Diseño y fabricación de maniquí prototipo antropomórfico para controles de calidad en tratamientos de Braquiterapia Ginecológica HDR 3D

Autores
Levenzon Levy, Nicolas; Fortuna, Gastón
Año de publicación
2023
Idioma
español castellano
Tipo de recurso
tesis de grado
Estado
versión publicada
Colaborador/a o director/a de tesis
Stefania, Maturana
Martinez, Ariel
Descripción
Proyecto Integrador (I.Biom.)--FCEFN-UNC, 2023
Fil: Levenzon Levy, Nicolas. Universidad Nacional de Córdoba. Facultad de Ciencias Exactas, Físicas y Naturales; Argentina.
Fil: Fortuna, Gastón. Universidad Nacional de Córdoba. Facultad de Ciencias Exactas, Físicas y Naturales; Argentina.
La Braquiterapia 3D HDR (high doce rate) es una técnica de irradiación localizada de alta precisión con mínimos efectos adversos, que mediante el uso de una fuente radioactiva milimétrica permite tratar lesiones en distintas localizaciones con una mayor cantidad de radiación emitida por unidad de tiempo utilizando un sistema de imágenes en tres dimensiones y técnicas de planificación avanzadas. El uso de imágenes para la planificación de tratamientos ha evolucionado de 2D a 3D, incluyendo tomografía computada y resonancia magnética. El Hospital Oncológico de Córdoba, realizó tratamientos de Braquiterapia intracavitaria ginecológica HDR durante seis años consecutivos con fuente de 192Ir y basándose en planificación bidimensional (2D) mediante el uso de películas de rayos X ortogonales. Con el objetivo de avanzar en tecnología y reducir costos, la institución adquirió un nuevo equipo de Braquiterapia para ser utilizado con fuente de 60Co y aplicadores de materiales tales que posibilitaron la adquisición de imágenes por Tomografía Computada y Resonancia Magnética con el fin de introducir un método de planificación tridimensional (3D). Entre los diversos controles que se deben realizar a las partes involucradas, como puede ser la dosis, tiempos de exposición, planificación, etc., en la braquiterapia, el más importante es el que involucra todo el proceso de principio a fin. Para poder llevar a cabo este trabajo se requiere de un maniquí o “phantom” lo más similar posible a un paciente para reproducir las condiciones exactas. En nuestro país no hay desarrollo comercial de maniquí específico para braquiterapia y las empresas que ofrecen dichos dispositivos son extranjeras. En el presente trabajo se desarrolló el proceso de diseño, fabricación y prueba de un maniquí antropomórfico para el control de calidad de principio a fin (End to end, E2E) de un tratamiento completo de braquiterapia ginecológica HDR 3D. Una vez corroborado el buen funcionamiento del dispositivo, con el fin de asegurar una correcta entrega del tratamiento, se realizaron conjuntamente mediciones de QA (quality assurance) sobre la unidad HDR, la fuente radiactiva y su conjunto dosimétrico, el equipo de adquisición de imágenes, el Sistema de Planificación de Tratamientos (TPS), y, por último, sobre el procedimiento en general. Además, se implementó una nueva técnica de prueba dosimétrica mediante placas radio-crómicas EBT3 (nombre de modelo de las placas radiosensibles) que resultaron muy satisfactorias e implica un nuevo avance en control de calidad en lo que respecta al área pública del país. Así mismo permitió realizar una comparación con resultados de mediciones anteriores arrojando resultados muy favorables y de alto impacto para la nueva técnica. La incorporación de este nuevo dispositivo y las pruebas realizadas con él permitieron actualizar el Programa de Garantía de Calidad de la institución, y establecer una base para poder implementarlo en otros servicios de Braquiterapia ginecológica HDR guiada por imágenes 3D del país. Además, cumple la función de capacitación de residentes y el perfeccionamiento de las técnicas de braquiterapia ginecológica intracavitaria.
