- In this manuscript we deal with recent advances in low-field Magnetic Resonance Imaging (MRI). The development of low-cost MRI solutions allowing portability and trustable diagnosis is a hot topic worldwide by these days. We analyze basic technical issues of recent examples of fixed-field instruments operating at low-field. Then we discuss pros and cons of the pre-polarized approach, from both physical and technical perspectives. Permanent magnet and electromagnet technology are confronted. Finally, magnetic field-cycling is introduced as an alternative MRI technique, where field-dependent experiments can be exploded for the development of new contrast mechanisms that are not feasible for fixed-field MRI instruments. As field cycled machines usually deals with switched currents in electromagnets, magnetic field instability and inhomogeneity are the main limiting factors affecting image quality. We finalize this manuscript discussing how it turns possible to overcome these limitations, thus opening new possibilities for the development of cost effective MRI technology.
Fil: Anoardo, Esteban. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - Córdoba. Instituto de Física Enrique Gaviola. Universidad Nacional de Córdoba. Instituto de Física Enrique Gaviola; Argentina. Universidad Nacional de Córdoba. Facultad de Matemática, Astronomía y Física; Argentina
Fil: Rodriguez, Gonzalo G.. University of New York; Estados Unidos - Materia
-
FIELD-CYCLING
LOW-FIELD
MAGNET
MRI
PRE-POLARIZED - Nivel de accesibilidad
- acceso abierto
- Condiciones de uso
- https://creativecommons.org/licenses/by-nc-nd/2.5/ar/
- Repositorio
- Institución
- Consejo Nacional de Investigaciones Científicas y Técnicas
- OAI Identificador
- oai:ri.conicet.gov.ar:11336/226475
Ver los metadatos del registro completo
id |
CONICETDig_b1922dc666f4bc74f7e8061ce8115e9b |
---|---|
oai_identifier_str |
oai:ri.conicet.gov.ar:11336/226475 |
network_acronym_str |
CONICETDig |
repository_id_str |
3498 |
network_name_str |
CONICET Digital (CONICET) |
spelling |
New challenges and opportunities for low-field MRIAnoardo, EstebanRodriguez, Gonzalo G.FIELD-CYCLINGLOW-FIELDMAGNETMRIPRE-POLARIZEDhttps://purl.org/becyt/ford/1.3https://purl.org/becyt/ford/1In this manuscript we deal with recent advances in low-field Magnetic Resonance Imaging (MRI). The development of low-cost MRI solutions allowing portability and trustable diagnosis is a hot topic worldwide by these days. We analyze basic technical issues of recent examples of fixed-field instruments operating at low-field. Then we discuss pros and cons of the pre-polarized approach, from both physical and technical perspectives. Permanent magnet and electromagnet technology are confronted. Finally, magnetic field-cycling is introduced as an alternative MRI technique, where field-dependent experiments can be exploded for the development of new contrast mechanisms that are not feasible for fixed-field MRI instruments. As field cycled machines usually deals with switched currents in electromagnets, magnetic field instability and inhomogeneity are the main limiting factors affecting image quality. We finalize this manuscript discussing how it turns possible to overcome these limitations, thus opening new possibilities for the development of cost effective MRI technology.Fil: Anoardo, Esteban. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - Córdoba. Instituto de Física Enrique Gaviola. Universidad Nacional de Córdoba. Instituto de Física Enrique Gaviola; Argentina. Universidad Nacional de Córdoba. Facultad de Matemática, Astronomía y Física; ArgentinaFil: Rodriguez, Gonzalo G.. University of New York; Estados UnidosElsevier2023-06info:eu-repo/semantics/articleinfo:eu-repo/semantics/publishedVersionhttp://purl.org/coar/resource_type/c_6501info:ar-repo/semantics/articuloapplication/pdfapplication/pdfapplication/pdfhttp://hdl.handle.net/11336/226475Anoardo, Esteban; Rodriguez, Gonzalo G.; New challenges and opportunities for low-field MRI; Elsevier; Journal of Magnetic Resonance Open; 14-15; 6-2023; 1-102666-4410CONICET DigitalCONICETenginfo:eu-repo/semantics/altIdentifier/url/https://linkinghub.elsevier.com/retrieve/pii/S2666441022000565info:eu-repo/semantics/altIdentifier/doi/10.1016/j.jmro.2022.100086info:eu-repo/semantics/openAccesshttps://creativecommons.org/licenses/by-nc-nd/2.5/ar/reponame:CONICET Digital (CONICET)instname:Consejo Nacional de Investigaciones Científicas y Técnicas2025-09-29T09:35:38Zoai:ri.conicet.gov.ar:11336/226475instacron:CONICETInstitucionalhttp://ri.conicet.gov.ar/Organismo científico-tecnológicoNo correspondehttp://ri.conicet.gov.ar/oai/requestdasensio@conicet.gov.ar; lcarlino@conicet.gov.arArgentinaNo correspondeNo correspondeNo correspondeopendoar:34982025-09-29 09:35:38.28CONICET Digital (CONICET) - Consejo Nacional de Investigaciones Científicas y Técnicasfalse |
dc.title.none.fl_str_mv |
New challenges and opportunities for low-field MRI |
title |
New challenges and opportunities for low-field MRI |
spellingShingle |
New challenges and opportunities for low-field MRI Anoardo, Esteban FIELD-CYCLING LOW-FIELD MAGNET MRI PRE-POLARIZED |
title_short |
New challenges and opportunities for low-field MRI |
title_full |
