Development of an impedimetric immunosensor to determine microcystin-LR: New approaches in the use of the electrochemical impedance spectroscopy was used in determining to determin...

Autores
Boffadossi, Micaela; Di Tocco, Aylen; Lassabe, Gabriel; Pirez Schirmer, Macarena; Robledo, Sebastian Noel; Fernandez, Hector; Zon, María Alicia; González Sapienza, Gualberto; Arevalo, Fernando Javier
Año de publicación
2020
Idioma
inglés
Tipo de recurso
artículo
Estado
versión publicada
Descripción
The development of new electrochemical immunosensors for the detection of environmental contaminants is of great interest due to their simplicity, high sensitivity, and extended analytical range. Because of the antibody is immobilized on the electrode, it is important to determine its loss of reactivity after immobilization. In this work, two aspects were addressed. First, we developed a new methodology based on electrochemical impedance (EI) to determine the kinetic parameters associated with immunoreactions carried out on the electrode. Second, an electrochemical immunosensor based on electrochemical impedance spectroscopy (EIE) was developed to determine microcystin-LR in drinking water samples. Microcystin-LR determination was based on a label-free non-competitive immunoassay. The electrochemical immunosensor shows a limit of detection (LOD) of 33 pg mL−1 (3.32 × 10−11 mol L−1 or 0.033 μg L−1) which is well below the WHO guideline recommendation of 1 μg L−1 and 40 times better than the LOD obtained using the same antibody in an optimized conventional competitive ELISA assay. In addition, an acceptable accuracy, with recovery percentages close to 100% were found. The label-free immunosensor is a valuable tool to monitor microcystin-LR in drinking water samples and the environment.
Fil: Boffadossi, Micaela. Universidad Nacional de Río Cuarto. Instituto para el Desarrollo Agroindustrial y de la Salud. - Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - Córdoba. Instituto para el Desarrollo Agroindustrial y de la Salud; Argentina. Universidad Nacional de Río Cuarto. Facultad de Ciencias Exactas Fisicoquímicas y Naturales. Departamento de Química; Argentina
Fil: Di Tocco, Aylen. Universidad Nacional de Río Cuarto. Instituto para el Desarrollo Agroindustrial y de la Salud. - Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - Córdoba. Instituto para el Desarrollo Agroindustrial y de la Salud; Argentina. Universidad Nacional de Río Cuarto. Facultad de Ciencias Exactas Fisicoquímicas y Naturales. Departamento de Química; Argentina
Fil: Lassabe, Gabriel. Universidad de la República; Uruguay
Fil: Pirez Schirmer, Macarena. Universidad de la República; Uruguay
Fil: Robledo, Sebastian Noel. Universidad Nacional de Río Cuarto. Instituto para el Desarrollo Agroindustrial y de la Salud. - Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - Córdoba. Instituto para el Desarrollo Agroindustrial y de la Salud; Argentina. Universidad Nacional de Río Cuarto. Facultad de Ingeniería. Departamento de Tecnología Química; Argentina
Fil: Fernandez, Hector. Universidad Nacional de Río Cuarto. Instituto para el Desarrollo Agroindustrial y de la Salud. - Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - Córdoba. Instituto para el Desarrollo Agroindustrial y de la Salud; Argentina. Universidad Nacional de Río Cuarto. Facultad de Ciencias Exactas Fisicoquímicas y Naturales. Departamento de Química; Argentina
Fil: Zon, María Alicia. Universidad Nacional de Río Cuarto. Instituto para el Desarrollo Agroindustrial y de la Salud. - Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - Córdoba. Instituto para el Desarrollo Agroindustrial y de la Salud; Argentina. Universidad Nacional de Río Cuarto. Facultad de Ciencias Exactas Fisicoquímicas y Naturales. Departamento de Química; Argentina
Fil: González Sapienza, Gualberto. Universidad de la República; Uruguay
Fil: Arevalo, Fernando Javier. Universidad Nacional de Río Cuarto. Instituto para el Desarrollo Agroindustrial y de la Salud. - Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - Córdoba. Instituto para el Desarrollo Agroindustrial y de la Salud; Argentina. Universidad Nacional de Río Cuarto. Facultad de Ciencias Exactas Fisicoquímicas y Naturales. Departamento de Química; Argentina