3D HDR (High dose rate) brachytherapy is a technique of localized high precision irradiation with minimal adverse effects, which, by using a millimetric radioactive source, allows treating lesions in different locations with a greater amount of radiation emitted per unit of time using a system of three-dimensional images and advanced planning techniques. The use of images for treatment planning has evolved from 2D to 3D, including computed tomography and magnetic resonance imaging. The Oncological Hospital of Córdoba carried out intracavitary gynecological HDR brachytherapy treatments for six consecutive years with a 192Ir source based on two-dimensional (2D) planning using orthogonal X-ray films. With the aim of advancing technology and reducing costs, the institution acquired a new brachytherapy equipment to be used with a 60Co source and applicators of such materials that enabled the acquisition of images by computed tomography and magnetic resonance imaging in order to introduce a three-dimensional (3D) planning method. Among the various controls that must be carried out on the involved parts, such as dose, exposure times, planning, etc., in brachytherapy, the most important is the one that involves the entire process from start to finish. To carry out this work, a mannequin or "phantom" is required that is as similar as possible to a patient to reproduce the exact conditions. In our country, there is no commercial development of a specific mannequin for brachytherapy, and the companies that offer such devices are foreign. In this work, the process of designing, manufacturing, and testing an anthropomorphic dummy for end-to-end (E2E) quality control of a complete 3D HDR gynecological brachytherapy treatment was developed. Once the proper functioning of the device was corroborated, in order to ensure a correct delivery of the treatment, QA (quality assurance) measurements were jointly performed on the HDR unit, the radioactive source and its dosimetric assembly, the imaging acquisition equipment, the Treatment Planning System (TPS), and finally on the procedure in general. In addition, a new dosimetric test technique was implemented using EBT3 radiochromic plates that were very satisfactory and represent a new advance in quality control in the public area of the country. It also allowed a comparison with previous measurement results, yielding very favorable and high-impact results for the new technique. The incorporation of this new device and the tests carried out with it allowed updating the institution's Quality Assurance Program and establishing a basis for implementing it in other services of 3D image-guided gynecological HDR brachytherapy in the country. Additionally, it serves as training for residents and the improvement of intracavitary gynecological brachytherapy techniques.
Fil: Levenzon Levy, Nicolas. Universidad Nacional de Córdoba. Facultad de Ciencias Exactas, Físicas y Naturales; Argentina.
Fil: Fortuna, Gastón. Universidad Nacional de Córdoba. Facultad de Ciencias Exactas, Físicas y Naturales; Argentina.
Materia
Proyecto Integrador (I.Biom.)
Braquiterapia HDR 3D
Dosimetría
Control de calidad
Maniquí antropomórfico
Test End to End
Tomografía computada
Planificación de tratamiento
Nivel de accesibilidad
acceso abierto
Condiciones de uso
Repositorio
Repositorio Digital Universitario (UNC)
Institución
Universidad Nacional de Córdoba
OAI Identificador
oai:rdu.unc.edu.ar:11086/552510

id RDUUNC_9cfebef1ef16a1f705a01298108b1350
oai_identifier_str oai:rdu.unc.edu.ar:11086/552510
network_acronym_str RDUUNC
repository_id_str 2572
network_name_str Repositorio Digital Universitario (UNC)
spelling Diseño y fabricación de maniquí prototipo antropomórfico para controles de calidad en tratamientos de Braquiterapia Ginecológica HDR 3DLevenzon Levy, NicolasFortuna, GastónProyecto Integrador (I.Biom.)Braquiterapia HDR 3DDosimetríaControl de calidadManiquí antropomórficoTest End to EndTomografía computadaPlanificación de tratamientoProyecto Integrador (I.Biom.)