Laboratory evaluation of two LISST-25X using river sediments
- Autores
- Filippa, Leonardo Marcelo; Freire, Luana; Trento, Alfredo Emilio; Alvarez, Ana Maria Teresita; Gallo, Marcos Esteban; Vinzón, Susana
- Año de publicación
- 2011
- Idioma
- inglés
- Tipo de recurso
- artículo
- Estado
- versión publicada
- Descripción
- There are different methods to determine particle size, most of them are only applicable in the laboratory. In order to describe the suspended fine sediment sizes where flocs may be part of the suspension, in situ measurements are essential. Thiswork shows the results of sediment size measurements done in the laboratory using two LISST-25X diffractometers, developed for in-situ measurements, and a Malvern Mastersizer 2000 diffractometer for laboratory use. The two LISST-25X models characterize the sample size through the Sauter Mean Diameter (SMD). Besides the SMD, the Malvern diffractometer determines the granulometric distributions of the samples. The tested samples are from different fluvial and estuarine environments (Paran¨¢, Salado and the Amazon Rivers) and their sizes (SMD) range from 4 to 300 ¦Ìm. The relationship between SMD obtained with the two LISST-25X and the Malvern diffractometers gave a determination coefficient of R2=0.98. Compared to the Malvern instrument, used as a reference, it was observed that the LISST instruments tend to overestimate the measured SMD for diameters lower than 20 ¦Ìm, and to underestimate the values of samples with larger diameters (N20 ¦Ìm). Aiming to obtain other characteristic diameters of the granulometric distribution from SMD measurements obtained with LISST-25X diffractometers, correlations between SMD and d10, d50 and d90, were established. The dependency of the measurement of sediment mass concentration obtained with LISST instruments, with the particle size, is also addressed.
Fil: Filippa, Leonardo Marcelo. Universidad Nacional del Litoral. Facultad de Ingeniería y Ciencias Hídricas; Argentina
Fil: Freire, Luana. Universidade Federal do Rio de Janeiro; Brasil
Fil: Trento, Alfredo Emilio. Universidad Nacional del Litoral. Facultad de Ingeniería y Ciencias Hídricas; Argentina
Fil: Alvarez, Ana Maria Teresita. Universidad Nacional del Litoral. Facultad de Ingeniería y Ciencias Hídricas; Argentina
Fil: Gallo, Marcos Esteban. Universidade Federal do Rio de Janeiro; Brasil
Fil: Vinzón, Susana. Universidade Federal do Rio de Janeiro; Brasil - Materia
-
LISST-25X
SAUTER_MEAN_DIAMETER_(SMD)
SEDIMENTS
GRANULOMETRIC_DISTRIBUTION - Nivel de accesibilidad
- acceso abierto
- Condiciones de uso
- https://creativecommons.org/licenses/by-nc-sa/2.5/ar/
- Repositorio
.jpg)
- Institución
- Consejo Nacional de Investigaciones Científicas y Técnicas
- OAI Identificador
- oai:ri.conicet.gov.ar:11336/103438
Ver los metadatos del registro completo
| id |
CONICETDig_7232beacd823750de68b181599ef3612 |
|---|---|
| oai_identifier_str |
oai:ri.conicet.gov.ar:11336/103438 |
| network_acronym_str |
CONICETDig |
| repository_id_str |
3498 |
| network_name_str |
CONICET Digital (CONICET) |
| spelling |