Materia
DETERMINATION OF AFFINITY CONSTANT BY ELECTROCHEMICAL IMPEDANCE
HETEROGENEOUS IMMUNOASSAY
MICROCYSTIN
Nivel de accesibilidad
acceso abierto
Condiciones de uso
https://creativecommons.org/licenses/by-nc-nd/2.5/ar/
Repositorio
CONICET Digital (CONICET)
Institución
Consejo Nacional de Investigaciones Científicas y Técnicas
OAI Identificador
oai:ri.conicet.gov.ar:11336/141985

id CONICETDig_5bf82e5310fb21fdbc5b45aae753392a
oai_identifier_str oai:ri.conicet.gov.ar:11336/141985
network_acronym_str CONICETDig
repository_id_str 3498
network_name_str CONICET Digital (CONICET)
spelling Development of an impedimetric immunosensor to determine microcystin-LR: New approaches in the use of the electrochemical impedance spectroscopy was used in determining to determine kinetic parameters of immunoreactionsBoffadossi, MicaelaDi Tocco, AylenLassabe, GabrielPirez Schirmer, MacarenaRobledo, Sebastian NoelFernandez, HectorZon, María AliciaGonzález Sapienza, GualbertoArevalo, Fernando JavierDETERMINATION OF AFFINITY CONSTANT BY ELECTROCHEMICAL IMPEDANCEHETEROGENEOUS IMMUNOASSAYMICROCYSTINhttps://purl.org/becyt/ford/1.4https://purl.org/becyt/ford/1The development of new electrochemical immunosensors for the detection of environmental contaminants is of great interest due to their simplicity, high sensitivity, and extended analytical range. Because of the antibody is immobilized on the electrode, it is important to determine its loss of reactivity after immobilization. In this work, two aspects were addressed. First, we developed a new methodology based on electrochemical impedance (EI) to determine the kinetic parameters associated with immunoreactions carried out on the electrode. Second, an electrochemical immunosensor based on electrochemical impedance spectroscopy (EIE) was developed to determine microcystin-LR in drinking water samples. Microcystin-LR determination was based on a label-free non-competitive immunoassay. The electrochemical immunosensor shows a limit of detection (LOD) of 33 pg mL−1 (3.32 × 10−11 mol L−1 or 0.033 μg L−1) which is well below the WHO guideline recommendation of 1 μg L−1 and 40 times better than the LOD obtained using the same antibody in an optimized conventional competitive ELISA assay. In addition, an acceptable accuracy, with recovery percentages close to 100% were found. The label-free immunosensor is a valuable tool to monitor microcystin-LR in drinking water samples and the environment.Fil: Boffadossi, Micaela. Universidad Nacional de Río Cuarto. Instituto para el Desarrollo Agroindustrial y de la Salud. - Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - Córdoba. Instituto para el Desarrollo Agroindustrial y de la Salud; Argentina. Universidad Nacional de Río Cuarto. Facultad de Ciencias Exactas Fisicoquímicas y Naturales. Departamento de Química; ArgentinaFil: Di Tocco, Aylen. Universidad Nacional de Río Cuarto. Instituto para el Desarrollo Agroindustrial y de la Salud. - Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - Córdoba. Instituto para el Desarrollo Agroindustrial y de la Salud; Argentina. Universidad Nacional de Río Cuarto. Facultad de Ciencias Exactas Fisicoquímicas y Naturales. Departamento de Química; ArgentinaFil: Lassabe, Gabriel. Universidad de la República; UruguayFil: Pirez Schirmer, Macarena. Universidad de la República; UruguayFil: Robledo, Sebastian Noel. Universidad Nacional de Río Cuarto. Instituto para el Desarrollo Agroindustrial y de la Salud. - Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - Córdoba. Instituto para el Desarrollo Agroindustrial y de la Salud; Argentina. Universidad Nacional de Río Cuarto. Facultad de Ingeniería. Departamento de Tecnología Química; ArgentinaFil: Fernandez, Hector. Universidad Nacional de Río Cuarto. Instituto para el Desarrollo Agroindustrial y de la Salud. - Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - Córdoba. Instituto para el Desarrollo Agroindustrial y de la Salud; Argentina. Universidad Nacional de Río Cuarto. Facultad de Ciencias Exactas Fisicoquímicas y Naturales. Departamento de Química; ArgentinaFil: Zon, María Alicia. Universidad Nacional de Río Cuarto. Instituto para el Desarrollo Agroindustrial y de la Salud. - Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - Córdoba. Instituto