--FCEFN-UNC, 2023Fil: Levenzon Levy, Nicolas. Universidad Nacional de Córdoba. Facultad de Ciencias Exactas, Físicas y Naturales; Argentina.Fil: Fortuna, Gastón. Universidad Nacional de Córdoba. Facultad de Ciencias Exactas, Físicas y Naturales; Argentina.La Braquiterapia 3D HDR (high doce rate) es una técnica de irradiación localizada de alta precisión con mínimos efectos adversos, que mediante el uso de una fuente radioactiva milimétrica permite tratar lesiones en distintas localizaciones con una mayor cantidad de radiación emitida por unidad de tiempo utilizando un sistema de imágenes en tres dimensiones y técnicas de planificación avanzadas. El uso de imágenes para la planificación de tratamientos ha evolucionado de 2D a 3D, incluyendo tomografía computada y resonancia magnética. El Hospital Oncológico de Córdoba, realizó tratamientos de Braquiterapia intracavitaria ginecológica HDR durante seis años consecutivos con fuente de 192Ir y basándose en planificación bidimensional (2D) mediante el uso de películas de rayos X ortogonales. Con el objetivo de avanzar en tecnología y reducir costos, la institución adquirió un nuevo equipo de Braquiterapia para ser utilizado con fuente de 60Co y aplicadores de materiales tales que posibilitaron la adquisición de imágenes por Tomografía Computada y Resonancia Magnética con el fin de introducir un método de planificación tridimensional (3D). Entre los diversos controles que se deben realizar a las partes involucradas, como puede ser la dosis, tiempos de exposición, planificación, etc., en la braquiterapia, el más importante es el que involucra todo el proceso de principio a fin. Para poder llevar a cabo este trabajo se requiere de un maniquí o “phantom” lo más similar posible a un paciente para reproducir las condiciones exactas. En nuestro país no hay desarrollo comercial de maniquí específico para braquiterapia y las empresas que ofrecen dichos dispositivos son extranjeras. En el presente trabajo se desarrolló el proceso de diseño, fabricación y prueba de un maniquí antropomórfico para el control de calidad de principio a fin (End to end, E2E) de un tratamiento completo de braquiterapia ginecológica HDR 3D. Una vez corroborado el buen funcionamiento del dispositivo, con el fin de asegurar una correcta entrega del tratamiento, se realizaron conjuntamente mediciones de QA (quality assurance) sobre la unidad HDR, la fuente radiactiva y su conjunto dosimétrico, el equipo de adquisición de imágenes, el Sistema de Planificación de Tratamientos (TPS), y, por último, sobre el procedimiento en general. Además, se implementó una nueva técnica de prueba dosimétrica mediante placas radio-crómicas EBT3 (nombre de modelo de las placas radiosensibles) que resultaron muy satisfactorias e implica un nuevo avance en control de calidad en lo que respecta al área pública del país. Así mismo permitió realizar una comparación con resultados de mediciones anteriores arrojando resultados muy favorables y de alto impacto para la nueva técnica. La incorporación de este nuevo dispositivo y las pruebas realizadas con él permitieron actualizar el Programa de Garantía de Calidad de la institución, y establecer una base para poder implementarlo en otros servicios de Braquiterapia ginecológica HDR guiada por imágenes 3D del país. Además, cumple la función de capacitación de residentes y el perfeccionamiento de las técnicas de braquiterapia ginecológica intracavitaria.3D HDR (High dose rate) brachytherapy is a technique of localized high precision irradiation with minimal adverse effects, which, by using a millimetric radioactive source, allows treating lesions in different locations with a greater amount of radiation emitted per unit of time using a system of three-dimensional images and advanced planning techniques. The use of images for treatment planning has evolved from 2D to 3D, including computed tomography and magnetic resonance imaging. The Oncological Hospital of Córdoba carried out intracavitary gynecological HDR brachytherapy treatments for six consecutive years with a 192Ir source based on two-dimensional (2D) planning using orthogonal X-ray films. With the aim of advancing technology and reducing costs, the institution acquired a new brachytherapy equipment to be used with a 60Co source and applicators of such materials that enabled the acquisition of images by computed tomography and magnetic resonance imaging in order to introduce a three-dimensional (3D) planning method. Among the various controls that must be carried out on the involved parts, such as dose, exposure times, planning, etc., in brachytherapy, the