Effects on Aortoiliac Fluid Dynamics After Endovascular Sealing of Abdominal Aneurysms

Autores
Casciaro, Mariano Ezequiel; Dottori, Javier Alejandro; El Batti, Salma; Alsac, Jean Marc; Mousseaux, Elie; Larrabide, Ignacio; Craiem, Damian
Año de publicación
2018
Idioma
inglés
Tipo de recurso
artículo
Estado
versión publicada
Descripción
Objectives: To evaluate the effects on aortoiliac fluid dynamics after the implantation of an endograft based on endovascular aneurysm sealing (EVAS) versus endovascular aneurysm repair (EVAR) strategy. Methods: An adaptive geometrical deformable model was used for aortic lumen segmentation in 8 patients before and after the surgery. Abdominal aneurysms were treated with an endograft based on the EVAS system (Nellix, n = 4) and with a device based on an anatomical fixation technology (n = 4). Pressure, blood velocity, and wall shear stress (WSS) were estimated at different aortic regions using computational fluid dynamics methods. Physiologic inlet/outlet flow values at the abdominal aorta, the celiac trunk, and the mesenteric and the renal arteries were set. Pressure references were set at iliac arteries outlet. Results: Maximum aneurysm sizes were similar for both groups in the preoperative scans. The lumen area was lower after EVAR (P <.05) and EVAS (P <.01) compared to preoperative aortic lumen sizes. Pressure increase was higher in the proximal abdominal aorta after EVAS compared to EVAR (2.3 ± 0.3 mm Hg vs 0.9 ± 0.3 mm Hg, P <.001). Peak blood velocities inside the endografts were 3-fold higher for EVAS compared to EVAR (54 ± 5 cm/s vs 17 ± 4 cm/s, P <.01). Velocities at the iliac arteries also remained higher for EVAS (38 ± 4 cm/s vs 24 ± 4 cm/s, P <.05). Peak WSS at the iliac arteries remained higher for EVAS compared to EVAR group (P <.05). Conclusion: The significant modification of the aortic bifurcation anatomy after EVAS alters aortoiliac fluid dynamics, showing a pressure impact at the renal arteries level and an acceleration of the blood velocity at the iliac region with a concomitant increase in peak WSS.
Fil: Casciaro, Mariano Ezequiel. Consejo Nacional de Investigaciones Científicas y Técnicas. Oficina de Coordinación Administrativa Houssay. Instituto de Medicina Traslacional, Trasplante y Bioingeniería. Fundación Favaloro. Instituto de Medicina Traslacional, Trasplante y Bioingeniería; Argentina
Fil: Dottori, Javier Alejandro. Consejo Nacional de Investigaciones Científicas y Técnicas; Argentina. Universidad Nacional del Centro de la Provincia de Buenos Aires. Facultad de Ciencias Exactas. Grupo de Plasmas Densos Magnetizados. Provincia de Buenos Aires. Gobernación. Comision de Investigaciones Científicas. Grupo de Plasmas Densos Magnetizados; Argentina
Fil: El Batti, Salma. Hopital Europeen Georges Pompidou; Francia
Fil: Alsac, Jean Marc. Hopital Europeen Georges Pompidou; Francia
Fil: Mousseaux, Elie. Hopital Europeen Georges Pompidou; Francia
Fil: Larrabide, Ignacio. Universidad Nacional del Centro de la Provincia de Buenos Aires. Facultad de Ciencias Exactas. Grupo de Plasmas Densos Magnetizados. Provincia de Buenos Aires. Gobernación. Comision de Investigaciones Científicas. Grupo de Plasmas Densos Magnetizados; Argentina. Consejo Nacional de Investigaciones Científicas y Técnicas; Argentina
Fil: Craiem, Damian. Hopital Europeen Georges Pompidou; Francia. Consejo Nacional de Investigaciones Científicas y Técnicas. Oficina de Coordinación Administrativa Houssay. Instituto de Medicina Traslacional, Trasplante y Bioingeniería. Fundación Favaloro. Instituto de Medicina Traslacional, Trasplante y Bioingeniería; Argentina
Materia
ABDOMINAL AORTIC ANEURYSM
COMPUTATIONAL FLUID DYNAMICS
ENDOVASCULAR REPAIR
