Fricke and polymer gel 2D dosimetry validation using Monte Carlo simulation
- Autores
- Vedelago, José Alberto; Chacón Obando, D.; Malano, Francisco Mauricio; Conejeros, R.; Figueroa, R.; Garcia, D.; González, G.; Romero, Marcelo Ricardo; Santibañez, M.; Strumia, Miriam Cristina; Velásquez, J.; Mattea, Facundo; Valente, M.
- Año de publicación
- 2016
- Idioma
- español castellano
- Tipo de recurso
- artículo
- Estado
- versión publicada
- Descripción
- Complexity in modern radiotherapy treatments demands advanced dosimetry systems for quality control. These systems must have several characteristics, such as high spatial resolution, tissue equivalence, three-dimensional resolution, and dose-integrating capabilities. In this scenario, gel dosimetry has proved to be a very promising option for quality assurance. In this study, the feasibility of Fricke and polymer gel dosimeters suitably shaped in form of thin layers and optically analyzed by visible light transmission imaging has been investigated for quality assurance in external radiotherapy. Dosimeter irradiation was carried out with a 6-MV photon beam (CLINAC 600C). The analysis of the irradiated dosimeters was done using two-dimensional optical transmission images. These dosimeters were compared with a treatment plan system using Monte Carlo simulations as a reference by means of a gamma test with parameters of 1 mm and 2%. Results show very good agreement between the different dosimetric systems: in the worst-case scenario, 98% of the analyzed points meet the test quality requirements. Therefore, gel dosimetry may be considered as a potential tool for the validation of other dosimetric systems.
Fil: Vedelago, José Alberto. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - Córdoba. Instituto de Física Enrique Gaviola. Universidad Nacional de Córdoba. Instituto de Física Enrique Gaviola; Argentina. Universidad Nacional de Córdoba; Argentina
Fil: Chacón Obando, D.. Universidad Nacional. Physics Department; Costa Rica. Universidad Nacional de Córdoba; Argentina
Fil: Malano, Francisco Mauricio. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - Córdoba. Instituto de Física Enrique Gaviola. Universidad Nacional de Córdoba. Instituto de Física Enrique Gaviola; Argentina. Universidad Nacional de Córdoba; Argentina
Fil: Conejeros, R.. Servicio de Radioterapia, Icos. Temuco, Chile;
Fil: Figueroa, R.. Universidad de la Frontera; Chile
Fil: Garcia, D.. Servicio de Imagenes por Resonancia Magnética; Chile
Fil: González, G.. Servicio de Imagenes por Resonancia Magnética; Chile
Fil: Romero, Marcelo Ricardo. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - Córdoba. Instituto Multidisciplinario de Biología Vegetal. Universidad Nacional de Córdoba. Facultad de Ciencias Exactas Físicas y Naturales. Instituto Multidisciplinario de Biología Vegetal; Argentina. Universidad Nacional de Córdoba; Argentina
Fil: Santibañez, M.. Servicio de Imagenes por Resonancia Magnética; Chile
Fil: Strumia, Miriam Cristina. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - Córdoba. Instituto Multidisciplinario de Biología Vegetal. Universidad Nacional de Córdoba. Facultad de Ciencias Exactas Físicas y Naturales. Instituto Multidisciplinario de Biología Vegetal; Argentina
Fil: Velásquez, J.. Servicio de Radioterapia; Chile
Fil: Mattea, Facundo. Universidad Nacional de Córdoba. Facultad de Ciencias Químicas. Departamento de Química Orgánica; Argentina. Consejo Nacional de Investigaciones Científicas y Técnicas; Argentina
Fil: Valente, M.. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - Córdoba. Instituto de Física Enrique Gaviola. Universidad Nacional de Córdoba. Instituto de Física Enrique Gaviola; Argentina. Universidad de La Frontera. Departamento de Ciencias Físicas; Chile - Materia
