Simple method to determine the resolution and sensitivity of systems for optoacoustic tomography
- Autores
- González, Martín Germán; Acosta, Eduardo Omar; Santiago, Guillermo
- Año de publicación
- 2018
- Idioma
- inglés
- Tipo de recurso
- artículo
- Estado
- versión publicada
- Descripción
- In this paper we present a method to determine the spatial resolution and sensitivity of systems for optoacoustic tomography (OAT). The method consists of obtaining the image of a sample based on a transparent film embedded in agarose. The film has a certain pattern made with a laser printer that allows to obtain the system spatial resolution. Moreover, since the damage threshold of the ink pattern is similar to that of living tissue, it is also possible to determine if the system is sensitive enough to be applied on biological samples. The method is straightforward, fast and repeatable, and was tested in a OAT system for obtaining two-dimensional images developed in our laboratory.
En este trabajo se presenta un método para determinar la resolución espacial y la sensibilidad de sistemas para tomografía optoacústica (TOA). Éste consiste en la obtención de la imagen de una muestra basada en una lámina transparente embebida en agarosa. La lámina posee un determinado patrón realizado con una impresora láser que permite determinar la resolución espacial del sistema. Por otro lado, como su umbral de daño es similar a del tejido vivo, también es posible establecer si el sistema posee la suficiente sensibilidad para ser usado en muestras biológicas. El método es directo, rápido y repetible, y fue probado en un sistema TOA para obtención de imágenes bidimensionales desarrollado en nuestro laboratorio.
Fil: González, Martín Germán. Consejo Nacional de Investigaciones Científicas y Técnicas; Argentina. Universidad de Buenos Aires. Facultad de Ingeniería; Argentina
Fil: Acosta, Eduardo Omar. Universidad de Buenos Aires. Facultad de Ingeniería; Argentina
Fil: Santiago, Guillermo. Universidad de Buenos Aires. Facultad de Ingeniería; Argentina - Materia
-
OPTOACOUSTIC TOMOGRAPHY
SPATIAL RESOLUTION
TWO-DIMENSIONAL IMAGES - Nivel de accesibilidad
- acceso abierto
- Condiciones de uso
- https://creativecommons.org/licenses/by-nc-nd/2.5/ar/
- Repositorio
- Institución
- Consejo Nacional de Investigaciones Científicas y Técnicas
- OAI Identificador
- oai:ri.conicet.gov.ar:11336/178236
Ver los metadatos del registro completo
id |
CONICETDig_5a3cf11b3e4b0cc3baa0d959bed5a059 |
---|---|
oai_identifier_str |
oai:ri.conicet.gov.ar:11336/178236 |
network_acronym_str |
CONICETDig |
repository_id_str |
3498 |
network_name_str |
CONICET Digital (CONICET) |
spelling |
Simple method to determine the resolution and sensitivity of systems for optoacoustic tomographyMétodo sencillo para determinar la resolución y sensibilidad de sistemas para tomografía optoacústicaGonzález, Martín GermánAcosta, Eduardo OmarSantiago, GuillermoOPTOACOUSTIC TOMOGRAPHYSPATIAL RESOLUTIONTWO-DIMENSIONAL IMAGEShttps://purl.org/becyt/ford/2.2https://purl.org/becyt/ford/2In this paper we present a method to determine the spatial resolution and sensitivity of systems for optoacoustic tomography (OAT). The method consists of obtaining the image of a sample based on a transparent film embedded in agarose. The film has a certain pattern made with a laser printer that allows to obtain the system spatial resolution. Moreover, since the damage threshold of the ink pattern is similar to that of living tissue, it is also possible to determine if the system is sensitive enough to be applied on biological samples. The method is straightforward, fast and repeatable, and was tested in a OAT system for obtaining two-dimensional images developed in our laboratory.En este trabajo se presenta un método para determinar la resolución espacial y la sensibilidad de sistemas para tomografía optoacústica (TOA). Éste consiste en la obtención de la imagen de una muestra basada en una lámina transparente embebida en agarosa. La lámina posee un determinado patrón realizado con una impresora láser que permite determinar la resolución espacial del sistema. Por