A Two-Wired Ultra-High Input Impedance Active Electrode
- Autores
- Guerrero, Federico Nicolás; Spinelli, Enrique Mario
- Año de publicación
- 2018
- Idioma
- inglés
- Tipo de recurso
- artículo
- Estado
- versión publicada
- Descripción
- This paper presents a novel two-wired active electrode that achieves ultrahigh input impedance using power supply bootstrapping. The proposed circuit reduces the input capacitance of a buffer amplifier while enabling measurements using leads with only two wires, providing a low-complexity and low-cost solution for interference rejection and artifact reduction in dc-coupled dry-contact biopotential measurements. An implemented prototype shows that, even using standard operational amplifiers, an input capacitance as low as 71 fF can be obtained, maintaining a high impedance in a 0-1 kHz bandwidth, sufficient for ECG, EEG, and EMG measurements. The circuit has a simple and easily replicable implementation that requires no individual adjustment. A common mode rejection ratio (CMRR) above 103 dB at 50 Hz was achieved and the increased rejection to interference due to the potential divider effect was experimentally tested maintaining a 92-dB CMRR at 50 Hz with a 1.2-M Ω source impedance unbalance. ECG measurements were conducted to validate the active electrode against a traditional alternative, and a test with dry-contact EEG electrodes was successfully conducted. Although the proposed circuit is intended to acquire superficial electrophysiological signals using dry electrodes, it can be used for measurement from other high-impedance sources, such as micropipette electrodes.
Fil: Guerrero, Federico Nicolas. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - La Plata. Instituto de Investigaciones en Electrónica, Control y Procesamiento de Señales. Universidad Nacional de La Plata. Instituto de Investigaciones en Electrónica, Control y Procesamiento de Señales; Argentina
Fil: Spinelli, Enrique Mario. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - La Plata. Instituto de Investigaciones en Electrónica, Control y Procesamiento de Señales. Universidad Nacional de La Plata. Instituto de Investigaciones en Electrónica, Control y Procesamiento de Señales; Argentina - Materia
-
ACTIVE ELECTRODES
BIOPOTENTIAL MEASUREMENTS
DRY-CONTACT ELECTRODES
POWER SUPPLY BOOTSTRAPPING
TWO-WIRED BUFFER - Nivel de accesibilidad
- acceso abierto
- Condiciones de uso
- https://creativecommons.org/licenses/by-nc-sa/2.5/ar/
- Repositorio
.jpg)
- Institución
- Consejo Nacional de Investigaciones Científicas y Técnicas
- OAI Identificador
- oai:ri.conicet.gov.ar:11336/93726
Ver los metadatos del registro completo
| id |
CONICETDig_777ada0cff95f679fafc73afc21f3160 |
|---|---|
| oai_identifier_str |
oai:ri.conicet.gov.ar:11336/93726 |
| network_acronym_str |
CONICETDig |
| repository_id_str |
3498 |
| network_name_str |
CONICET Digital (CONICET) |
| spelling |
A Two-Wired Ultra-High Input Impedance Active ElectrodeGuerrero, Federico NicolásSpinelli, Enrique MarioACTIVE ELECTRODESBIOPOTENTIAL MEASUREMENTSDRY-CONTACT ELECTRODESPOWER SUPPLY BOOTSTRAPPINGTWO-WIRED BUFFERhttps://purl.org/becyt/ford/2.2https://purl.org/becyt/ford/2This paper presents a novel two-wired active electrode that achieves ultrahigh input impedance using power supply bootstrapping. The proposed circuit reduces the input capacitance of a buffer amplifier while enabling measurements using leads with only two wires, providing a low-complexity and low-cost solution for interference rejection and artifact reduction in dc-coupled dry-contact biopotential measurements. An implemented prototype shows that, even using standard operational amplifiers, an input capacitance as low as 71 fF can be obtained, maintaining a high impedance in a 0-1 kHz bandwidth, sufficient for ECG, EEG, and EMG measurements. The circuit has a simple and easily replicable implementation that requires no individual adjustment. A common mode rejection ratio (CMRR) above 103 dB at 50 Hz was achieved and the increased rejection to interference due to the potential divider effect was experimentally tested maintaining a 92-dB CMRR at 50 Hz with a 1.2-M Ω source impedance unbalance. ECG measurements were conducted to validate the active electrode against a traditional alternative, and a test with dry-contact EEG electrodes was successfully conducted. Although the proposed circuit is intended to acquire superficial electrophysiological signals using dry electrodes, it can be used for measurement from other high-impedance sources, such as micropipette electrodes.Fil: Guerrero, Federico Nicolas. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - La Plata. Instituto de Investigaciones en Electrónica, Control y Procesamiento de Señales. Universidad Nacional de La Plata. Instituto de Investigaciones en Electrónica, Control y Procesamiento de Señales; ArgentinaFil: Spinelli, Enrique Mario. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - La Plata. Instituto de Investigaciones en Electrónica, Control y Procesamiento de Señales. Universidad Nacional de La Plata. Instituto de Investigaciones en Electrónica, Control y Procesamiento de Señales; ArgentinaInstitute of Electrical and Electronics Engineers2018-04info:eu-repo/semantics/articleinfo:eu-repo/semantics/publishedVersionhttp://purl.org/coar/resource_type/c_6501info:ar-repo/semantics/articuloapplication/pdfapplication/pdfapplication/pdfhttp://hdl.handle.net/11336/93726Guerrero, Federico Nicolás; Spinelli, Enrique Mario; A Two-Wired Ultra-High Input Impedance Active Electrode; Institute of Electrical and Electronics Engineers; IEEE Transactions On Biomedical Circuits And Systems; 12; 2; 4-2018; 437-4451932-4545CONICET DigitalCONICETenginfo:eu-repo/semantics/altIdentifier/url/http://ieeexplore.ieee.org/document/8305528/info:eu-repo/semantics/altIdentifier/doi/10.1109/TBCAS.2018.2796581info:eu-repo/semantics/openAccesshttps://creativecommons.org/licenses/by-nc-sa/2.5/ar/reponame:CONICET Digital (CONICET)instname:Consejo Nacional de Investigaciones Científicas y Técnicas2025-11-05T09:48:08Zoai:ri.conicet.gov.ar:11336/93726instacron:CONICETInstitucionalhttp://ri.conicet.gov.ar/Organismo científico-tecnológicoNo correspondehttp://ri.conicet.gov.ar/oai/requestdasensio@conicet.gov.ar; lcarlino@conicet.gov.arArgentinaNo correspondeNo correspondeNo correspondeopendoar:34982025-11-05 09:48:08.66CONICET Digital (CONICET) - Consejo Nacional de Investigaciones Científicas y Técnicasfalse |
| dc.title.none.fl_str_mv |
A Two-Wired Ultra-High Input Impedance Active Electrode |
| title |
A Two-Wired Ultra-High Input Impedance Active Electrode |
| spellingShingle |
A Two-Wired Ultra-High Input Impedance Active Electrode Guerrero, Federico Nicolás ACTIVE ELECTRODES BIOPOTENTIAL MEASUREMENTS DRY-CONTACT ELECTRODES POWER SUPPLY BOOTSTRAPPING TWO-WIRED BUFFER |
| title_short |
A Two-Wired Ultra-High Input Impedance Active Electrode |
| title_full |
A Two-Wired Ultra-High Input Impedance Active Electrode |
| title_fullStr |
A Two-Wired Ultra-High Input Impedance Active Electrode |
| title_full_unstemmed |
A Two-Wired Ultra-High Input Impedance Active Electrode |
| title_sort |
A Two-Wired Ultra-High Input Impedance Active Electrode |
| dc.creator.none.fl_str_mv |
Guerrero, Federico Nicolás Spinelli, Enrique Mario |
| author |
Guerrero, Federico Nicolás |
| author_facet |
Guerrero, Federico Nicolás Spinelli, Enrique Mario |
| author_role |
author |
| author2 |
Spinelli, Enrique Mario |
| author2_role |
author |
| dc.subject.none.fl_str_mv |
ACTIVE ELECTRODES BIOPOTENTIAL MEASUREMENTS DRY-CONTACT ELECTRODES POWER SUPPLY BOOTSTRAPPING TWO-WIRED BUFFER |
| topic |
ACTIVE ELECTRODES BIOPOTENTIAL MEASUREMENTS DRY-CONTACT ELECTRODES POWER SUPPLY BOOTSTRAPPING TWO-WIRED BUFFER |
| purl_subject.fl_str_mv |
https://purl.org/becyt/ford/2.2 https://purl.org/becyt/ford/2 |
| dc.description.none.fl_txt_mv |
This paper presents a novel two-wired active electrode that achieves ultrahigh input impedance using power supply bootstrapping. The proposed circuit reduces the input capacitance of a buffer amplifier while enabling measurements using leads with only two wires, providing a low-complexity and low-cost solution for interference rejection and artifact reduction in dc-coupled dry-contact biopotential measurements. An implemented prototype shows that, even using standard operational amplifiers, an input capacitance as low as 71 fF can be obtained, maintaining a high impedance in a 0-1 kHz bandwidth, sufficient for ECG, EEG, and EMG measurements. The circuit has a simple and easily replicable implementation that requires no individual adjustment. A common mode rejection ratio (CMRR) above 103 dB at 50 Hz was achieved and the increased rejection to interference due to the potential divider effect was experimentally tested maintaining a 92-dB CMRR at 50 Hz with a 1.2-M Ω source impedance unbalance. ECG measurements were conducted to validate the active electrode against a traditional alternative, and a test with dry-contact EEG electrodes was successfully conducted. Although the proposed circuit is intended to acquire superficial electrophysiological signals using dry electrodes, it can be used for measurement from other high-impedance sources, such as micropipette electrodes. Fil: Guerrero, Federico Nicolas. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - La Plata. Instituto de Investigaciones en Electrónica, Control y Procesamiento de Señales. Universidad Nacional de La Plata. Instituto de Investigaciones en Electrónica, Control y Procesamiento de Señales; Argentina Fil: Spinelli, Enrique Mario. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - La Plata. Instituto de Investigaciones en Electrónica, Control y Procesamiento de Señales. Universidad Nacional de La Plata. Instituto de Investigaciones en Electrónica, Control y Procesamiento de Señales; Argentina |
| description |
This paper presents a novel two-wired active electrode that achieves ultrahigh input impedance using power supply bootstrapping. The proposed circuit reduces the input capacitance of a buffer amplifier while enabling measurements using leads with only two wires, providing a low-complexity and low-cost solution for interference rejection and artifact reduction in dc-coupled dry-contact biopotential measurements. An implemented prototype shows that, even using standard operational amplifiers, an input capacitance as low as 71 fF can be obtained, maintaining a high impedance in a 0-1 kHz bandwidth, sufficient for ECG, EEG, and EMG measurements. The circuit has a simple and easily replicable implementation that requires no individual adjustment. A common mode rejection ratio (CMRR) above 103 dB at 50 Hz was achieved and the increased rejection to interference due to the potential divider effect was experimentally tested maintaining a 92-dB CMRR at 50 Hz with a 1.2-M Ω source impedance unbalance. ECG measurements were conducted to validate the active electrode against a traditional alternative, and a test with dry-contact EEG electrodes was successfully conducted. Although the proposed circuit is intended to acquire superficial electrophysiological signals using dry electrodes, it can be used for measurement from other high-impedance sources, such as micropipette electrodes. |
| publishDate |
2018 |
| dc.date.none.fl_str_mv |
2018-04 |
| dc.type.none.fl_str_mv |
info:eu-repo/semantics/article info:eu-repo/semantics/publishedVersion http://purl.org/coar/resource_type/c_6501 info:ar-repo/semantics/articulo |
| format |
article |
| status_str |
publishedVersion |
| dc.identifier.none.fl_str_mv |
http://hdl.handle.net/11336/93726 Guerrero, Federico Nicolás; Spinelli, Enrique Mario; A Two-Wired Ultra-High Input Impedance Active Electrode; Institute of Electrical and Electronics Engineers; IEEE Transactions On Biomedical Circuits And Systems; 12; 2; 4-2018; 437-445 1932-4545 CONICET Digital CONICET |
| url |
http://hdl.handle.net/11336/93726 |
| identifier_str_mv |
Guerrero, Federico Nicolás; Spinelli, Enrique Mario; A Two-Wired Ultra-High Input Impedance Active Electrode; Institute of Electrical and Electronics Engineers; IEEE Transactions On Biomedical Circuits And Systems; 12; 2; 4-2018; 437-445 1932-4545 CONICET Digital CONICET |
| dc.language.none.fl_str_mv |
eng |
| language |
eng |
| dc.relation.none.fl_str_mv |
info:eu-repo/semantics/altIdentifier/url/http://ieeexplore.ieee.org/document/8305528/ info:eu-repo/semantics/altIdentifier/doi/10.1109/TBCAS.2018.2796581 |
| dc.rights.none.fl_str_mv |
info:eu-repo/semantics/openAccess https://creativecommons.org/licenses/by-nc-sa/2.5/ar/ |
| eu_rights_str_mv |
openAccess |
| rights_invalid_str_mv |
https://creativecommons.org/licenses/by-nc-sa/2.5/ar/ |
| dc.format.none.fl_str_mv |
application/pdf application/pdf application/pdf |
| dc.publisher.none.fl_str_mv |
Institute of Electrical and Electronics Engineers |
| publisher.none.fl_str_mv |
Institute of Electrical and Electronics Engineers |
| dc.source.none.fl_str_mv |
reponame:CONICET Digital (CONICET) instname:Consejo Nacional de Investigaciones Científicas y Técnicas |
| reponame_str |
CONICET Digital (CONICET) |
| collection |
CONICET Digital (CONICET) |
| instname_str |
Consejo Nacional de Investigaciones Científicas y Técnicas |
| repository.name.fl_str_mv |
CONICET Digital (CONICET) - Consejo Nacional de Investigaciones Científicas y Técnicas |
| repository.mail.fl_str_mv |
dasensio@conicet.gov.ar; lcarlino@conicet.gov.ar |
| _version_ |
1847977158449823744 |
| score |
13.087074 |