-
Gel Dosimetry
Quality Assurance
Gamma Index
Monte Carlo Simulation
Fricke Gel - Nivel de accesibilidad
- acceso abierto
- Condiciones de uso
- Atribución-NoComercial-CompartirIgual 2.5 Argentina (CC BY-NC-SA 2.5 AR)
- Repositorio
- Institución
- Consejo Nacional de Investigaciones Científicas y Técnicas
- OAI Identificador
- oai:ri.conicet.gov.ar:11336/23759
Ver los metadatos del registro completo
id |
CONICETDig_e8ede41346de7e49e3013e8d25a6302e |
---|---|
oai_identifier_str |
oai:ri.conicet.gov.ar:11336/23759 |
network_acronym_str |
CONICETDig |
repository_id_str |
3498 |
network_name_str |
CONICET Digital (CONICET) |
spelling |
Fricke and polymer gel 2D dosimetry validation using Monte Carlo simulationVedelago, José AlbertoChacón Obando, D.Malano, Francisco MauricioConejeros, R.Figueroa, R.Garcia, D.González, G.Romero, Marcelo RicardoSantibañez, M.Strumia, Miriam CristinaVelásquez, J.Mattea, FacundoValente, M.Gel DosimetryQuality AssuranceGamma IndexMonte Carlo SimulationFricke Gelhttps://purl.org/becyt/ford/1.3https://purl.org/becyt/ford/1https://purl.org/becyt/ford/1.3https://purl.org/becyt/ford/1Complexity in modern radiotherapy treatments demands advanced dosimetry systems for quality control. These systems must have several characteristics, such as high spatial resolution, tissue equivalence, three-dimensional resolution, and dose-integrating capabilities. In this scenario, gel dosimetry has proved to be a very promising option for quality assurance. In this study, the feasibility of Fricke and polymer gel dosimeters suitably shaped in form of thin layers and optically analyzed by visible light transmission imaging has been investigated for quality assurance in external radiotherapy. Dosimeter irradiation was carried out with a 6-MV photon beam (CLINAC 600C). The analysis of the irradiated dosimeters was done using two-dimensional optical transmission images. These dosimeters were compared with a treatment plan system using Monte Carlo simulations as a reference by means of a gamma test with parameters of 1 mm and 2%. Results show very good agreement between the different dosimetric systems: in the worst-case scenario, 98% of the analyzed points meet the test quality requirements. Therefore, gel dosimetry may be considered as a potential tool for the validation of other dosimetric systems.Fil: Vedelago, José Alberto. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - Córdoba. Instituto de Física Enrique Gaviola. Universidad Nacional de Córdoba. Instituto de Física Enrique Gaviola; Argentina. Universidad Nacional de Córdoba; ArgentinaFil: Chacón Obando, D.. Universidad Nacional. Physics Department; Costa Rica. Universidad Nacional de Córdoba; ArgentinaFil: Malano, Francisco Mauricio. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - Córdoba. Instituto de Física Enrique Gaviola. Universidad Nacional de Córdoba. Instituto de Física Enrique Gaviola; Argentina. Universidad Nacional de Córdoba; ArgentinaFil: Conejeros, R.. Servicio de Radioterapia, Icos. Temuco, Chile;Fil: Figueroa, R.. Universidad de la Frontera; ChileFil: Garcia, D.. Servicio de Imagenes por Resonancia Magnética; ChileFil: González, G.. Servicio de Imagenes por Resonancia Magnética; ChileFil: Romero, Marcelo Ricardo. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - Córdoba. Instituto Multidisciplinario de Biología Vegetal. Universidad Nacional de Córdoba. Facultad de Ciencias Exactas Físicas y Naturales. Instituto Multidisciplinario de Biología Vegetal; Argentina. Universidad Nacional de Córdoba; ArgentinaFil: Santibañez, M.. Servicio de Imagenes por Resonancia Magnética; ChileFil: Strumia, Miriam Cristina. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - Córdoba. Instituto Multidisciplinario de Biología Vegetal. Universidad Nacional de Córdoba. Facultad de Ciencias Exactas Físicas y Naturales. Instituto Multidisciplinario de Biología Vegetal; ArgentinaFil: Velásquez, J.. Servicio de Radioterapia; ChileFil: Mattea, Facundo. Universidad Nacional de Córdoba. Facultad de Ciencias Químicas. Departamento de Química Orgánica; Argentina. Consejo Nacional de Investigaciones Científicas y Técnicas; ArgentinaFil: Valente, M.. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - Córdoba. Instituto de Física Enrique Gaviola. Universidad Nacional de Córdoba. Instituto de Física Enrique Gaviola; Argentina. Universidad de La Frontera. Departamento de Ciencias Físicas; ChileElsevier2016-08info:eu-repo/semantics/articleinfo:eu-repo/semantics/publishedVersionhttp://purl.org/coar/resource_type/c_6501info:ar-repo/semantics/articuloapplication/pdfapplication/pdfapplication/pdfapplication/pdfapplication/pdfapplication/pdfapplication/pdfhttp://hdl.handle.net/11336/23759Vedelago, José Alberto; Chacón Obando, D.; Malano, Francisco Mauricio; Conejeros, R.; Figueroa, R.; et al.; Fricke and polymer gel 2D dosimetry validation using Monte Carlo simulation; Elsevier; Radiation Measurements; 91; 8-2016; 54-641350-4487CONICET DigitalCONICETspainfo:eu-repo/semantics/altIdentifier/url/http://www.sciencedirect.com/science/article/pii/S1350448716301202info:eu-repo/semantics/altIdentifier/doi/10.1016/j.radmeas.2016.05.003info:eu-repo/semantics/openAccessAtribución-NoComercial-CompartirIgual 2.5 Argentina (CC BY-NC-SA 2.5 AR)https://creativecommons.org/licenses/by-nc-sa/2.5/ar/reponame:CONICET Digital (CONICET)instname:Consejo Nacional de Investigaciones Científicas y Técnicas2025-09-29T09:37:06Zoai:ri.conicet.gov.ar:11336/23759instacron:CONICETInstitucionalhttp://ri.conicet.gov.ar/Organismo científico-tecnológicoNo correspondehttp://ri.conicet.gov.ar/oai/requestdasensio@conicet.gov.ar; lcarlino@conicet.gov.arArgentinaNo correspondeNo correspondeNo correspondeopendoar:34982025-09-29 09:37:07.285CONICET Digital (CONICET) - Consejo Nacional de Investigaciones Científicas y Técnicasfalse |
dc.title.none.fl_str_mv |
Fricke and polymer gel 2D dosimetry validation using Monte Carlo simulation |
title |
Fricke and polymer gel 2D dosimetry validation using Monte Carlo simulation |
spellingShingle |
Fricke and polymer gel 2D dosimetry validation using Monte Carlo simulation Vedelago, José Alberto Gel Dosimetry Quality Assurance Gamma Index Monte Carlo Simulation Fricke Gel |
title_short |
Fricke and polymer gel 2D dosimetry validation using Monte Carlo simulation |
title_full |
Fricke and polymer gel 2D dosimetry validation using Monte Carlo simulation |
title_fullStr |
Fricke and polymer gel 2D dosimetry validation using Monte Carlo simulation |
title_full_unstemmed |
Fricke and polymer gel 2D dosimetry validation using Monte Carlo simulation |
title_sort |
Fricke and polymer gel 2D dosimetry validation using Monte Carlo simulation |
dc.creator.none.fl_str_mv |
Vedelago, José Alberto Chacón Obando, D. Malano, Francisco Mauricio Conejeros, R. Figueroa, R. Garcia, D. González, G. Romero, Marcelo Ricardo Santibañez, M. Strumia, Miriam Cristina Velásquez, J. Mattea, Facundo Valente, M. |
author |
Vedelago, José Alberto |
author_facet |
Vedelago, José Alberto Chacón Obando, D. Malano, Francisco Mauricio Conejeros, R. Figueroa, R. Garcia, D. González, G. Romero, Marcelo Ricardo Santibañez, M. Strumia, Miriam Cristina Velásquez, J. Mattea, Facundo Valente, M. |
author_role |
author |
author2 |
Chacón Obando, D. Malano, Francisco Mauricio Conejeros, R. Figueroa, R. Garcia, D. González, G. Romero, Marcelo Ricardo Santibañez, M. Strumia, Miriam Cristina Velásquez, J. Mattea, Facundo Valente, M. |