para el Desarrollo Agroindustrial y de la Salud; Argentina. Universidad Nacional de Río Cuarto. Facultad de Ciencias Exactas Fisicoquímicas y Naturales. Departamento de Química; ArgentinaFil: González Sapienza, Gualberto. Universidad de la República; UruguayFil: Arevalo, Fernando Javier. Universidad Nacional de Río Cuarto. Instituto para el Desarrollo Agroindustrial y de la Salud. - Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - Córdoba. Instituto para el Desarrollo Agroindustrial y de la Salud; Argentina. Universidad Nacional de Río Cuarto. Facultad de Ciencias Exactas Fisicoquímicas y Naturales. Departamento de Química; ArgentinaPergamon-Elsevier Science Ltd2020-09info:eu-repo/semantics/articleinfo:eu-repo/semantics/publishedVersionhttp://purl.org/coar/resource_type/c_6501info:ar-repo/semantics/articuloapplication/pdfapplication/pdfapplication/pdfapplication/pdfapplication/pdfhttp://hdl.handle.net/11336/141985Boffadossi, Micaela; Di Tocco, Aylen; Lassabe, Gabriel; Pirez Schirmer, Macarena; Robledo, Sebastian Noel; et al.; Development of an impedimetric immunosensor to determine microcystin-LR: New approaches in the use of the electrochemical impedance spectroscopy was used in determining to determine kinetic parameters of immunoreactions; Pergamon-Elsevier Science Ltd; Electrochimica Acta; 353; 9-2020; 1-8; 1366210013-4686CONICET DigitalCONICETenginfo:eu-repo/semantics/altIdentifier/url/https://www.sciencedirect.com/science/article/abs/pii/S0013468620310148info:eu-repo/semantics/altIdentifier/doi/10.1016/j.electacta.2020.136621info:eu-repo/semantics/openAccesshttps://creativecommons.org/licenses/by-nc-nd/2.5/ar/reponame:CONICET Digital (CONICET)instname:Consejo Nacional de Investigaciones Científicas y Técnicas2025-09-29T10:03:42Zoai:ri.conicet.gov.ar:11336/141985instacron:CONICETInstitucionalhttp://ri.conicet.gov.ar/Organismo científico-tecnológicoNo correspondehttp://ri.conicet.gov.ar/oai/requestdasensio@conicet.gov.ar; lcarlino@conicet.gov.arArgentinaNo correspondeNo correspondeNo correspondeopendoar:34982025-09-29 10:03:42.47CONICET Digital (CONICET) - Consejo Nacional de Investigaciones Científicas y Técnicasfalse
dc.title.none.fl_str_mv Development of an impedimetric immunosensor to determine microcystin-LR: New approaches in the use of the electrochemical impedance spectroscopy was used in determining to determine kinetic parameters of immunoreactions
title Development of an impedimetric immunosensor to determine microcystin-LR: New approaches in the use of the electrochemical impedance spectroscopy was used in determining to determine kinetic parameters of immunoreactions
spellingShingle Development of an impedimetric immunosensor to determine microcystin-LR: New approaches in the use of the electrochemical impedance spectroscopy was used in determining to determine kinetic parameters of immunoreactions
Boffadossi, Micaela
DETERMINATION OF AFFINITY CONSTANT BY ELECTROCHEMICAL IMPEDANCE
HETEROGENEOUS IMMUNOASSAY
MICROCYSTIN
title_short Development of an impedimetric immunosensor to determine microcystin-LR: New approaches in the use of the electrochemical impedance spectroscopy was used in determining to determine kinetic parameters of immunoreactions
title_full Development of an impedimetric immunosensor to determine microcystin-LR: New approaches in the use of the electrochemical impedance spectroscopy was used in determining to determine kinetic parameters of immunoreactions
title_fullStr Development of an impedimetric immunosensor to determine microcystin-LR: New approaches in the use of the electrochemical impedance spectroscopy was used in determining to determine kinetic parameters of immunoreactions
title_full_unstemmed Development of an impedimetric immunosensor to determine microcystin-LR: New approaches in the use of the electrochemical impedance spectroscopy was used in determining to determine kinetic parameters of immunoreactions
title_sort Development of an impedimetric immunosensor to determine microcystin-LR: New approaches in the use of the electrochemical impedance spectroscopy was used in determining to determine kinetic parameters of immunoreactions
dc.creator.none.fl_str_mv Boffadossi, Micaela
Di Tocco, Aylen
Lassabe, Gabriel
Pirez Schirmer, Macarena