GEOMETRIC DEFORMABLE MODELS
NELLIX ENDOGRAFT
SAC ANCHORING PROSTHESIS
Nivel de accesibilidad
acceso abierto
Condiciones de uso
https://creativecommons.org/licenses/by-nc-sa/2.5/ar/
Repositorio
CONICET Digital (CONICET)
Institución
Consejo Nacional de Investigaciones Científicas y Técnicas
OAI Identificador
oai:ri.conicet.gov.ar:11336/95936

id CONICETDig_5153c92bbb1924979c133deb02a2ea57
oai_identifier_str oai:ri.conicet.gov.ar:11336/95936
network_acronym_str CONICETDig
repository_id_str 3498
network_name_str CONICET Digital (CONICET)
spelling Effects on Aortoiliac Fluid Dynamics After Endovascular Sealing of Abdominal AneurysmsCasciaro, Mariano EzequielDottori, Javier AlejandroEl Batti, SalmaAlsac, Jean MarcMousseaux, ElieLarrabide, IgnacioCraiem, DamianABDOMINAL AORTIC ANEURYSMCOMPUTATIONAL FLUID DYNAMICSENDOVASCULAR REPAIRGEOMETRIC DEFORMABLE MODELSNELLIX ENDOGRAFTSAC ANCHORING PROSTHESIShttps://purl.org/becyt/ford/1.2https://purl.org/becyt/ford/1https://purl.org/becyt/ford/1.2https://purl.org/becyt/ford/1Objectives: To evaluate the effects on aortoiliac fluid dynamics after the implantation of an endograft based on endovascular aneurysm sealing (EVAS) versus endovascular aneurysm repair (EVAR) strategy. Methods: An adaptive geometrical deformable model was used for aortic lumen segmentation in 8 patients before and after the surgery. Abdominal aneurysms were treated with an endograft based on the EVAS system (Nellix, n = 4) and with a device based on an anatomical fixation technology (n = 4). Pressure, blood velocity, and wall shear stress (WSS) were estimated at different aortic regions using computational fluid dynamics methods. Physiologic inlet/outlet flow values at the abdominal aorta, the celiac trunk, and the mesenteric and the renal arteries were set. Pressure references were set at iliac arteries outlet. Results: Maximum aneurysm sizes were similar for both groups in the preoperative scans. The lumen area was lower after EVAR (P <.05) and EVAS (P <.01) compared to preoperative aortic lumen sizes. Pressure increase was higher in the proximal abdominal aorta after EVAS compared to EVAR (2.3 ± 0.3 mm Hg vs 0.9 ± 0.3 mm Hg, P <.001). Peak blood velocities inside the endografts were 3-fold higher for EVAS compared to EVAR (54 ± 5 cm/s vs 17 ± 4 cm/s, P <.01). Velocities at the iliac arteries also remained higher for EVAS (38 ± 4 cm/s vs 24 ± 4 cm/s, P <.05). Peak WSS at the iliac arteries remained higher for EVAS compared to EVAR group (P <.05). Conclusion: The significant modification of the aortic bifurcation anatomy after EVAS alters aortoiliac fluid dynamics, showing a pressure impact at the renal arteries level and an acceleration of the blood velocity at the iliac region with a concomitant increase in peak WSS.Fil: Casciaro, Mariano Ezequiel. Consejo Nacional de Investigaciones Científicas y Técnicas. Oficina de Coordinación Administrativa Houssay. Instituto de Medicina Traslacional, Trasplante y Bioingeniería. Fundación Favaloro. Instituto de Medicina Traslacional, Trasplante y Bioingeniería; ArgentinaFil: Dottori, Javier Alejandro. Consejo Nacional de Investigaciones Científicas y Técnicas; Argentina. Universidad Nacional del Centro de la Provincia de Buenos Aires. Facultad de Ciencias Exactas. Grupo de Plasmas Densos Magnetizados. Provincia de Buenos Aires. Gobernación. Comision de Investigaciones Científicas. Grupo de Plasmas Densos Magnetizados; ArgentinaFil: El Batti, Salma. Hopital Europeen Georges Pompidou; FranciaFil: Alsac, Jean Marc. Hopital Europeen Georges Pompidou; FranciaFil: Mousseaux, Elie. Hopital Europeen Georges Pompidou; FranciaFil: Larrabide, Ignacio. Universidad Nacional del Centro de la Provincia de Buenos Aires. Facultad