most important is the one that involves the entire process from start to finish. To carry out this work, a mannequin or "phantom" is required that is as similar as possible to a patient to reproduce the exact conditions. In our country, there is no commercial development of a specific mannequin for brachytherapy, and the companies that offer such devices are foreign. In this work, the process of designing, manufacturing, and testing an anthropomorphic dummy for end-to-end (E2E) quality control of a complete 3D HDR gynecological brachytherapy treatment was developed. Once the proper functioning of the device was corroborated, in order to ensure a correct delivery of the treatment, QA (quality assurance) measurements were jointly performed on the HDR unit, the radioactive source and its dosimetric assembly, the imaging acquisition equipment, the Treatment Planning System (TPS), and finally on the procedure in general. In addition, a new dosimetric test technique was implemented using EBT3 radiochromic plates that were very satisfactory and represent a new advance in quality control in the public area of the country. It also allowed a comparison with previous measurement results, yielding very favorable and high-impact results for the new technique. The incorporation of this new device and the tests carried out with it allowed updating the institution's Quality Assurance Program and establishing a basis for implementing it in other services of 3D image-guided gynecological HDR brachytherapy in the country. Additionally, it serves as training for residents and the improvement of intracavitary gynecological brachytherapy techniques.Fil: Levenzon Levy, Nicolas. Universidad Nacional de Córdoba. Facultad de Ciencias Exactas, Físicas y Naturales; Argentina.Fil: Fortuna, Gastón. Universidad Nacional de Córdoba. Facultad de Ciencias Exactas, Físicas y Naturales; Argentina.Stefania, MaturanaMartinez, Ariel2023info:eu-repo/semantics/bachelorThesisinfo:eu-repo/semantics/publishedVersionhttp://purl.org/coar/resource_type/c_7a1finfo:ar-repo/semantics/tesisDeGradoapplication/pdfhttp://hdl.handle.net/11086/552510spainfo:eu-repo/semantics/openAccessreponame:Repositorio Digital Universitario (UNC)instname:Universidad Nacional de Córdobainstacron:UNC2025-09-29T13:43:19Zoai:rdu.unc.edu.ar:11086/552510Institucionalhttps://rdu.unc.edu.ar/Universidad públicaNo correspondehttp://rdu.unc.edu.ar/oai/snrdoca.unc@gmail.comArgentinaNo correspondeNo correspondeNo correspondeopendoar:25722025-09-29 13:43:19.451Repositorio Digital Universitario (UNC) - Universidad Nacional de Córdobafalse
dc.title.none.fl_str_mv Diseño y fabricación de maniquí prototipo antropomórfico para controles de calidad en tratamientos de Braquiterapia Ginecológica HDR 3D
title Diseño y fabricación de maniquí prototipo antropomórfico para controles de calidad en tratamientos de Braquiterapia Ginecológica HDR 3D
spellingShingle Diseño y fabricación de maniquí prototipo antropomórfico para controles de calidad en tratamientos de Braquiterapia Ginecológica HDR 3D
Levenzon Levy, Nicolas
Proyecto Integrador (I.Biom.)
Braquiterapia HDR 3D
Dosimetría
Control de calidad
Maniquí antropomórfico
Test End to End
Tomografía computada
Planificación de tratamiento
title_short Diseño y fabricación de maniquí prototipo antropomórfico para controles de calidad en tratamientos de Braquiterapia Ginecológica HDR 3D
title_full Diseño y fabricación de maniquí prototipo antropomórfico para controles de calidad en tratamientos de Braquiterapia Ginecológica HDR 3D
title_fullStr Diseño y fabricación de maniquí prototipo antropomórfico para controles de calidad en tratamientos de Braquiterapia Ginecológica HDR 3D
title_full_unstemmed Diseño y fabricación de maniquí prototipo antropomórfico para controles de calidad en tratamientos de Braquiterapia Ginecológica HDR 3D
title_sort Diseño y fabricación de maniquí prototipo antropomórfico para controles de calidad en tratamientos de Braquiterapia Ginecológica HDR 3D
dc.creator.none.fl_str_mv Levenzon Levy, Nicolas
Fortuna, Gastón
author Levenzon Levy, Nicolas
author_facet Levenzon Levy, Nicolas
Fortuna, Gastón
author_role author
author2 Fortuna, Gastón
author2_role author
dc.contributor.none.fl_str_mv Stefania, Maturana
Martinez, Ariel
dc.subject.none.fl_str_mv Proyecto Integrador (I.Biom.)
