A Two-Wired Ultra-High Input Impedance Active Electrode

Autores
Guerrero, Federico Nicolás; Spinelli, Enrique Mario
Año de publicación
2018
Idioma
inglés
Tipo de recurso
artículo
Estado
versión publicada
Descripción
This paper presents a novel two-wired active electrode that achieves ultrahigh input impedance using power supply bootstrapping. The proposed circuit reduces the input capacitance of a buffer amplifier while enabling measurements using leads with only two wires, providing a low-complexity and low-cost solution for interference rejection and artifact reduction in dc-coupled dry-contact biopotential measurements. An implemented prototype shows that, even using standard operational amplifiers, an input capacitance as low as 71 fF can be obtained, maintaining a high impedance in a 0-1 kHz bandwidth, sufficient for ECG, EEG, and EMG measurements. The circuit has a simple and easily replicable implementation that requires no individual adjustment. A common mode rejection ratio (CMRR) above 103 dB at 50 Hz was achieved and the increased rejection to interference due to the potential divider effect was experimentally tested maintaining a 92-dB CMRR at 50 Hz with a 1.2-M Ω source impedance unbalance. ECG measurements were conducted to validate the active electrode against a traditional alternative, and a test with dry-contact EEG electrodes was successfully conducted. Although the proposed circuit is intended to acquire superficial electrophysiological signals using dry electrodes, it can be used for measurement from other high-impedance sources, such as micropipette electrodes.
Fil: Guerrero, Federico Nicolas. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - La Plata. Instituto de Investigaciones en Electrónica, Control y Procesamiento de Señales. Universidad Nacional de La Plata. Instituto de Investigaciones en Electrónica, Control y Procesamiento de Señales; Argentina
Fil: Spinelli, Enrique Mario. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - La Plata. Instituto de Investigaciones en Electrónica, Control y Procesamiento de Señales. Universidad Nacional de La Plata. Instituto de Investigaciones en Electrónica, Control y Procesamiento de Señales; Argentina
Materia
ACTIVE ELECTRODES
BIOPOTENTIAL MEASUREMENTS
DRY-CONTACT ELECTRODES
POWER SUPPLY BOOTSTRAPPING
TWO-WIRED BUFFER
Nivel de accesibilidad
acceso abierto
Condiciones de uso
https://creativecommons.org/licenses/by-nc-sa/2.5/ar/
Repositorio
CONICET Digital (CONICET)
Institución
Consejo Nacional de Investigaciones Científicas y Técnicas
OAI Identificador
oai:ri.conicet.gov.ar:11336/93726

id CONICETDig_777ada0cff95f679fafc73afc21f3160
oai_identifier_str oai:ri.conicet.gov.ar:11336/93726
network_acronym_str CONICETDig
repository_id_str 3498
network_name_str CONICET Digital (CONICET)
spelling A Two-Wired Ultra-High Input Impedance Active ElectrodeGuerrero, Federico NicolásSpinelli, Enrique MarioACTIVE ELECTRODESBIOPOTENTIAL MEASUREMENTSDRY-CONTACT ELECTRODESPOWER SUPPLY BOOTSTRAPPINGTWO-WIRED BUFFERhttps://purl.org/becyt/ford/2.2https://purl.org/becyt/ford/2This paper presents a novel two-wired active electrode that achieves ultrahigh input impedance using power supply bootstrapping. The proposed circuit reduces the input capacitance of a buffer amplifier while enabling measurements using leads with only two wires, providing a low-complexity and low-cost solution for interference rejection and artifact reduction in dc-coupled dry-contact biopotential measurements. An implemented prototype shows that, even using standard operational amplifiers, an input capacitance as low as 71 fF can be obtained, maintaining a high impedance in a 0-1 kHz bandwidth, sufficient for ECG, EEG, and EMG measurements. The circuit has a simple and easily replicable implementation that requires no individual adjustment. A common mode rejection ratio (CMRR) above 103 dB at 50 Hz was achieved and the increased rejection to interference due to the potential divider effect was experimentally tested maintaining a 92-dB CMRR at 50 Hz with a 1.2-M Ω source impedance unbalance. ECG measurements were conducted to validate the active electrode against a traditional alternative, and a test with dry-contact EEG electrodes was successfully conducted. Although the proposed circuit is intended to acquire superficial electrophysiological signals using dry electrodes, it can be used for measurement from other high-impedance sources, such as micropipette electrodes.Fil: Guerrero, Federico Nicolas. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - La Plata. Instituto de Investigaciones en Electrónica, Control y Procesamiento de Señales. Universidad Nacional de La Plata. Instituto de Investigaciones en Electrónica, Control y Procesamiento de Señales; ArgentinaFil: Spinelli, Enrique Mario. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - La Plata. Instituto de Investigaciones en Electrónica, Control y Procesamiento de Señales. Universidad Nacional de La Plata. Instituto de Investigaciones en Electrónica, Control y Procesamiento de Señales; ArgentinaInstitute of Electrical and Electronics Engineers2018-04info:eu-repo/semantics/articleinfo:eu-repo/semantics/publishedVersionhttp://purl.org/coar/resource_type/c_6501info:ar-repo/semantics/articuloapplication/pdfapplication/pdfapplication/pdfhttp://hdl.handle.net/11336/93726Guerrero, Federico Nicolás; Spinelli, Enrique Mario; A Two-Wired Ultra-High Input Impedance Active Electrode; Institute of Electrical and Electronics Engineers; IEEE Transactions On Biomedical Circuits And Systems; 12; 2; 4-2018; 437-4451932-4545CONICET DigitalCONICETenginfo:eu-repo/semantics/altIdentifier/url/http://ieeexplore.ieee.org/document/8305528/info:eu-repo/semantics/altIdentifier/doi/10.1109/TBCAS.2018.2796581info:eu-repo/semantics/openAccesshttps://creativecommons.org/licenses/by-nc-sa/2.5/ar/reponame:CONICET Digital (CONICET)instname:Consejo Nacional de Investigaciones Científicas y Técnicas2025-09-03T09:50:47Zoai:ri.conicet.gov.ar:11336/93726instacron:CONICETInstitucionalhttp://ri.conicet.gov.ar/Organismo científico-tecnológicoNo correspondehttp://ri.conicet.gov.ar/oai/requestdasensio@conicet.gov.ar; lcarlino@conicet.gov.arArgentinaNo correspondeNo correspondeNo correspondeopendoar:34982025-09-03 09:50:47.991CONICET Digital (CONICET) - Consejo Nacional de Investigaciones Científicas y Técnicasfalse
dc.title.none.fl_str_mv A Two-Wired Ultra-High Input Impedance Active Electrode
title A Two-Wired Ultra-High Input Impedance Active Electrode
spellingShingle A Two-Wired Ultra-High Input Impedance Active Electrode
Guerrero, Federico Nicolás
ACTIVE ELECTRODES
BIOPOTENTIAL MEASUREMENTS
DRY-CONTACT ELECTRODES
POWER SUPPLY BOOTSTRAPPING
TWO-WIRED BUFFER
title_short A Two-Wired Ultra-High Input Impedance Active Electrode
title_full A Two-Wired Ultra-High Input Impedance Active Electrode
title_fullStr A Two-Wired Ultra-High Input Impedance Active Electrode
title_full_unstemmed A Two-Wired Ultra-High Input Impedance Active Electrode
title_sort A Two-Wired Ultra-High Input Impedance Active Electrode
dc.creator.none.fl_str_mv Guerrero, Federico Nicolás
Spinelli, Enrique Mario
author Guerrero, Federico Nicolás
author_facet Guerrero, Federico Nicolás
Spinelli, Enrique Mario
author_role author
author2 Spinelli, Enrique Mario
author2_role author
dc.subject.none.fl_str_mv ACTIVE ELECTRODES
BIOPOTENTIAL MEASUREMENTS
DRY-CONTACT ELECTRODES
POWER SUPPLY BOOTSTRAPPING
TWO-WIRED BUFFER
topic ACTIVE ELECTRODES
BIOPOTENTIAL MEASUREMENTS
DRY-CONTACT ELECTRODES
POWER SUPPLY BOOTSTRAPPING
TWO-WIRED BUFFER
purl_subject.fl_str_mv https://purl.org/becyt/ford/2.2
https://purl.org/becyt/ford/2
dc.description.none.fl_txt_mv This paper presents a novel two-wired active electrode that achieves ultrahigh input impedance using power supply bootstrapping. The proposed circuit reduces the input capacitance of a buffer amplifier while enabling measurements using leads with only two wires, providing a low-complexity and low-cost solution for interference rejection and artifact reduction in dc-coupled dry-contact biopotential measurements. An implemented prototype shows that, even using standard operational amplifiers, an input capacitance as low as 71 fF can be obtained, maintaining a high impedance in a 0-1 kHz bandwidth, sufficient for ECG, EEG, and EMG measurements. The circuit has a simple and easily replicable implementation that requires no individual adjustment. A common mode rejection ratio (CMRR) above 103 dB at 50 Hz was achieved and the increased rejection to interference due to the potential divider effect was experimentally tested maintaining a 92-dB CMRR at 50 Hz with a 1.2-M Ω source impedance unbalance. ECG measurements were conducted to validate the active electrode against a traditional alternative, and a test with dry-contact EEG electrodes was successfully conducted. Although the proposed circuit is intended to acquire superficial electrophysiological signals using dry electrodes, it can be used for measurement from other high-impedance sources, such as micropipette electrodes.