de Ciencias Exactas. Grupo de Plasmas Densos Magnetizados. Provincia de Buenos Aires. Gobernación. Comision de Investigaciones Científicas. Grupo de Plasmas Densos Magnetizados; Argentina. Consejo Nacional de Investigaciones Científicas y Técnicas; ArgentinaFil: Craiem, Damian. Hopital Europeen Georges Pompidou; Francia. Consejo Nacional de Investigaciones Científicas y Técnicas. Oficina de Coordinación Administrativa Houssay. Instituto de Medicina Traslacional, Trasplante y Bioingeniería. Fundación Favaloro. Instituto de Medicina Traslacional, Trasplante y Bioingeniería; ArgentinaSAGE Publications2018-11info:eu-repo/semantics/articleinfo:eu-repo/semantics/publishedVersionhttp://purl.org/coar/resource_type/c_6501info:ar-repo/semantics/articuloapplication/pdfapplication/pdfapplication/pdfapplication/pdfhttp://hdl.handle.net/11336/95936Casciaro, Mariano Ezequiel; Dottori, Javier Alejandro; El Batti, Salma; Alsac, Jean Marc; Mousseaux, Elie; et al.; Effects on Aortoiliac Fluid Dynamics After Endovascular Sealing of Abdominal Aneurysms; SAGE Publications; Vascular And Endovascular Surgery; 52; 8; 11-2018; 621-6281538-5744CONICET DigitalCONICETenginfo:eu-repo/semantics/altIdentifier/url/http://journals.sagepub.com/doi/10.1177/1538574418791059info:eu-repo/semantics/altIdentifier/doi/10.1177/1538574418791059info:eu-repo/semantics/openAccesshttps://creativecommons.org/licenses/by-nc-sa/2.5/ar/reponame:CONICET Digital (CONICET)instname:Consejo Nacional de Investigaciones Científicas y Técnicas2025-09-29T09:46:34Zoai:ri.conicet.gov.ar:11336/95936instacron:CONICETInstitucionalhttp://ri.conicet.gov.ar/Organismo científico-tecnológicoNo correspondehttp://ri.conicet.gov.ar/oai/requestdasensio@conicet.gov.ar; lcarlino@conicet.gov.arArgentinaNo correspondeNo correspondeNo correspondeopendoar:34982025-09-29 09:46:35.037CONICET Digital (CONICET) - Consejo Nacional de Investigaciones Científicas y Técnicasfalse
dc.title.none.fl_str_mv Effects on Aortoiliac Fluid Dynamics After Endovascular Sealing of Abdominal Aneurysms
title Effects on Aortoiliac Fluid Dynamics After Endovascular Sealing of Abdominal Aneurysms
spellingShingle Effects on Aortoiliac Fluid Dynamics After Endovascular Sealing of Abdominal Aneurysms
Casciaro, Mariano Ezequiel
ABDOMINAL AORTIC ANEURYSM
COMPUTATIONAL FLUID DYNAMICS
ENDOVASCULAR REPAIR
GEOMETRIC DEFORMABLE MODELS
NELLIX ENDOGRAFT
SAC ANCHORING PROSTHESIS
title_short Effects on Aortoiliac Fluid Dynamics After Endovascular Sealing of Abdominal Aneurysms
title_full Effects on Aortoiliac Fluid Dynamics After Endovascular Sealing of Abdominal Aneurysms
title_fullStr Effects on Aortoiliac Fluid Dynamics After Endovascular Sealing of Abdominal Aneurysms
title_full_unstemmed Effects on Aortoiliac Fluid Dynamics After Endovascular Sealing of Abdominal Aneurysms
title_sort Effects on Aortoiliac Fluid Dynamics After Endovascular Sealing of Abdominal Aneurysms
dc.creator.none.fl_str_mv Casciaro, Mariano Ezequiel
Dottori, Javier Alejandro
El Batti, Salma
Alsac, Jean Marc
Mousseaux, Elie
Larrabide, Ignacio
Craiem, Damian
author Casciaro, Mariano Ezequiel
author_facet Casciaro, Mariano Ezequiel
Dottori, Javier Alejandro
El Batti, Salma
Alsac, Jean Marc
Mousseaux, Elie
Larrabide, Ignacio
Craiem, Damian
author_role author
author2 Dottori, Javier Alejandro
El Batti, Salma
Alsac, Jean Marc
Mousseaux, Elie
Larrabide, Ignacio
Craiem, Damian
author2_role author
author
author
author
author
author
dc.subject.none.fl_str_mv ABDOMINAL AORTIC ANEURYSM
COMPUTATIONAL FLUID DYNAMICS
ENDOVASCULAR REPAIR
GEOMETRIC DEFORMABLE MODELS
NELLIX ENDOGRAFT
SAC ANCHORING PROSTHESIS
topic ABDOMINAL AORTIC ANEURYSM
COMPUTATIONAL FLUID DYNAMICS
ENDOVASCULAR REPAIR
GEOMETRIC DEFORMABLE MODELS