Robledo, Sebastian Noel
Fernandez, Hector
Zon, María Alicia
González Sapienza, Gualberto
Arevalo, Fernando Javier
author Boffadossi, Micaela
author_facet Boffadossi, Micaela
Di Tocco, Aylen
Lassabe, Gabriel
Pirez Schirmer, Macarena
Robledo, Sebastian Noel
Fernandez, Hector
Zon, María Alicia
González Sapienza, Gualberto
Arevalo, Fernando Javier
author_role author
author2 Di Tocco, Aylen
Lassabe, Gabriel
Pirez Schirmer, Macarena
Robledo, Sebastian Noel
Fernandez, Hector
Zon, María Alicia
González Sapienza, Gualberto
Arevalo, Fernando Javier
author2_role author
author
author
author
author
author
author
author
dc.subject.none.fl_str_mv DETERMINATION OF AFFINITY CONSTANT BY ELECTROCHEMICAL IMPEDANCE
HETEROGENEOUS IMMUNOASSAY
MICROCYSTIN
topic DETERMINATION OF AFFINITY CONSTANT BY ELECTROCHEMICAL IMPEDANCE
HETEROGENEOUS IMMUNOASSAY
MICROCYSTIN
purl_subject.fl_str_mv https://purl.org/becyt/ford/1.4
https://purl.org/becyt/ford/1
dc.description.none.fl_txt_mv The development of new electrochemical immunosensors for the detection of environmental contaminants is of great interest due to their simplicity, high sensitivity, and extended analytical range. Because of the antibody is immobilized on the electrode, it is important to determine its loss of reactivity after immobilization. In this work, two aspects were addressed. First, we developed a new methodology based on electrochemical impedance (EI) to determine the kinetic parameters associated with immunoreactions carried out on the electrode. Second, an electrochemical immunosensor based on electrochemical impedance spectroscopy (EIE) was developed to determine microcystin-LR in drinking water samples. Microcystin-LR determination was based on a label-free non-competitive immunoassay. The electrochemical immunosensor shows a limit of detection (LOD) of 33 pg mL−1 (3.32 × 10−11 mol L−1 or 0.033 μg L−1) which is well below the WHO guideline recommendation of 1 μg L−1 and 40 times better than the LOD obtained using the same antibody in an optimized conventional competitive ELISA assay. In addition, an acceptable accuracy, with recovery percentages close to 100% were found. The label-free immunosensor is a valuable tool to monitor microcystin-LR in drinking water samples and the environment.
Fil: Boffadossi, Micaela. Universidad Nacional de Río Cuarto. Instituto para el Desarrollo Agroindustrial y de la Salud. - Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - Córdoba. Instituto para el Desarrollo Agroindustrial y de la Salud; Argentina. Universidad Nacional de Río Cuarto. Facultad de Ciencias Exactas Fisicoquímicas y Naturales. Departamento de Química; Argentina
Fil: Di Tocco, Aylen. Universidad Nacional de Río Cuarto. Instituto para el Desarrollo Agroindustrial y de la Salud. - Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - Córdoba. Instituto para el Desarrollo Agroindustrial y de la Salud; Argentina. Universidad Nacional de Río Cuarto. Facultad de Ciencias Exactas Fisicoquímicas y Naturales. Departamento de Química; Argentina
Fil: Lassabe, Gabriel. Universidad de la República; Uruguay
Fil: Pirez Schirmer, Macarena. Universidad de la República; Uruguay
Fil: Robledo, Sebastian Noel. Universidad Nacional de Río Cuarto. Instituto para el Desarrollo Agroindustrial y de la Salud. - Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - Córdoba. Instituto para el Desarrollo Agroindustrial y de la Salud; Argentina. Universidad Nacional de Río Cuarto. Facultad de Ingeniería. Departamento de Tecnología Química; Argentina
Fil: Fernandez, Hector. Universidad Nacional de Río Cuarto. Instituto para el Desarrollo Agroindustrial y de la Salud. - Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - Córdoba. Instituto para el Desarrollo Agroindustrial y de la Salud; Argentina. Universidad Nacional de Río Cuarto. Facultad de Ciencias Exactas Fisicoquímicas y Naturales. Departamento de Química; Argentina
Fil: Zon, María Alicia. Universidad Nacional de Río Cuarto. Instituto para el Desarrollo Agroindustrial y de la Salud. - Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - Córdoba. Instituto para el Desarrollo Agroindustrial y de la Salud; Argentina. Universidad Nacional de Río Cuarto. Facultad de Ciencias Exactas Fisicoquímicas y Naturales. Departamento de Química; Argentina
