Effect of the Electrode Impedance in Improved Buffer Amplifier for Bioelectric Recordings
- Autores
- Valverde, Esteban Raul; Arini, Pedro David; Biagetti, Marcelo; Bertran, Guillermo Claudio; Quinteiro, Ricardo Alberto
- Año de publicación
- 2004
- Idioma
- inglés
- Tipo de recurso
- artículo
- Estado
- versión publicada
- Descripción
- We analyzed the effects of the electrode impedance on the transfer response of a one-stage improved buffer amplifier. The electrode DC resistance (Rd) modifies the one-stage buffer transfer response. We found a limit electrode resistance (Rd(lim)) which depends on the transfer damping factor (∈). If Rd is lower than 86.5 kW, the transfer response of the buffer fulfils the American Heart Association, AHA, recommendations, but when Rd is greater than Rd(lim) it must be cautiously weighted-up because its influence in the transfer response becomes appreciable. The maximum Rd that can be driven by the buffer is 1.2 MW. Higher values never fulfil the AHA recommendations. Therefore, electrodes with higher impedance should not be used with this kind of buffer. On the other hand, when this buffer is used to build-in an instrumentation amplifier (IA) for bipolar recording, the common-mode rejection ratio (CMRR) is sensitive to the electrode type used.
Fil: Valverde, Esteban Raul. Consejo Nacional de Investigaciones Científicas y Técnicas; Argentina. Universidad Favaloro; Argentina
Fil: Arini, Pedro David. Consejo Nacional de Investigaciones Científicas y Técnicas. Oficina de Coordinación Administrativa Saavedra 15. Instituto Argentino de Matemática Alberto Calderón; Argentina. Universidad Favaloro; Argentina
Fil: Biagetti, Marcelo. Universidad Favaloro; Argentina
Fil: Bertran, Guillermo Claudio. Universidad Favaloro; Argentina. Consejo Nacional de Investigaciones Científicas y Técnicas; Argentina
Fil: Quinteiro, Ricardo Alberto. Universidad Favaloro; Argentina - Materia
-
ELECTRODE IMPEDANCE
BUFFER AMPLIFIER
BIOELECTRIC RECORDINGS - Nivel de accesibilidad
- acceso abierto
- Condiciones de uso
- https://creativecommons.org/licenses/by-nc-sa/2.5/ar/
- Repositorio
- Institución
- Consejo Nacional de Investigaciones Científicas y Técnicas
- OAI Identificador
- oai:ri.conicet.gov.ar:11336/110162
Ver los metadatos del registro completo
id |
CONICETDig_05dad25e1b453e97e762e816860cd5b3 |
---|---|
oai_identifier_str |
oai:ri.conicet.gov.ar:11336/110162 |
network_acronym_str |
CONICETDig |
repository_id_str |
3498 |
network_name_str |
CONICET Digital (CONICET) |
spelling |
Effect of the Electrode Impedance in Improved Buffer Amplifier for Bioelectric RecordingsValverde, Esteban RaulArini, Pedro DavidBiagetti, MarceloBertran, Guillermo ClaudioQuinteiro, Ricardo AlbertoELECTRODE IMPEDANCEBUFFER AMPLIFIERBIOELECTRIC RECORDINGShttps://purl.org/becyt/ford/2.2https://purl.org/becyt/ford/2We analyzed the effects of the electrode impedance on the transfer response of a one-stage improved buffer amplifier. The electrode DC resistance (Rd) modifies the one-stage buffer transfer response. We found a limit electrode resistance (Rd(lim)) which depends on the transfer damping factor (∈). If Rd is lower than 86.5 kW, the transfer response of the buffer fulfils the American Heart Association, AHA, recommendations, but when Rd is greater than Rd(lim) it must be cautiously weighted-up because its influence in the transfer response becomes appreciable. The maximum Rd that can be driven by the buffer is 1.2 MW. Higher values never fulfil the AHA recommendations. Therefore, electrodes with higher impedance should not be used with this kind of buffer. On the other hand, when this buffer is used to build-in an instrumentation amplifier (IA) for bipolar recording, the common-mode rejection ratio (CMRR) is sensitive to the electrode type used.Fil: Valverde, Esteban Raul. Consejo Nacional de Investigaciones Científicas y Técnicas; Argentina. Universidad Favaloro; ArgentinaFil: Arini, Pedro David. Consejo Nacional de Investigaciones Científicas y Técnicas. Oficina de Coordinación Administrativa Saavedra 15. Instituto Argentino de Matemática Alberto