NELLIX ENDOGRAFT
SAC ANCHORING PROSTHESIS
purl_subject.fl_str_mv https://purl.org/becyt/ford/1.2
https://purl.org/becyt/ford/1
https://purl.org/becyt/ford/1.2
https://purl.org/becyt/ford/1
dc.description.none.fl_txt_mv Objectives: To evaluate the effects on aortoiliac fluid dynamics after the implantation of an endograft based on endovascular aneurysm sealing (EVAS) versus endovascular aneurysm repair (EVAR) strategy. Methods: An adaptive geometrical deformable model was used for aortic lumen segmentation in 8 patients before and after the surgery. Abdominal aneurysms were treated with an endograft based on the EVAS system (Nellix, n = 4) and with a device based on an anatomical fixation technology (n = 4). Pressure, blood velocity, and wall shear stress (WSS) were estimated at different aortic regions using computational fluid dynamics methods. Physiologic inlet/outlet flow values at the abdominal aorta, the celiac trunk, and the mesenteric and the renal arteries were set. Pressure references were set at iliac arteries outlet. Results: Maximum aneurysm sizes were similar for both groups in the preoperative scans. The lumen area was lower after EVAR (P <.05) and EVAS (P <.01) compared to preoperative aortic lumen sizes. Pressure increase was higher in the proximal abdominal aorta after EVAS compared to EVAR (2.3 ± 0.3 mm Hg vs 0.9 ± 0.3 mm Hg, P <.001). Peak blood velocities inside the endografts were 3-fold higher for EVAS compared to EVAR (54 ± 5 cm/s vs 17 ± 4 cm/s, P <.01). Velocities at the iliac arteries also remained higher for EVAS (38 ± 4 cm/s vs 24 ± 4 cm/s, P <.05). Peak WSS at the iliac arteries remained higher for EVAS compared to EVAR group (P <.05). Conclusion: The significant modification of the aortic bifurcation anatomy after EVAS alters aortoiliac fluid dynamics, showing a pressure impact at the renal arteries level and an acceleration of the blood velocity at the iliac region with a concomitant increase in peak WSS.
Fil: Casciaro, Mariano Ezequiel. Consejo Nacional de Investigaciones Científicas y Técnicas. Oficina de Coordinación Administrativa Houssay. Instituto de Medicina Traslacional, Trasplante y Bioingeniería. Fundación Favaloro. Instituto de Medicina Traslacional, Trasplante y Bioingeniería; Argentina
Fil: Dottori, Javier Alejandro. Consejo Nacional de Investigaciones Científicas y Técnicas; Argentina. Universidad Nacional del Centro de la Provincia de Buenos Aires. Facultad de Ciencias Exactas. Grupo de Plasmas Densos Magnetizados. Provincia de Buenos Aires. Gobernación. Comision de Investigaciones Científicas. Grupo de Plasmas Densos Magnetizados; Argentina
Fil: El Batti, Salma. Hopital Europeen Georges Pompidou; Francia
Fil: Alsac, Jean Marc. Hopital Europeen Georges Pompidou; Francia
Fil: Mousseaux, Elie. Hopital Europeen Georges Pompidou; Francia
Fil: Larrabide, Ignacio. Universidad Nacional del Centro de la Provincia de Buenos Aires. Facultad de Ciencias Exactas. Grupo de Plasmas Densos Magnetizados. Provincia de Buenos Aires. Gobernación. Comision de Investigaciones Científicas. Grupo de Plasmas Densos Magnetizados; Argentina. Consejo Nacional de Investigaciones Científicas y Técnicas; Argentina
Fil: Craiem, Damian. Hopital Europeen Georges Pompidou; Francia. Consejo Nacional de Investigaciones Científicas y Técnicas. Oficina de Coordinación Administrativa Houssay. Instituto de Medicina Traslacional, Trasplante y Bioingeniería. Fundación Favaloro. Instituto de Medicina Traslacional, Trasplante y Bioingeniería; Argentina