Fil: Guerrero, Federico Nicolas. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - La Plata. Instituto de Investigaciones en Electrónica, Control y Procesamiento de Señales. Universidad Nacional de La Plata. Instituto de Investigaciones en Electrónica, Control y Procesamiento de Señales; Argentina
Fil: Spinelli, Enrique Mario. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - La Plata. Instituto de Investigaciones en Electrónica, Control y Procesamiento de Señales. Universidad Nacional de La Plata. Instituto de Investigaciones en Electrónica, Control y Procesamiento de Señales; Argentina
description This paper presents a novel two-wired active electrode that achieves ultrahigh input impedance using power supply bootstrapping. The proposed circuit reduces the input capacitance of a buffer amplifier while enabling measurements using leads with only two wires, providing a low-complexity and low-cost solution for interference rejection and artifact reduction in dc-coupled dry-contact biopotential measurements. An implemented prototype shows that, even using standard operational amplifiers, an input capacitance as low as 71 fF can be obtained, maintaining a high impedance in a 0-1 kHz bandwidth, sufficient for ECG, EEG, and EMG measurements. The circuit has a simple and easily replicable implementation that requires no individual adjustment. A common mode rejection ratio (CMRR) above 103 dB at 50 Hz was achieved and the increased rejection to interference due to the potential divider effect was experimentally tested maintaining a 92-dB CMRR at 50 Hz with a 1.2-M Ω source impedance unbalance. ECG measurements were conducted to validate the active electrode against a traditional alternative, and a test with dry-contact EEG electrodes was successfully conducted. Although the proposed circuit is intended to acquire superficial electrophysiological signals using dry electrodes, it can be used for measurement from other high-impedance sources, such as micropipette electrodes.
publishDate 2018
dc.date.none.fl_str_mv 2018-04
dc.type.none.fl_str_mv info:eu-repo/semantics/article
info:eu-repo/semantics/publishedVersion
http://purl.org/coar/resource_type/c_6501
info:ar-repo/semantics/articulo
format article
status_str publishedVersion
dc.identifier.none.fl_str_mv http://hdl.handle.net/11336/93726
Guerrero, Federico Nicolás; Spinelli, Enrique Mario; A Two-Wired Ultra-High Input Impedance Active Electrode; Institute of Electrical and Electronics Engineers; IEEE Transactions On Biomedical Circuits And Systems; 12; 2; 4-2018; 437-445
1932-4545
CONICET Digital
CONICET
url http://hdl.handle.net/11336/93726
identifier_str_mv Guerrero, Federico Nicolás; Spinelli, Enrique Mario; A Two-Wired Ultra-High Input Impedance Active Electrode; Institute of Electrical and Electronics Engineers; IEEE Transactions On Biomedical Circuits And Systems; 12; 2; 4-2018; 437-445
1932-4545
CONICET Digital
CONICET
dc.language.none.fl_str_mv eng
language eng
dc.relation.none.fl_str_mv info:eu-repo/semantics/altIdentifier/url/http://ieeexplore.ieee.org/document/8305528/
info:eu-repo/semantics/altIdentifier/doi/10.1109/TBCAS.2018.2796581
dc.rights.none.fl_str_mv info:eu-repo/semantics/openAccess
https://creativecommons.org/licenses/by-nc-sa/2.5/ar/
eu_rights_str_mv openAccess
rights_invalid_str_mv https://creativecommons.org/licenses/by-nc-sa/2.5/ar/
dc.format.none.fl_str_mv application/pdf
application/pdf
application/pdf
dc.publisher.none.fl_str_mv Institute of Electrical and Electronics Engineers
publisher.none.fl_str_mv Institute of Electrical and Electronics Engineers
dc.source.none.fl_str_mv reponame:CONICET Digital (CONICET)
instname:Consejo Nacional de Investigaciones Científicas y Técnicas
reponame_str CONICET Digital (CONICET)
collection CONICET Digital (CONICET)
instname_str Consejo Nacional de Investigaciones Científicas y Técnicas
repository.name.fl_str_mv CONICET Digital (CONICET) - Consejo Nacional de Investigaciones Científicas y Técnicas
repository.mail.fl_str_mv dasensio@conicet.gov.ar; lcarlino@conicet.gov.ar
_version_ 1842269053567107072
score 13.13397