Laboratory evaluation of two LISST-25X using river sedimentsFilippa, Leonardo MarceloFreire, LuanaTrento, Alfredo EmilioAlvarez, Ana Maria TeresitaGallo, Marcos EstebanVinzón, SusanaLISST-25XSAUTER_MEAN_DIAMETER_(SMD)SEDIMENTSGRANULOMETRIC_DISTRIBUTIONhttps://purl.org/becyt/ford/2.7https://purl.org/becyt/ford/2There are different methods to determine particle size, most of them are only applicable in the laboratory. In order to describe the suspended fine sediment sizes where flocs may be part of the suspension, in situ measurements are essential. Thiswork shows the results of sediment size measurements done in the laboratory using two LISST-25X diffractometers, developed for in-situ measurements, and a Malvern Mastersizer 2000 diffractometer for laboratory use. The two LISST-25X models characterize the sample size through the Sauter Mean Diameter (SMD). Besides the SMD, the Malvern diffractometer determines the granulometric distributions of the samples. The tested samples are from different fluvial and estuarine environments (Paran¨¢, Salado and the Amazon Rivers) and their sizes (SMD) range from 4 to 300 ¦Ìm. The relationship between SMD obtained with the two LISST-25X and the Malvern diffractometers gave a determination coefficient of R2=0.98. Compared to the Malvern instrument, used as a reference, it was observed that the LISST instruments tend to overestimate the measured SMD for diameters lower than 20 ¦Ìm, and to underestimate the values of samples with larger diameters (N20 ¦Ìm). Aiming to obtain other characteristic diameters of the granulometric distribution from SMD measurements obtained with LISST-25X diffractometers, correlations between SMD and d10, d50 and d90, were established. The dependency of the measurement of sediment mass concentration obtained with LISST instruments, with the particle size, is also addressed.Fil: Filippa, Leonardo Marcelo. Universidad Nacional del Litoral. Facultad de Ingeniería y Ciencias Hídricas; ArgentinaFil: Freire, Luana. Universidade Federal do Rio de Janeiro; BrasilFil: Trento, Alfredo Emilio. Universidad Nacional del Litoral. Facultad de Ingeniería y Ciencias Hídricas; ArgentinaFil: Alvarez, Ana Maria Teresita. Universidad Nacional del Litoral. Facultad de Ingeniería y Ciencias Hídricas; ArgentinaFil: Gallo, Marcos Esteban. Universidade Federal do Rio de Janeiro; BrasilFil: Vinzón, Susana. Universidade Federal do Rio de Janeiro; BrasilElsevier Science2011-07info:eu-repo/semantics/articleinfo:eu-repo/semantics/publishedVersionhttp://purl.org/coar/resource_type/c_6501info:ar-repo/semantics/articuloapplication/pdfapplication/pdfapplication/pdfapplication/pdfhttp://hdl.handle.net/11336/103438Filippa, Leonardo Marcelo; Freire, Luana; Trento, Alfredo Emilio; Alvarez, Ana Maria Teresita; Gallo, Marcos Esteban; et al.; Laboratory evaluation of two LISST-25X using river sediments; Elsevier Science; Sedimentary Geology; 238; 3-4; 7-2011; 268-2760037-0738CONICET DigitalCONICETenginfo:eu-repo/semantics/altIdentifier/doi/10.1016/j.sedgeo.2011.04.017info:eu-repo/semantics/openAccesshttps://creativecommons.org/licenses/by-nc-sa/2.5/ar/reponame:CONICET Digital (CONICET)instname:Consejo Nacional de Investigaciones Científicas y Técnicas2025-10-22T11:00:16Zoai:ri.conicet.gov.ar:11336/103438instacron:CONICETInstitucionalhttp://ri.conicet.gov.ar/Organismo científico-tecnológicoNo correspondehttp://ri.conicet.gov.ar/oai/requestdasensio@conicet.gov.ar; lcarlino@conicet.gov.arArgentinaNo correspondeNo correspondeNo correspondeopendoar:34982025-10-22 11:00:16.764CONICET Digital (CONICET) - Consejo Nacional de Investigaciones Científicas y Técnicasfalse |
| dc.title.none.fl_str_mv |
Laboratory evaluation of two LISST-25X using river sediments |
| title |
Laboratory evaluation of two LISST-25X using river sediments |
| spellingShingle |
Laboratory evaluation of two LISST-25X using river sediments Filippa, Leonardo Marcelo LISST-25X SAUTER_MEAN_DIAMETER_(SMD) SEDIMENTS GRANULOMETRIC_DISTRIBUTION |