Braquiterapia HDR 3D
Dosimetría
Control de calidad
Maniquí antropomórfico
Test End to End
Tomografía computada
Planificación de tratamiento
topic Proyecto Integrador (I.Biom.)
Braquiterapia HDR 3D
Dosimetría
Control de calidad
Maniquí antropomórfico
Test End to End
Tomografía computada
Planificación de tratamiento
dc.description.none.fl_txt_mv Proyecto Integrador (I.Biom.)--FCEFN-UNC, 2023
Fil: Levenzon Levy, Nicolas. Universidad Nacional de Córdoba. Facultad de Ciencias Exactas, Físicas y Naturales; Argentina.
Fil: Fortuna, Gastón. Universidad Nacional de Córdoba. Facultad de Ciencias Exactas, Físicas y Naturales; Argentina.
La Braquiterapia 3D HDR (high doce rate) es una técnica de irradiación localizada de alta precisión con mínimos efectos adversos, que mediante el uso de una fuente radioactiva milimétrica permite tratar lesiones en distintas localizaciones con una mayor cantidad de radiación emitida por unidad de tiempo utilizando un sistema de imágenes en tres dimensiones y técnicas de planificación avanzadas. El uso de imágenes para la planificación de tratamientos ha evolucionado de 2D a 3D, incluyendo tomografía computada y resonancia magnética. El Hospital Oncológico de Córdoba, realizó tratamientos de Braquiterapia intracavitaria ginecológica HDR durante seis años consecutivos con fuente de 192Ir y basándose en planificación bidimensional (2D) mediante el uso de películas de rayos X ortogonales. Con el objetivo de avanzar en tecnología y reducir costos, la institución adquirió un nuevo equipo de Braquiterapia para ser utilizado con fuente de 60Co y aplicadores de materiales tales que posibilitaron la adquisición de imágenes por Tomografía Computada y Resonancia Magnética con el fin de introducir un método de planificación tridimensional (3D). Entre los diversos controles que se deben realizar a las partes involucradas, como puede ser la dosis, tiempos de exposición, planificación, etc., en la braquiterapia, el más importante es el que involucra todo el proceso de principio a fin. Para poder llevar a cabo este trabajo se requiere de un maniquí o “phantom” lo más similar posible a un paciente para reproducir las condiciones exactas. En nuestro país no hay desarrollo comercial de maniquí específico para braquiterapia y las empresas que ofrecen dichos dispositivos son extranjeras. En el presente trabajo se desarrolló el proceso de diseño, fabricación y prueba de un maniquí antropomórfico para el control de calidad de principio a fin (End to end, E2E) de un tratamiento completo de braquiterapia ginecológica HDR 3D. Una vez corroborado el buen funcionamiento del dispositivo, con el fin de asegurar una correcta entrega del tratamiento, se realizaron conjuntamente mediciones de QA (quality assurance) sobre la unidad HDR, la fuente radiactiva y su conjunto dosimétrico, el equipo de adquisición de imágenes, el Sistema de Planificación de Tratamientos (TPS), y, por último, sobre el procedimiento en general. Además, se implementó una nueva técnica de prueba dosimétrica mediante placas radio-crómicas EBT3 (nombre de modelo de las placas radiosensibles) que resultaron muy satisfactorias e implica un nuevo avance en control de calidad en lo que respecta al área pública del país. Así mismo permitió realizar una comparación con resultados de mediciones anteriores arrojando resultados muy favorables y de alto impacto para la nueva técnica. La incorporación de este nuevo dispositivo y las pruebas realizadas con él permitieron actualizar el Programa de Garantía de Calidad de la institución, y establecer una base para poder implementarlo en otros servicios de Braquiterapia ginecológica HDR guiada por imágenes 3D del país. Además, cumple la función de capacitación de residentes y el perfeccionamiento de las técnicas de braquiterapia ginecológica intracavitaria.