description Objectives: To evaluate the effects on aortoiliac fluid dynamics after the implantation of an endograft based on endovascular aneurysm sealing (EVAS) versus endovascular aneurysm repair (EVAR) strategy. Methods: An adaptive geometrical deformable model was used for aortic lumen segmentation in 8 patients before and after the surgery. Abdominal aneurysms were treated with an endograft based on the EVAS system (Nellix, n = 4) and with a device based on an anatomical fixation technology (n = 4). Pressure, blood velocity, and wall shear stress (WSS) were estimated at different aortic regions using computational fluid dynamics methods. Physiologic inlet/outlet flow values at the abdominal aorta, the celiac trunk, and the mesenteric and the renal arteries were set. Pressure references were set at iliac arteries outlet. Results: Maximum aneurysm sizes were similar for both groups in the preoperative scans. The lumen area was lower after EVAR (P <.05) and EVAS (P <.01) compared to preoperative aortic lumen sizes. Pressure increase was higher in the proximal abdominal aorta after EVAS compared to EVAR (2.3 ± 0.3 mm Hg vs 0.9 ± 0.3 mm Hg, P <.001). Peak blood velocities inside the endografts were 3-fold higher for EVAS compared to EVAR (54 ± 5 cm/s vs 17 ± 4 cm/s, P <.01). Velocities at the iliac arteries also remained higher for EVAS (38 ± 4 cm/s vs 24 ± 4 cm/s, P <.05). Peak WSS at the iliac arteries remained higher for EVAS compared to EVAR group (P <.05). Conclusion: The significant modification of the aortic bifurcation anatomy after EVAS alters aortoiliac fluid dynamics, showing a pressure impact at the renal arteries level and an acceleration of the blood velocity at the iliac region with a concomitant increase in peak WSS.
publishDate 2018
dc.date.none.fl_str_mv 2018-11
dc.type.none.fl_str_mv info:eu-repo/semantics/article
info:eu-repo/semantics/publishedVersion
http://purl.org/coar/resource_type/c_6501
info:ar-repo/semantics/articulo
format article
status_str publishedVersion
dc.identifier.none.fl_str_mv http://hdl.handle.net/11336/95936
Casciaro, Mariano Ezequiel; Dottori, Javier Alejandro; El Batti, Salma; Alsac, Jean Marc; Mousseaux, Elie; et al.; Effects on Aortoiliac Fluid Dynamics After Endovascular Sealing of Abdominal Aneurysms; SAGE Publications; Vascular And Endovascular Surgery; 52; 8; 11-2018; 621-628
1538-5744
CONICET Digital
CONICET
url http://hdl.handle.net/11336/95936
identifier_str_mv Casciaro, Mariano Ezequiel; Dottori, Javier Alejandro; El Batti, Salma; Alsac, Jean Marc; Mousseaux, Elie; et al.; Effects on Aortoiliac Fluid Dynamics After Endovascular Sealing of Abdominal Aneurysms; SAGE Publications; Vascular And Endovascular Surgery; 52; 8; 11-2018; 621-628
1538-5744
CONICET Digital
CONICET
dc.language.none.fl_str_mv eng
language eng
dc.relation.none.fl_str_mv info:eu-repo/semantics/altIdentifier/url/http://journals.sagepub.com/doi/10.1177/1538574418791059
info:eu-repo/semantics/altIdentifier/doi/10.1177/1538574418791059
dc.rights.none.fl_str_mv info:eu-repo/semantics/openAccess
https://creativecommons.org/licenses/by-nc-sa/2.5/ar/
eu_rights_str_mv openAccess
rights_invalid_str_mv https://creativecommons.org/licenses/by-nc-sa/2.5/ar/
dc.format.none.fl_str_mv application/pdf
application/pdf
application/pdf
application/pdf
dc.publisher.none.fl_str_mv SAGE Publications
publisher.none.fl_str_mv SAGE Publications
dc.source.none.fl_str_mv reponame:CONICET Digital (CONICET)
instname:Consejo Nacional de Investigaciones Científicas y Técnicas
reponame_str CONICET Digital (CONICET)
collection CONICET Digital (CONICET)
instname_str Consejo Nacional de Investigaciones Científicas y Técnicas
repository.name.fl_str_mv CONICET Digital (CONICET) - Consejo Nacional de Investigaciones Científicas y Técnicas
repository.mail.fl_str_mv dasensio@conicet.gov.ar; lcarlino@conicet.gov.ar
_version_ 1844613454678196224
score 13.070432