Universidad Nacional de San Luis. Facultad de Química, Bioquímica y Farmacia. Instituto de Química de San Luis; ArgentinaFil: Ferreira, Sergio Luis Costa. Universidade Federal da Bahia; BrasilFil: Gasquez, José Antonio. Universidad Nacional de San Luis; ArgentinaFil: Olsina, Roberto Antonio. Consejo Nacional de Investigaciones Científicas y Técnicas; Argentina. Universidad Nacional de San Luis; ArgentinaFil: Martinez, Luis Dante. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - San Luis. Instituto de Química de San Luis. Universidad Nacional de San Luis. Facultad de Química, Bioquímica y Farmacia. Instituto de Química de San Luis; ArgentinaElsevier Science2004-08info:eu-repo/semantics/articleinfo:eu-repo/semantics/publishedVersionhttp://purl.org/coar/resource_type/c_6501info:ar-repo/semantics/articuloapplication/pdfapplication/pdfapplication/pdfhttp://hdl.handle.net/11336/97974Cerutti, Estela Soledad; Ferreira, Sergio Luis Costa; Gasquez, José Antonio; Olsina, Roberto Antonio; Martinez, Luis Dante; Optimization of the preconcentration system of cadmium with 1(2-thiazolylazo)-p-cresol using a knotted reactor and flame atomic absorption spectrometric detection; Elsevier Science; Journal of Hazardous Materials; 112; 3; 8-2004; 279-2830304-3894CONICET DigitalCONICETenginfo:eu-repo/semantics/altIdentifier/doi/10.1016/j.jhazmat.2004.05.017info:eu-repo/semantics/altIdentifier/url/https://www.sciencedirect.com/science/article/pii/S0304389404002626info:eu-repo/semantics/openAccesshttps://creativecommons.org/licenses/by-nc-nd/2.5/ar/reponame:CONICET Digital (CONICET)instname:Consejo Nacional de Investigaciones Científicas y Técnicas2025-09-03T09:53:51Zoai:ri.conicet.gov.ar:11336/97974instacron:CONICETInstitucionalhttp://ri.conicet.gov.ar/Organismo científico-tecnológicoNo correspondehttp://ri.conicet.gov.ar/oai/requestdasensio@conicet.gov.ar; lcarlino@conicet.gov.arArgentinaNo correspondeNo correspondeNo correspondeopendoar:34982025-09-03 09:53:52.18CONICET Digital (CONICET) - Consejo Nacional de Investigaciones Científicas y Técnicasfalse |
dc.title.none.fl_str_mv |
Optimization of the preconcentration system of cadmium with 1(2-thiazolylazo)-p-cresol using a knotted reactor and flame atomic absorption spectrometric detection |
title |
Optimization of the preconcentration system of cadmium with 1(2-thiazolylazo)-p-cresol using a knotted reactor and flame atomic absorption spectrometric detection |
spellingShingle |
Optimization of the preconcentration system of cadmium with 1(2-thiazolylazo)-p-cresol using a knotted reactor and flame atomic absorption spectrometric detection Cerutti, Estela Soledad CADMIUM DETERMINATION FAAS KNOTTED REACTOR ON-LINE PRECONCENTRATION WATER SAMPLES, 1(2-THIAZOLYLAZO)-P-CRESOL |
title_short |
Optimization of the preconcentration system of cadmium with 1(2-thiazolylazo)-p-cresol using a knotted reactor and flame atomic absorption spectrometric detection |
title_full |
Optimization of the preconcentration system of cadmium with 1(2-thiazolylazo)-p-cresol using a knotted reactor and flame atomic absorption spectrometric detection |
title_fullStr |
Optimization of the preconcentration system of cadmium with 1(2-thiazolylazo)-p-cresol using a knotted reactor and flame atomic absorption spectrometric detection |
title_full_unstemmed |
Optimization of the preconcentration system of cadmium with 1(2-thiazolylazo)-p-cresol using a knotted reactor and flame atomic absorption spectrometric detection |
title_sort |
Optimization of the preconcentration system of cadmium with 1(2-thiazolylazo)-p-cresol using a knotted reactor and flame atomic absorption spectrometric detection |
dc.creator.none.fl_str_mv |
Cerutti, Estela Soledad Ferreira, Sergio Luis Costa Gasquez, José Antonio Olsina, Roberto Antonio Martinez, Luis Dante |
author |
Cerutti, Estela Soledad |
author_facet |
Cerutti, Estela Soledad Ferreira, Sergio Luis Costa Gasquez, José Antonio Olsina, Roberto Antonio Martinez, Luis Dante |
author_role |
author |
author2 |
Ferreira, Sergio Luis Costa Gasquez, José Antonio Olsina, Roberto Antonio Martinez, Luis Dante |
author2_role |
author author author author |
dc.subject.none.fl_str_mv |
CADMIUM DETERMINATION FAAS KNOTTED REACTOR ON-LINE PRECONCENTRATION WATER SAMPLES, 1(2-THIAZOLYLAZO)-P-CRESOL |
topic |
CADMIUM DETERMINATION FAAS KNOTTED REACTOR ON-LINE PRECONCENTRATION WATER SAMPLES, 1(2-THIAZOLYLAZO)-P-CRESOL |
purl_subject.fl_str_mv |
https://purl.org/becyt/ford/1.4 https://purl.org/becyt/ford/1 |
dc.description.none.fl_txt_mv |