Fil: González Sapienza, Gualberto. Universidad de la República; Uruguay
Fil: Arevalo, Fernando Javier. Universidad Nacional de Río Cuarto. Instituto para el Desarrollo Agroindustrial y de la Salud. - Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - Córdoba. Instituto para el Desarrollo Agroindustrial y de la Salud; Argentina. Universidad Nacional de Río Cuarto. Facultad de Ciencias Exactas Fisicoquímicas y Naturales. Departamento de Química; Argentina
description The development of new electrochemical immunosensors for the detection of environmental contaminants is of great interest due to their simplicity, high sensitivity, and extended analytical range. Because of the antibody is immobilized on the electrode, it is important to determine its loss of reactivity after immobilization. In this work, two aspects were addressed. First, we developed a new methodology based on electrochemical impedance (EI) to determine the kinetic parameters associated with immunoreactions carried out on the electrode. Second, an electrochemical immunosensor based on electrochemical impedance spectroscopy (EIE) was developed to determine microcystin-LR in drinking water samples. Microcystin-LR determination was based on a label-free non-competitive immunoassay. The electrochemical immunosensor shows a limit of detection (LOD) of 33 pg mL−1 (3.32 × 10−11 mol L−1 or 0.033 μg L−1) which is well below the WHO guideline recommendation of 1 μg L−1 and 40 times better than the LOD obtained using the same antibody in an optimized conventional competitive ELISA assay. In addition, an acceptable accuracy, with recovery percentages close to 100% were found. The label-free immunosensor is a valuable tool to monitor microcystin-LR in drinking water samples and the environment.
publishDate 2020
dc.date.none.fl_str_mv 2020-09
dc.type.none.fl_str_mv info:eu-repo/semantics/article
info:eu-repo/semantics/publishedVersion
http://purl.org/coar/resource_type/c_6501
info:ar-repo/semantics/articulo
format article
status_str publishedVersion
dc.identifier.none.fl_str_mv http://hdl.handle.net/11336/141985
Boffadossi, Micaela; Di Tocco, Aylen; Lassabe, Gabriel; Pirez Schirmer, Macarena; Robledo, Sebastian Noel; et al.; Development of an impedimetric immunosensor to determine microcystin-LR: New approaches in the use of the electrochemical impedance spectroscopy was used in determining to determine kinetic parameters of immunoreactions; Pergamon-Elsevier Science Ltd; Electrochimica Acta; 353; 9-2020; 1-8; 136621
0013-4686
CONICET Digital
CONICET
url http://hdl.handle.net/11336/141985
identifier_str_mv Boffadossi, Micaela; Di Tocco, Aylen; Lassabe, Gabriel; Pirez Schirmer, Macarena; Robledo, Sebastian Noel; et al.; Development of an impedimetric immunosensor to determine microcystin-LR: New approaches in the use of the electrochemical impedance spectroscopy was used in determining to determine kinetic parameters of immunoreactions; Pergamon-Elsevier Science Ltd; Electrochimica Acta; 353; 9-2020; 1-8; 136621
0013-4686
CONICET Digital
CONICET
dc.language.none.fl_str_mv eng
language eng
dc.relation.none.fl_str_mv info:eu-repo/semantics/altIdentifier/url/https://www.sciencedirect.com/science/article/abs/pii/S0013468620310148
info:eu-repo/semantics/altIdentifier/doi/10.1016/j.electacta.2020.136621
dc.rights.none.fl_str_mv info:eu-repo/semantics/openAccess
https://creativecommons.org/licenses/by-nc-nd/2.5/ar/
eu_rights_str_mv openAccess
rights_invalid_str_mv https://creativecommons.org/licenses/by-nc-nd/2.5/ar/
dc.format.none.fl_str_mv application/pdf
application/pdf
application/pdf
application/pdf
application/pdf
dc.publisher.none.fl_str_mv Pergamon-Elsevier Science Ltd
publisher.none.fl_str_mv Pergamon-Elsevier Science Ltd
dc.source.none.fl_str_mv reponame:CONICET Digital (CONICET)
instname:Consejo Nacional de Investigaciones Científicas y Técnicas
reponame_str CONICET Digital (CONICET)
collection CONICET Digital (CONICET)
instname_str Consejo Nacional de Investigaciones Científicas y Técnicas
repository.name.fl_str_mv CONICET Digital (CONICET) - Consejo Nacional de Investigaciones Científicas y Técnicas
repository.mail.fl_str_mv dasensio@conicet.gov.ar; lcarlino@conicet.gov.ar
_version_ 1844613855929434112
score 13.070432