otro lado, como su umbral de daño es similar a del tejido vivo, también es posible establecer si el sistema posee la suficiente sensibilidad para ser usado en muestras biológicas. El método es directo, rápido y repetible, y fue probado en un sistema TOA para obtención de imágenes bidimensionales desarrollado en nuestro laboratorio.Fil: González, Martín Germán. Consejo Nacional de Investigaciones Científicas y Técnicas; Argentina. Universidad de Buenos Aires. Facultad de Ingeniería; ArgentinaFil: Acosta, Eduardo Omar. Universidad de Buenos Aires. Facultad de Ingeniería; ArgentinaFil: Santiago, Guillermo. Universidad de Buenos Aires. Facultad de Ingeniería; ArgentinaUniversidad de Buenos Aires. Facultad de Ingeniería2018-12info:eu-repo/semantics/articleinfo:eu-repo/semantics/publishedVersionhttp://purl.org/coar/resource_type/c_6501info:ar-repo/semantics/articuloapplication/pdfapplication/pdfhttp://hdl.handle.net/11336/178236González, Martín Germán; Acosta, Eduardo Omar; Santiago, Guillermo; Simple method to determine the resolution and sensitivity of systems for optoacoustic tomography; Universidad de Buenos Aires. Facultad de Ingeniería; Elektron; 2; 2; 12-2018; 63-662525-0159CONICET DigitalCONICETenginfo:eu-repo/semantics/altIdentifier/url/http://elektron.fi.uba.ar/index.php/elektron/article/view/47info:eu-repo/semantics/openAccesshttps://creativecommons.org/licenses/by-nc-nd/2.5/ar/reponame:CONICET Digital (CONICET)instname:Consejo Nacional de Investigaciones Científicas y Técnicas2025-09-03T09:55:58Zoai:ri.conicet.gov.ar:11336/178236instacron:CONICETInstitucionalhttp://ri.conicet.gov.ar/Organismo científico-tecnológicoNo correspondehttp://ri.conicet.gov.ar/oai/requestdasensio@conicet.gov.ar; lcarlino@conicet.gov.arArgentinaNo correspondeNo correspondeNo correspondeopendoar:34982025-09-03 09:55:58.849CONICET Digital (CONICET) - Consejo Nacional de Investigaciones Científicas y Técnicasfalse |
dc.title.none.fl_str_mv |
Simple method to determine the resolution and sensitivity of systems for optoacoustic tomography Método sencillo para determinar la resolución y sensibilidad de sistemas para tomografía optoacústica |
title |
Simple method to determine the resolution and sensitivity of systems for optoacoustic tomography |
spellingShingle |
Simple method to determine the resolution and sensitivity of systems for optoacoustic tomography González, Martín Germán OPTOACOUSTIC TOMOGRAPHY SPATIAL RESOLUTION TWO-DIMENSIONAL IMAGES |
title_short |
Simple method to determine the resolution and sensitivity of systems for optoacoustic tomography |
title_full |
Simple method to determine the resolution and sensitivity of systems for optoacoustic tomography |
title_fullStr |
Simple method to determine the resolution and sensitivity of systems for optoacoustic tomography |
title_full_unstemmed |
Simple method to determine the resolution and sensitivity of systems for optoacoustic tomography |
title_sort |
Simple method to determine the resolution and sensitivity of systems for optoacoustic tomography |
dc.creator.none.fl_str_mv |
González, Martín Germán Acosta, Eduardo Omar Santiago, Guillermo |
author |
González, Martín Germán |
author_facet |
González, Martín Germán Acosta, Eduardo Omar Santiago, Guillermo |
author_role |
author |
author2 |
Acosta, Eduardo Omar Santiago, Guillermo |
author2_role |
author author |
dc.subject.none.fl_str_mv |
OPTOACOUSTIC TOMOGRAPHY SPATIAL RESOLUTION TWO-DIMENSIONAL IMAGES |
topic |
OPTOACOUSTIC TOMOGRAPHY SPATIAL RESOLUTION TWO-DIMENSIONAL IMAGES |
purl_subject.fl_str_mv |
https://purl.org/becyt/ford/2.2 https://purl.org/becyt/ford/2 |
dc.description.none.fl_txt_mv |