The present paper proposes an on-line preconcentration procedure for cadmium determination in drinking water samples. It is based on the precipitation of cadmium(II) ions on a knotted reactor (KR) using 1(2-thiazolylazo)-p-cresol (TAC) as complexing reagent. The optimization step was performed using a full factorial design involving the variables: pH, eluent concentration (nitric acid) and TAC concentration. The results of this experiment demonstrated that these variables at chosen levels are not statistically significant. Under optimized experimental established conditions, analytical parameters for the preconcentration method were: a detection limit of 40.0 ng/l, precision as relative standard deviation (RSD) of 1.2 and 1.0%, for cadmium concentration of 2.5 and 20.0 μg/l, respectively. The preconcentration factor considering the slopes of the analytical curves with and without preconcentration is 23 for a sample volume of 10 ml. This system shows a sampling frequency of 25 h-1. In order to check the accuracy, the standard reference material, NIST SRM 1643d trace elements in water was analyzed. A comparison, using t-test demonstrates that there is not significant difference among the achieved results with proposed method and the certified values. The addition/recovery experiments in the samples analyzed demonstrated the accuracy and applicability of the system developed for cadmium determination in water samples. Fil: Cerutti, Estela Soledad. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - San Luis. Instituto de Química de San Luis. Universidad Nacional de San Luis. Facultad de Química, Bioquímica y Farmacia. Instituto de Química de San Luis; Argentina Fil: Ferreira, Sergio Luis Costa. Universidade Federal da Bahia; Brasil Fil: Gasquez, José Antonio. Universidad Nacional de San Luis; Argentina Fil: Olsina, Roberto Antonio. Consejo Nacional de Investigaciones Científicas y Técnicas; Argentina. Universidad Nacional de San Luis; Argentina Fil: Martinez, Luis Dante. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - San Luis. Instituto de Química de San Luis. Universidad Nacional de San Luis. Facultad de Química, Bioquímica y Farmacia. Instituto de Química de San Luis; Argentina |
description |
The present paper proposes an on-line preconcentration procedure for cadmium determination in drinking water samples. It is based on the precipitation of cadmium(II) ions on a knotted reactor (KR) using 1(2-thiazolylazo)-p-cresol (TAC) as complexing reagent. The optimization step was performed using a full factorial design involving the variables: pH, eluent concentration (nitric acid) and TAC concentration. The results of this experiment demonstrated that these variables at chosen levels are not statistically significant. Under optimized experimental established conditions, analytical parameters for the preconcentration method were: a detection limit of 40.0 ng/l, precision as relative standard deviation (RSD) of 1.2 and 1.0%, for cadmium concentration of 2.5 and 20.0 μg/l, respectively. The preconcentration factor considering the slopes of the analytical curves with and without preconcentration is 23 for a sample volume of 10 ml. This system shows a sampling frequency of 25 h-1. In order to check the accuracy, the standard reference material, NIST SRM 1643d trace elements in water was analyzed. A comparison, using t-test demonstrates that there is not significant difference among the achieved results with proposed method and the certified values. The addition/recovery experiments in the samples analyzed demonstrated the accuracy and applicability of the system developed for cadmium determination in water samples. |
publishDate |
2004 |
dc.date.none.fl_str_mv |
2004-08 |
dc.type.none.fl_str_mv |
info:eu-repo/semantics/article info:eu-repo/semantics/publishedVersion http://purl.org/coar/resource_type/c_6501 info:ar-repo/semantics/articulo |
format |
article |
status_str |
publishedVersion |
dc.identifier.none.fl_str_mv |
http://hdl.handle.net/11336/97974 Cerutti, Estela Soledad; Ferreira, Sergio Luis Costa; Gasquez, José Antonio; Olsina, Roberto Antonio; Martinez, Luis Dante; Optimization of the preconcentration system of cadmium with 1(2-thiazolylazo)-p-cresol using a knotted reactor and flame atomic absorption spectrometric detection; Elsevier Science; Journal of Hazardous Materials; 112; 3; 8-2004; 279-283 0304-3894 CONICET Digital CONICET |
url |
http://hdl.handle.net/11336/97974 |
identifier_str_mv |
Cerutti, Estela Soledad; Ferreira, Sergio Luis Costa; Gasquez, José Antonio; Olsina, Roberto Antonio; Martinez, Luis Dante; Optimization of the preconcentration system of cadmium with 1(2-thiazolylazo)-p-cresol using a knotted reactor and flame atomic absorption spectrometric detection; Elsevier Science; Journal of Hazardous Materials; 112; 3; 8-2004; 279-283 0304-3894 CONICET Digital CONICET |
dc.language.none.fl_str_mv |
eng |
language |
eng |
dc.relation.none.fl_str_mv |
info:eu-repo/semantics/altIdentifier/doi/10.1016/j.jhazmat.2004.05.017 info:eu-repo/semantics/altIdentifier/url/https://www.sciencedirect.com/science/article/pii/S0304389404002626 |
dc.rights.none.fl_str_mv |
info:eu-repo/semantics/openAccess https://creativecommons.org/licenses/by-nc-nd/2.5/ar/ |
eu_rights_str_mv |
openAccess |
rights_invalid_str_mv |
https://creativecommons.org/licenses/by-nc-nd/2.5/ar/ |
dc.format.none.fl_str_mv |
application/pdf application/pdf application/pdf |
dc.publisher.none.fl_str_mv |
Elsevier Science |
publisher.none.fl_str_mv |
Elsevier Science |
dc.source.none.fl_str_mv |
reponame:CONICET Digital (CONICET) instname:Consejo Nacional de Investigaciones Científicas y Técnicas |
reponame_str |
CONICET Digital (CONICET) |
collection |
CONICET Digital (CONICET) |
instname_str |
Consejo Nacional de Investigaciones Científicas y Técnicas |
repository.name.fl_str_mv |
CONICET Digital (CONICET) - Consejo Nacional de Investigaciones Científicas y Técnicas |
repository.mail.fl_str_mv |
dasensio@conicet.gov.ar; lcarlino@conicet.gov.ar |
_version_ |
1842269253521113088 |
score |
13.13397 |