New challenges and opportunities for low-field MRI |
title_fullStr |
New challenges and opportunities for low-field MRI |
title_full_unstemmed |
New challenges and opportunities for low-field MRI |
title_sort |
New challenges and opportunities for low-field MRI |
dc.creator.none.fl_str_mv |
Anoardo, Esteban Rodriguez, Gonzalo G. |
author |
Anoardo, Esteban |
author_facet |
Anoardo, Esteban Rodriguez, Gonzalo G. |
author_role |
author |
author2 |
Rodriguez, Gonzalo G. |
author2_role |
author |
dc.subject.none.fl_str_mv |
FIELD-CYCLING LOW-FIELD MAGNET MRI PRE-POLARIZED |
topic |
FIELD-CYCLING LOW-FIELD MAGNET MRI PRE-POLARIZED |
purl_subject.fl_str_mv |
https://purl.org/becyt/ford/1.3 https://purl.org/becyt/ford/1 |
dc.description.none.fl_txt_mv |
In this manuscript we deal with recent advances in low-field Magnetic Resonance Imaging (MRI). The development of low-cost MRI solutions allowing portability and trustable diagnosis is a hot topic worldwide by these days. We analyze basic technical issues of recent examples of fixed-field instruments operating at low-field. Then we discuss pros and cons of the pre-polarized approach, from both physical and technical perspectives. Permanent magnet and electromagnet technology are confronted. Finally, magnetic field-cycling is introduced as an alternative MRI technique, where field-dependent experiments can be exploded for the development of new contrast mechanisms that are not feasible for fixed-field MRI instruments. As field cycled machines usually deals with switched currents in electromagnets, magnetic field instability and inhomogeneity are the main limiting factors affecting image quality. We finalize this manuscript discussing how it turns possible to overcome these limitations, thus opening new possibilities for the development of cost effective MRI technology. Fil: Anoardo, Esteban. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - Córdoba. Instituto de Física Enrique Gaviola. Universidad Nacional de Córdoba. Instituto de Física Enrique Gaviola; Argentina. Universidad Nacional de Córdoba. Facultad de Matemática, Astronomía y Física; Argentina Fil: Rodriguez, Gonzalo G.. University of New York; Estados Unidos |
description |
In this manuscript we deal with recent advances in low-field Magnetic Resonance Imaging (MRI). The development of low-cost MRI solutions allowing portability and trustable diagnosis is a hot topic worldwide by these days. We analyze basic technical issues of recent examples of fixed-field instruments operating at low-field. Then we discuss pros and cons of the pre-polarized approach, from both physical and technical perspectives. Permanent magnet and electromagnet technology are confronted. Finally, magnetic field-cycling is introduced as an alternative MRI technique, where field-dependent experiments can be exploded for the development of new contrast mechanisms that are not feasible for fixed-field MRI instruments. As field cycled machines usually deals with switched currents in electromagnets, magnetic field instability and inhomogeneity are the main limiting factors affecting image quality. We finalize this manuscript discussing how it turns possible to overcome these limitations, thus opening new possibilities for the development of cost effective MRI technology. |
publishDate |
2023 |
dc.date.none.fl_str_mv |
2023-06 |
dc.type.none.fl_str_mv |
info:eu-repo/semantics/article info:eu-repo/semantics/publishedVersion http://purl.org/coar/resource_type/c_6501 info:ar-repo/semantics/articulo |
format |
article |
status_str |
publishedVersion |
dc.identifier.none.fl_str_mv |
http://hdl.handle.net/11336/226475 Anoardo, Esteban; Rodriguez, Gonzalo G.; New challenges and opportunities for low-field MRI; Elsevier; Journal of Magnetic Resonance Open; 14-15; 6-2023; 1-10 2666-4410 CONICET Digital CONICET |
url |
http://hdl.handle.net/11336/226475 |
identifier_str_mv |
Anoardo, Esteban; Rodriguez, Gonzalo G.; New challenges and opportunities for low-field MRI; Elsevier; Journal of Magnetic Resonance Open; 14-15; 6-2023; 1-10 2666-4410 CONICET Digital CONICET |
dc.language.none.fl_str_mv |
eng |
language |
eng |
dc.relation.none.fl_str_mv |
info:eu-repo/semantics/altIdentifier/url/https://linkinghub.elsevier.com/retrieve/pii/S2666441022000565 info:eu-repo/semantics/altIdentifier/doi/10.1016/j.jmro.2022.100086 |
dc.rights.none.fl_str_mv |
info:eu-repo/semantics/openAccess https://creativecommons.org/licenses/by-nc-nd/2.5/ar/ |
eu_rights_str_mv |
openAccess |
rights_invalid_str_mv |
https://creativecommons.org/licenses/by-nc-nd/2.5/ar/ |
dc.format.none.fl_str_mv |
application/pdf application/pdf application/pdf |
dc.publisher.none.fl_str_mv |
Elsevier |
publisher.none.fl_str_mv |
Elsevier |
dc.source.none.fl_str_mv |
reponame:CONICET Digital (CONICET) instname:Consejo Nacional de Investigaciones Científicas y Técnicas |
reponame_str |
CONICET Digital (CONICET) |
collection |
CONICET Digital (CONICET) |
instname_str |
Consejo Nacional de Investigaciones Científicas y Técnicas |
repository.name.fl_str_mv |
CONICET Digital (CONICET) - Consejo Nacional de Investigaciones Científicas y Técnicas |
repository.mail.fl_str_mv |
dasensio@conicet.gov.ar; lcarlino@conicet.gov.ar |
_version_ |
1844613111778115584 |
score |
13.070432 |