Calderón; Argentina. Universidad Favaloro; ArgentinaFil: Biagetti, Marcelo. Universidad Favaloro; ArgentinaFil: Bertran, Guillermo Claudio. Universidad Favaloro; Argentina. Consejo Nacional de Investigaciones Científicas y Técnicas; ArgentinaFil: Quinteiro, Ricardo Alberto. Universidad Favaloro; ArgentinaTaylor & Francis2004-12info:eu-repo/semantics/articleinfo:eu-repo/semantics/publishedVersionhttp://purl.org/coar/resource_type/c_6501info:ar-repo/semantics/articuloapplication/pdfapplication/pdfapplication/pdfapplication/pdfhttp://hdl.handle.net/11336/110162Valverde, Esteban Raul; Arini, Pedro David; Biagetti, Marcelo; Bertran, Guillermo Claudio; Quinteiro, Ricardo Alberto; Effect of the Electrode Impedance in Improved Buffer Amplifier for Bioelectric Recordings; Taylor & Francis; Journal Of Medical Engineering And Technology; 28; 5; 12-2004; 217-2220309-19021464-522XCONICET DigitalCONICETenginfo:eu-repo/semantics/altIdentifier/url/https://www.tandfonline.com/doi/abs/10.1080/03091900410001662323info:eu-repo/semantics/altIdentifier/doi/10.1080/03091900410001662323info:eu-repo/semantics/openAccesshttps://creativecommons.org/licenses/by-nc-sa/2.5/ar/reponame:CONICET Digital (CONICET)instname:Consejo Nacional de Investigaciones Científicas y Técnicas2025-09-29T09:42:33Zoai:ri.conicet.gov.ar:11336/110162instacron:CONICETInstitucionalhttp://ri.conicet.gov.ar/Organismo científico-tecnológicoNo correspondehttp://ri.conicet.gov.ar/oai/requestdasensio@conicet.gov.ar; lcarlino@conicet.gov.arArgentinaNo correspondeNo correspondeNo correspondeopendoar:34982025-09-29 09:42:34.263CONICET Digital (CONICET) - Consejo Nacional de Investigaciones Científicas y Técnicasfalse |
dc.title.none.fl_str_mv |
Effect of the Electrode Impedance in Improved Buffer Amplifier for Bioelectric Recordings |
title |
Effect of the Electrode Impedance in Improved Buffer Amplifier for Bioelectric Recordings |
spellingShingle |
Effect of the Electrode Impedance in Improved Buffer Amplifier for Bioelectric Recordings Valverde, Esteban Raul ELECTRODE IMPEDANCE BUFFER AMPLIFIER BIOELECTRIC RECORDINGS |
title_short |
Effect of the Electrode Impedance in Improved Buffer Amplifier for Bioelectric Recordings |
title_full |
Effect of the Electrode Impedance in Improved Buffer Amplifier for Bioelectric Recordings |
title_fullStr |
Effect of the Electrode Impedance in Improved Buffer Amplifier for Bioelectric Recordings |
title_full_unstemmed |
Effect of the Electrode Impedance in Improved Buffer Amplifier for Bioelectric Recordings |
title_sort |
Effect of the Electrode Impedance in Improved Buffer Amplifier for Bioelectric Recordings |
dc.creator.none.fl_str_mv |
Valverde, Esteban Raul Arini, Pedro David Biagetti, Marcelo Bertran, Guillermo Claudio Quinteiro, Ricardo Alberto |
author |
Valverde, Esteban Raul |
author_facet |
Valverde, Esteban Raul Arini, Pedro David Biagetti, Marcelo Bertran, Guillermo Claudio Quinteiro, Ricardo Alberto |
author_role |
author |
author2 |
Arini, Pedro David Biagetti, Marcelo Bertran, Guillermo Claudio Quinteiro, Ricardo Alberto |
author2_role |
author author author author |
dc.subject.none.fl_str_mv |
ELECTRODE IMPEDANCE BUFFER AMPLIFIER BIOELECTRIC RECORDINGS |
topic |
ELECTRODE IMPEDANCE BUFFER AMPLIFIER BIOELECTRIC RECORDINGS |
purl_subject.fl_str_mv |
https://purl.org/becyt/ford/2.2 https://purl.org/becyt/ford/2 |
dc.description.none.fl_txt_mv |
We analyzed the effects of the electrode impedance on the transfer response of a one-stage improved buffer amplifier. The electrode DC resistance (Rd) modifies the one-stage buffer transfer response. We found a limit electrode resistance (Rd(lim)) which depends on the transfer damping factor (∈). If Rd is lower than 86.5 kW, the transfer response of the buffer fulfils the American Heart Association, AHA, recommendations, but when Rd is greater than Rd(lim) it must be cautiously weighted-up because its influence in the transfer response becomes appreciable. The maximum Rd that can be driven by the buffer is 1.2 MW. Higher values never fulfil the AHA recommendations. Therefore, electrodes with higher impedance should not be used with