| title_short |
Laboratory evaluation of two LISST-25X using river sediments |
| title_full |
Laboratory evaluation of two LISST-25X using river sediments |
| title_fullStr |
Laboratory evaluation of two LISST-25X using river sediments |
| title_full_unstemmed |
Laboratory evaluation of two LISST-25X using river sediments |
| title_sort |
Laboratory evaluation of two LISST-25X using river sediments |
| dc.creator.none.fl_str_mv |
Filippa, Leonardo Marcelo Freire, Luana Trento, Alfredo Emilio Alvarez, Ana Maria Teresita Gallo, Marcos Esteban Vinzón, Susana |
| author |
Filippa, Leonardo Marcelo |
| author_facet |
Filippa, Leonardo Marcelo Freire, Luana Trento, Alfredo Emilio Alvarez, Ana Maria Teresita Gallo, Marcos Esteban Vinzón, Susana |
| author_role |
author |
| author2 |
Freire, Luana Trento, Alfredo Emilio Alvarez, Ana Maria Teresita Gallo, Marcos Esteban Vinzón, Susana |
| author2_role |
author author author author author |
| dc.subject.none.fl_str_mv |
LISST-25X SAUTER_MEAN_DIAMETER_(SMD) SEDIMENTS GRANULOMETRIC_DISTRIBUTION |
| topic |
LISST-25X SAUTER_MEAN_DIAMETER_(SMD) SEDIMENTS GRANULOMETRIC_DISTRIBUTION |
| purl_subject.fl_str_mv |
https://purl.org/becyt/ford/2.7 https://purl.org/becyt/ford/2 |
| dc.description.none.fl_txt_mv |
There are different methods to determine particle size, most of them are only applicable in the laboratory. In order to describe the suspended fine sediment sizes where flocs may be part of the suspension, in situ measurements are essential. Thiswork shows the results of sediment size measurements done in the laboratory using two LISST-25X diffractometers, developed for in-situ measurements, and a Malvern Mastersizer 2000 diffractometer for laboratory use. The two LISST-25X models characterize the sample size through the Sauter Mean Diameter (SMD). Besides the SMD, the Malvern diffractometer determines the granulometric distributions of the samples. The tested samples are from different fluvial and estuarine environments (Paran¨¢, Salado and the Amazon Rivers) and their sizes (SMD) range from 4 to 300 ¦Ìm. The relationship between SMD obtained with the two LISST-25X and the Malvern diffractometers gave a determination coefficient of R2=0.98. Compared to the Malvern instrument, used as a reference, it was observed that the LISST instruments tend to overestimate the measured SMD for diameters lower than 20 ¦Ìm, and to underestimate the values of samples with larger diameters (N20 ¦Ìm). Aiming to obtain other characteristic diameters of the granulometric distribution from SMD measurements obtained with LISST-25X diffractometers, correlations between SMD and d10, d50 and d90, were established. The dependency of the measurement of sediment mass concentration obtained with LISST instruments, with the particle size, is also addressed. Fil: Filippa, Leonardo Marcelo. Universidad Nacional del Litoral. Facultad de Ingeniería y Ciencias Hídricas; Argentina Fil: Freire, Luana. Universidade Federal do Rio de Janeiro; Brasil Fil: Trento, Alfredo Emilio. Universidad Nacional del Litoral. Facultad de Ingeniería y Ciencias Hídricas; Argentina Fil: Alvarez, Ana Maria Teresita. Universidad Nacional del Litoral. Facultad de Ingeniería y Ciencias Hídricas; Argentina Fil: Gallo, Marcos Esteban. Universidade Federal do Rio de Janeiro; Brasil Fil: Vinzón, Susana. Universidade Federal do Rio de Janeiro; Brasil |
| description |