In this paper we present a method to determine the spatial resolution and sensitivity of systems for optoacoustic tomography (OAT). The method consists of obtaining the image of a sample based on a transparent film embedded in agarose. The film has a certain pattern made with a laser printer that allows to obtain the system spatial resolution. Moreover, since the damage threshold of the ink pattern is similar to that of living tissue, it is also possible to determine if the system is sensitive enough to be applied on biological samples. The method is straightforward, fast and repeatable, and was tested in a OAT system for obtaining two-dimensional images developed in our laboratory. En este trabajo se presenta un método para determinar la resolución espacial y la sensibilidad de sistemas para tomografía optoacústica (TOA). Éste consiste en la obtención de la imagen de una muestra basada en una lámina transparente embebida en agarosa. La lámina posee un determinado patrón realizado con una impresora láser que permite determinar la resolución espacial del sistema. Por otro lado, como su umbral de daño es similar a del tejido vivo, también es posible establecer si el sistema posee la suficiente sensibilidad para ser usado en muestras biológicas. El método es directo, rápido y repetible, y fue probado en un sistema TOA para obtención de imágenes bidimensionales desarrollado en nuestro laboratorio. Fil: González, Martín Germán. Consejo Nacional de Investigaciones Científicas y Técnicas; Argentina. Universidad de Buenos Aires. Facultad de Ingeniería; Argentina Fil: Acosta, Eduardo Omar. Universidad de Buenos Aires. Facultad de Ingeniería; Argentina Fil: Santiago, Guillermo. Universidad de Buenos Aires. Facultad de Ingeniería; Argentina |
description |
In this paper we present a method to determine the spatial resolution and sensitivity of systems for optoacoustic tomography (OAT). The method consists of obtaining the image of a sample based on a transparent film embedded in agarose. The film has a certain pattern made with a laser printer that allows to obtain the system spatial resolution. Moreover, since the damage threshold of the ink pattern is similar to that of living tissue, it is also possible to determine if the system is sensitive enough to be applied on biological samples. The method is straightforward, fast and repeatable, and was tested in a OAT system for obtaining two-dimensional images developed in our laboratory. |
publishDate |
2018 |
dc.date.none.fl_str_mv |
2018-12 |
dc.type.none.fl_str_mv |
info:eu-repo/semantics/article info:eu-repo/semantics/publishedVersion http://purl.org/coar/resource_type/c_6501 info:ar-repo/semantics/articulo |
format |
article |
status_str |
publishedVersion |
dc.identifier.none.fl_str_mv |
http://hdl.handle.net/11336/178236 González, Martín Germán; Acosta, Eduardo Omar; Santiago, Guillermo; Simple method to determine the resolution and sensitivity of systems for optoacoustic tomography; Universidad de Buenos Aires. Facultad de Ingeniería; Elektron; 2; 2; 12-2018; 63-66 2525-0159 CONICET Digital CONICET |
url |
http://hdl.handle.net/11336/178236 |
identifier_str_mv |
González, Martín Germán; Acosta, Eduardo Omar; Santiago, Guillermo; Simple method to determine the resolution and sensitivity of systems for optoacoustic tomography; Universidad de Buenos Aires. Facultad de Ingeniería; Elektron; 2; 2; 12-2018; 63-66 2525-0159 CONICET Digital CONICET |
dc.language.none.fl_str_mv |
eng |
language |
eng |
dc.relation.none.fl_str_mv |
info:eu-repo/semantics/altIdentifier/url/http://elektron.fi.uba.ar/index.php/elektron/article/view/47 |
dc.rights.none.fl_str_mv |
info:eu-repo/semantics/openAccess https://creativecommons.org/licenses/by-nc-nd/2.5/ar/ |
eu_rights_str_mv |
openAccess |
rights_invalid_str_mv |
https://creativecommons.org/licenses/by-nc-nd/2.5/ar/ |
dc.format.none.fl_str_mv |
application/pdf application/pdf |
dc.publisher.none.fl_str_mv |
Universidad de Buenos Aires. Facultad de Ingeniería |
publisher.none.fl_str_mv |
Universidad de Buenos Aires. Facultad de Ingeniería |
dc.source.none.fl_str_mv |
reponame:CONICET Digital (CONICET) instname:Consejo Nacional de Investigaciones Científicas y Técnicas |
reponame_str |
CONICET Digital (CONICET) |
collection |
CONICET Digital (CONICET) |
instname_str |
Consejo Nacional de Investigaciones Científicas y Técnicas |
repository.name.fl_str_mv |
CONICET Digital (CONICET) - Consejo Nacional de Investigaciones Científicas y Técnicas |
repository.mail.fl_str_mv |
dasensio@conicet.gov.ar; lcarlino@conicet.gov.ar |
_version_ |
1842269374879105024 |
score |
13.13397 |