this kind of buffer. On the other hand, when this buffer is used to build-in an instrumentation amplifier (IA) for bipolar recording, the common-mode rejection ratio (CMRR) is sensitive to the electrode type used. Fil: Valverde, Esteban Raul. Consejo Nacional de Investigaciones Científicas y Técnicas; Argentina. Universidad Favaloro; Argentina Fil: Arini, Pedro David. Consejo Nacional de Investigaciones Científicas y Técnicas. Oficina de Coordinación Administrativa Saavedra 15. Instituto Argentino de Matemática Alberto Calderón; Argentina. Universidad Favaloro; Argentina Fil: Biagetti, Marcelo. Universidad Favaloro; Argentina Fil: Bertran, Guillermo Claudio. Universidad Favaloro; Argentina. Consejo Nacional de Investigaciones Científicas y Técnicas; Argentina Fil: Quinteiro, Ricardo Alberto. Universidad Favaloro; Argentina |
description |
We analyzed the effects of the electrode impedance on the transfer response of a one-stage improved buffer amplifier. The electrode DC resistance (Rd) modifies the one-stage buffer transfer response. We found a limit electrode resistance (Rd(lim)) which depends on the transfer damping factor (∈). If Rd is lower than 86.5 kW, the transfer response of the buffer fulfils the American Heart Association, AHA, recommendations, but when Rd is greater than Rd(lim) it must be cautiously weighted-up because its influence in the transfer response becomes appreciable. The maximum Rd that can be driven by the buffer is 1.2 MW. Higher values never fulfil the AHA recommendations. Therefore, electrodes with higher impedance should not be used with this kind of buffer. On the other hand, when this buffer is used to build-in an instrumentation amplifier (IA) for bipolar recording, the common-mode rejection ratio (CMRR) is sensitive to the electrode type used. |
publishDate |
2004 |
dc.date.none.fl_str_mv |
2004-12 |
dc.type.none.fl_str_mv |
info:eu-repo/semantics/article info:eu-repo/semantics/publishedVersion http://purl.org/coar/resource_type/c_6501 info:ar-repo/semantics/articulo |
format |
article |
status_str |
publishedVersion |
dc.identifier.none.fl_str_mv |
http://hdl.handle.net/11336/110162 Valverde, Esteban Raul; Arini, Pedro David; Biagetti, Marcelo; Bertran, Guillermo Claudio; Quinteiro, Ricardo Alberto; Effect of the Electrode Impedance in Improved Buffer Amplifier for Bioelectric Recordings; Taylor & Francis; Journal Of Medical Engineering And Technology; 28; 5; 12-2004; 217-222 0309-1902 1464-522X CONICET Digital CONICET |
url |
http://hdl.handle.net/11336/110162 |
identifier_str_mv |
Valverde, Esteban Raul; Arini, Pedro David; Biagetti, Marcelo; Bertran, Guillermo Claudio; Quinteiro, Ricardo Alberto; Effect of the Electrode Impedance in Improved Buffer Amplifier for Bioelectric Recordings; Taylor & Francis; Journal Of Medical Engineering And Technology; 28; 5; 12-2004; 217-222 0309-1902 1464-522X CONICET Digital CONICET |
dc.language.none.fl_str_mv |
eng |
language |
eng |
dc.relation.none.fl_str_mv |
info:eu-repo/semantics/altIdentifier/url/https://www.tandfonline.com/doi/abs/10.1080/03091900410001662323 info:eu-repo/semantics/altIdentifier/doi/10.1080/03091900410001662323 |
dc.rights.none.fl_str_mv |
info:eu-repo/semantics/openAccess https://creativecommons.org/licenses/by-nc-sa/2.5/ar/ |
eu_rights_str_mv |
openAccess |
rights_invalid_str_mv |
https://creativecommons.org/licenses/by-nc-sa/2.5/ar/ |
dc.format.none.fl_str_mv |
application/pdf application/pdf application/pdf application/pdf |
dc.publisher.none.fl_str_mv |
Taylor & Francis |
publisher.none.fl_str_mv |
Taylor & Francis |
dc.source.none.fl_str_mv |
reponame:CONICET Digital (CONICET) instname:Consejo Nacional de Investigaciones Científicas y Técnicas |
reponame_str |
CONICET Digital (CONICET) |
collection |
CONICET Digital (CONICET) |
instname_str |
Consejo Nacional de Investigaciones Científicas y Técnicas |
repository.name.fl_str_mv |
CONICET Digital (CONICET) - Consejo Nacional de Investigaciones Científicas y Técnicas |
repository.mail.fl_str_mv |
dasensio@conicet.gov.ar; lcarlino@conicet.gov.ar |
_version_ |
1844613341095395328 |
score |
13.070432 |