There are different methods to determine particle size, most of them are only applicable in the laboratory. In order to describe the suspended fine sediment sizes where flocs may be part of the suspension, in situ measurements are essential. Thiswork shows the results of sediment size measurements done in the laboratory using two LISST-25X diffractometers, developed for in-situ measurements, and a Malvern Mastersizer 2000 diffractometer for laboratory use. The two LISST-25X models characterize the sample size through the Sauter Mean Diameter (SMD). Besides the SMD, the Malvern diffractometer determines the granulometric distributions of the samples. The tested samples are from different fluvial and estuarine environments (Paran¨¢, Salado and the Amazon Rivers) and their sizes (SMD) range from 4 to 300 ¦Ìm. The relationship between SMD obtained with the two LISST-25X and the Malvern diffractometers gave a determination coefficient of R2=0.98. Compared to the Malvern instrument, used as a reference, it was observed that the LISST instruments tend to overestimate the measured SMD for diameters lower than 20 ¦Ìm, and to underestimate the values of samples with larger diameters (N20 ¦Ìm). Aiming to obtain other characteristic diameters of the granulometric distribution from SMD measurements obtained with LISST-25X diffractometers, correlations between SMD and d10, d50 and d90, were established. The dependency of the measurement of sediment mass concentration obtained with LISST instruments, with the particle size, is also addressed. |
| publishDate |
2011 |
| dc.date.none.fl_str_mv |
2011-07 |
| dc.type.none.fl_str_mv |
info:eu-repo/semantics/article info:eu-repo/semantics/publishedVersion http://purl.org/coar/resource_type/c_6501 info:ar-repo/semantics/articulo |
| format |
article |
| status_str |
publishedVersion |
| dc.identifier.none.fl_str_mv |
http://hdl.handle.net/11336/103438 Filippa, Leonardo Marcelo; Freire, Luana; Trento, Alfredo Emilio; Alvarez, Ana Maria Teresita; Gallo, Marcos Esteban; et al.; Laboratory evaluation of two LISST-25X using river sediments; Elsevier Science; Sedimentary Geology; 238; 3-4; 7-2011; 268-276 0037-0738 CONICET Digital CONICET |
| url |
http://hdl.handle.net/11336/103438 |
| identifier_str_mv |
Filippa, Leonardo Marcelo; Freire, Luana; Trento, Alfredo Emilio; Alvarez, Ana Maria Teresita; Gallo, Marcos Esteban; et al.; Laboratory evaluation of two LISST-25X using river sediments; Elsevier Science; Sedimentary Geology; 238; 3-4; 7-2011; 268-276 0037-0738 CONICET Digital CONICET |
| dc.language.none.fl_str_mv |
eng |
| language |
eng |
| dc.relation.none.fl_str_mv |
info:eu-repo/semantics/altIdentifier/doi/10.1016/j.sedgeo.2011.04.017 |
| dc.rights.none.fl_str_mv |
info:eu-repo/semantics/openAccess https://creativecommons.org/licenses/by-nc-sa/2.5/ar/ |
| eu_rights_str_mv |
openAccess |
| rights_invalid_str_mv |
https://creativecommons.org/licenses/by-nc-sa/2.5/ar/ |
| dc.format.none.fl_str_mv |
application/pdf application/pdf application/pdf application/pdf |
| dc.publisher.none.fl_str_mv |
Elsevier Science |
| publisher.none.fl_str_mv |
Elsevier Science |
| dc.source.none.fl_str_mv |
reponame:CONICET Digital (CONICET) instname:Consejo Nacional de Investigaciones Científicas y Técnicas |
| reponame_str |
CONICET Digital (CONICET) |
| collection |
CONICET Digital (CONICET) |
| instname_str |
Consejo Nacional de Investigaciones Científicas y Técnicas |
| repository.name.fl_str_mv |
CONICET Digital (CONICET) - Consejo Nacional de Investigaciones Científicas y Técnicas |
| repository.mail.fl_str_mv |
dasensio@conicet.gov.ar; lcarlino@conicet.gov.ar |
| _version_ |
1846781160371257344 |
| score |
12.982451 |