author2_role |
author author author author author author author author author author author author |
dc.subject.none.fl_str_mv |
Gel Dosimetry Quality Assurance Gamma Index Monte Carlo Simulation Fricke Gel |
topic |
Gel Dosimetry Quality Assurance Gamma Index Monte Carlo Simulation Fricke Gel |
purl_subject.fl_str_mv |
https://purl.org/becyt/ford/1.3 https://purl.org/becyt/ford/1 https://purl.org/becyt/ford/1.3 https://purl.org/becyt/ford/1 |
dc.description.none.fl_txt_mv |
Complexity in modern radiotherapy treatments demands advanced dosimetry systems for quality control. These systems must have several characteristics, such as high spatial resolution, tissue equivalence, three-dimensional resolution, and dose-integrating capabilities. In this scenario, gel dosimetry has proved to be a very promising option for quality assurance. In this study, the feasibility of Fricke and polymer gel dosimeters suitably shaped in form of thin layers and optically analyzed by visible light transmission imaging has been investigated for quality assurance in external radiotherapy. Dosimeter irradiation was carried out with a 6-MV photon beam (CLINAC 600C). The analysis of the irradiated dosimeters was done using two-dimensional optical transmission images. These dosimeters were compared with a treatment plan system using Monte Carlo simulations as a reference by means of a gamma test with parameters of 1 mm and 2%. Results show very good agreement between the different dosimetric systems: in the worst-case scenario, 98% of the analyzed points meet the test quality requirements. Therefore, gel dosimetry may be considered as a potential tool for the validation of other dosimetric systems. Fil: Vedelago, José Alberto. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - Córdoba. Instituto de Física Enrique Gaviola. Universidad Nacional de Córdoba. Instituto de Física Enrique Gaviola; Argentina. Universidad Nacional de Córdoba; Argentina Fil: Chacón Obando, D.. Universidad Nacional. Physics Department; Costa Rica. Universidad Nacional de Córdoba; Argentina Fil: Malano, Francisco Mauricio. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - Córdoba. Instituto de Física Enrique Gaviola. Universidad Nacional de Córdoba. Instituto de Física Enrique Gaviola; Argentina. Universidad Nacional de Córdoba; Argentina Fil: Conejeros, R.. Servicio de Radioterapia, Icos. Temuco, Chile; Fil: Figueroa, R.. Universidad de la Frontera; Chile Fil: Garcia, D.. Servicio de Imagenes por Resonancia Magnética; Chile Fil: González, G.. Servicio de Imagenes por Resonancia Magnética; Chile Fil: Romero, Marcelo Ricardo. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - Córdoba. Instituto Multidisciplinario de Biología Vegetal. Universidad Nacional de Córdoba. Facultad de Ciencias Exactas Físicas y Naturales. Instituto Multidisciplinario de Biología Vegetal; Argentina. Universidad Nacional de Córdoba; Argentina Fil: Santibañez, M.. Servicio de Imagenes por Resonancia Magnética; Chile Fil: Strumia, Miriam Cristina. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - Córdoba. Instituto Multidisciplinario de Biología Vegetal. Universidad Nacional de Córdoba. Facultad de Ciencias Exactas Físicas y Naturales. Instituto Multidisciplinario de Biología Vegetal; Argentina Fil: Velásquez, J.. Servicio de Radioterapia; Chile Fil: Mattea, Facundo. Universidad Nacional de Córdoba. Facultad de Ciencias Químicas. Departamento de Química Orgánica; Argentina. Consejo Nacional de Investigaciones Científicas y Técnicas; Argentina Fil: Valente, M.. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - Córdoba. Instituto de Física Enrique Gaviola. Universidad Nacional de Córdoba. Instituto de Física Enrique Gaviola; Argentina. Universidad de La Frontera. Departamento de Ciencias Físicas; Chile |