3D HDR (High dose rate) brachytherapy is a technique of localized high precision irradiation with minimal adverse effects, which, by using a millimetric radioactive source, allows treating lesions in different locations with a greater amount of radiation emitted per unit of time using a system of three-dimensional images and advanced planning techniques. The use of images for treatment planning has evolved from 2D to 3D, including computed tomography and magnetic resonance imaging. The Oncological Hospital of Córdoba carried out intracavitary gynecological HDR brachytherapy treatments for six consecutive years with a 192Ir source based on two-dimensional (2D) planning using orthogonal X-ray films. With the aim of advancing technology and reducing costs, the institution acquired a new brachytherapy equipment to be used with a 60Co source and applicators of such materials that enabled the acquisition of images by computed tomography and magnetic resonance imaging in order to introduce a three-dimensional (3D) planning method. Among the various controls that must be carried out on the involved parts, such as dose, exposure times, planning, etc., in brachytherapy, the most important is the one that involves the entire process from start to finish. To carry out this work, a mannequin or "phantom" is required that is as similar as possible to a patient to reproduce the exact conditions. In our country, there is no commercial development of a specific mannequin for brachytherapy, and the companies that offer such devices are foreign. In this work, the process of designing, manufacturing, and testing an anthropomorphic dummy for end-to-end (E2E) quality control of a complete 3D HDR gynecological brachytherapy treatment was developed. Once the proper functioning of the device was corroborated, in order to ensure a correct delivery of the treatment, QA (quality assurance) measurements were jointly performed on the HDR unit, the radioactive source and its dosimetric assembly, the imaging acquisition equipment, the Treatment Planning System (TPS), and finally on the procedure in general. In addition, a new dosimetric test technique was implemented using EBT3 radiochromic plates that were very satisfactory and represent a new advance in quality control in the public area of the country. It also allowed a comparison with previous measurement results, yielding very favorable and high-impact results for the new technique. The incorporation of this new device and the tests carried out with it allowed updating the institution's Quality Assurance Program and establishing a basis for implementing it in other services of 3D image-guided gynecological HDR brachytherapy in the country. Additionally, it serves as training for residents and the improvement of intracavitary gynecological brachytherapy techniques.
Fil: Levenzon Levy, Nicolas. Universidad Nacional de Córdoba. Facultad de Ciencias Exactas, Físicas y Naturales; Argentina.
Fil: Fortuna, Gastón. Universidad Nacional de Córdoba. Facultad de Ciencias Exactas, Físicas y Naturales; Argentina.
description Proyecto Integrador (I.Biom.)--FCEFN-UNC, 2023
publishDate 2023
dc.date.none.fl_str_mv 2023
dc.type.none.fl_str_mv info:eu-repo/semantics/bachelorThesis
info:eu-repo/semantics/publishedVersion
http://purl.org/coar/resource_type/c_7a1f
info:ar-repo/semantics/tesisDeGrado
format bachelorThesis
status_str publishedVersion
dc.identifier.none.fl_str_mv http://hdl.handle.net/11086/552510
url http://hdl.handle.net/11086/552510
dc.language.none.fl_str_mv spa
language spa
dc.rights.none.fl_str_mv info:eu-repo/semantics/openAccess
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv application/pdf
dc.source.none.fl_str_mv reponame:Repositorio Digital Universitario (UNC)
instname:Universidad Nacional de Córdoba
instacron:UNC
reponame_str Repositorio Digital Universitario (UNC)
collection Repositorio Digital Universitario (UNC)
instname_str Universidad Nacional de Córdoba
instacron_str UNC
institution UNC
repository.name.fl_str_mv Repositorio Digital Universitario (UNC) - Universidad Nacional de Córdoba
repository.mail.fl_str_mv oca.unc@gmail.com
_version_ 1844618952766914560
score 13.070432