description |
Complexity in modern radiotherapy treatments demands advanced dosimetry systems for quality control. These systems must have several characteristics, such as high spatial resolution, tissue equivalence, three-dimensional resolution, and dose-integrating capabilities. In this scenario, gel dosimetry has proved to be a very promising option for quality assurance. In this study, the feasibility of Fricke and polymer gel dosimeters suitably shaped in form of thin layers and optically analyzed by visible light transmission imaging has been investigated for quality assurance in external radiotherapy. Dosimeter irradiation was carried out with a 6-MV photon beam (CLINAC 600C). The analysis of the irradiated dosimeters was done using two-dimensional optical transmission images. These dosimeters were compared with a treatment plan system using Monte Carlo simulations as a reference by means of a gamma test with parameters of 1 mm and 2%. Results show very good agreement between the different dosimetric systems: in the worst-case scenario, 98% of the analyzed points meet the test quality requirements. Therefore, gel dosimetry may be considered as a potential tool for the validation of other dosimetric systems. |
publishDate |
2016 |
dc.date.none.fl_str_mv |
2016-08 |
dc.type.none.fl_str_mv |
info:eu-repo/semantics/article info:eu-repo/semantics/publishedVersion http://purl.org/coar/resource_type/c_6501 info:ar-repo/semantics/articulo |
format |
article |
status_str |
publishedVersion |
dc.identifier.none.fl_str_mv |
http://hdl.handle.net/11336/23759 Vedelago, José Alberto; Chacón Obando, D.; Malano, Francisco Mauricio; Conejeros, R.; Figueroa, R.; et al.; Fricke and polymer gel 2D dosimetry validation using Monte Carlo simulation; Elsevier; Radiation Measurements; 91; 8-2016; 54-64 1350-4487 CONICET Digital CONICET |
url |
http://hdl.handle.net/11336/23759 |
identifier_str_mv |
Vedelago, José Alberto; Chacón Obando, D.; Malano, Francisco Mauricio; Conejeros, R.; Figueroa, R.; et al.; Fricke and polymer gel 2D dosimetry validation using Monte Carlo simulation; Elsevier; Radiation Measurements; 91; 8-2016; 54-64 1350-4487 CONICET Digital CONICET |
dc.language.none.fl_str_mv |
spa |
language |
spa |
dc.relation.none.fl_str_mv |
info:eu-repo/semantics/altIdentifier/url/http://www.sciencedirect.com/science/article/pii/S1350448716301202 info:eu-repo/semantics/altIdentifier/doi/10.1016/j.radmeas.2016.05.003 |
dc.rights.none.fl_str_mv |
info:eu-repo/semantics/openAccess Atribución-NoComercial-CompartirIgual 2.5 Argentina (CC BY-NC-SA 2.5 AR) https://creativecommons.org/licenses/by-nc-sa/2.5/ar/ |
eu_rights_str_mv |
openAccess |
rights_invalid_str_mv |
Atribución-NoComercial-CompartirIgual 2.5 Argentina (CC BY-NC-SA 2.5 AR) https://creativecommons.org/licenses/by-nc-sa/2.5/ar/ |
dc.format.none.fl_str_mv |
application/pdf application/pdf application/pdf application/pdf application/pdf application/pdf application/pdf |
dc.publisher.none.fl_str_mv |
Elsevier |
publisher.none.fl_str_mv |
Elsevier |
dc.source.none.fl_str_mv |
reponame:CONICET Digital (CONICET) instname:Consejo Nacional de Investigaciones Científicas y Técnicas |
reponame_str |
CONICET Digital (CONICET) |
collection |
CONICET Digital (CONICET) |
instname_str |
Consejo Nacional de Investigaciones Científicas y Técnicas |
repository.name.fl_str_mv |
CONICET Digital (CONICET) - Consejo Nacional de Investigaciones Científicas y Técnicas |
repository.mail.fl_str_mv |
dasensio@conicet.gov.ar; lcarlino@conicet.gov.ar |
_version_ |
1844613168154804224 |
score |
13.070432 |