Characterization of digital dispersive spectrometers by low coherence interferometry
- Autores
- Martínez Matos, Ó.; Rickenstorff, C.; Zamora, S.; Izquierdo, J. G.; Vaveliuk, Pablo
- Año de publicación
- 2017
- Idioma
- inglés
- Tipo de recurso
- artículo
- Estado
- versión publicada
- Descripción
- We propose a procedure to determine the spectral response of digital dispersive spectrometers without previous knowledge of any parameter of the system. The method consists of applying the Fourier transform spectroscopy technique to each pixel of the detection plane, a CCD camera, to obtain its individual spectral response. From this simple procedure, the system-point spread function and the effect of the finite pixel width are taken into account giving rise to a response matrix that fully characterizes the spectrometer. Using the response matrix information we find the resolving power of a given spectrometer, predict in advance its response to any virtual input spectrum and improve numerically the spectrometer's resolution. We consider that the presented approach could be useful in most spectroscopic branches such as in computational spectroscopy, optical coherence tomography, hyperspectral imaging, spectral interferometry and analytical chemistry, among others.
Fil: Martínez Matos, Ó.. Universidad Complutense de Madrid; España
Fil: Rickenstorff, C.. Universidad Complutense de Madrid; España
Fil: Zamora, S.. Universidad Complutense de Madrid; España
Fil: Izquierdo, J. G.. Universidad Complutense de Madrid; España
Fil: Vaveliuk, Pablo. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - La Plata. Centro de Investigaciones Ópticas. Provincia de Buenos Aires. Gobernación. Comisión de Investigaciones Científicas. Centro de Investigaciones Ópticas. Universidad Nacional de La Plata. Centro de Investigaciones Ópticas; Argentina - Materia
-
Spectrometers
Interferometry - Nivel de accesibilidad
- acceso abierto
- Condiciones de uso
- https://creativecommons.org/licenses/by/2.5/ar/
- Repositorio
- Institución
- Consejo Nacional de Investigaciones Científicas y Técnicas
- OAI Identificador
- oai:ri.conicet.gov.ar:11336/49266
Ver los metadatos del registro completo
id |
CONICETDig_3eac223200bc62e94e5f34d998005e75 |
---|---|
oai_identifier_str |
oai:ri.conicet.gov.ar:11336/49266 |
network_acronym_str |
CONICETDig |
repository_id_str |
3498 |
network_name_str |
CONICET Digital (CONICET) |
spelling |
Characterization of digital dispersive spectrometers by low coherence interferometryMartínez Matos, Ó.Rickenstorff, C.Zamora, S.Izquierdo, J. G.Vaveliuk, PabloSpectrometersInterferometryhttps://purl.org/becyt/ford/1.3https://purl.org/becyt/ford/1We propose a procedure to determine the spectral response of digital dispersive spectrometers without previous knowledge of any parameter of the system. The method consists of applying the Fourier transform spectroscopy technique to each pixel of the detection plane, a CCD camera, to obtain its individual spectral response. From this simple procedure, the system-point spread function and the effect of the finite pixel width are taken into account giving rise to a response matrix that fully characterizes the spectrometer. Using the response matrix information we find the resolving power of a given spectrometer, predict in advance its response to any virtual input spectrum and improve numerically the spectrometer's resolution. We consider that the presented approach could be useful in most spectroscopic branches such as in computational spectroscopy, optical coherence tomography, hyperspectral imaging, spectral interferometry and analytical chemistry, among others.Fil: Martínez Matos, Ó.. Universidad Complutense de Madrid; EspañaFil: Rickenstorff, C.. Universidad Complutense de Madrid; EspañaFil: Zamora, S.. Universidad Complutense de Madrid; EspañaFil: Izquierdo, J. G.. Universidad Complutense de Madrid; EspañaFil: Vaveliuk, Pablo. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - La Plata. Centro de Investigaciones Ópticas. Provincia de Buenos Aires. Gobernación. Comisión de Investigaciones Científicas. Centro de Investigaciones Ópticas. Universidad Nacional de La Plata. Centro de Investigaciones Ópticas; ArgentinaOptical Society of America2017-02info:eu-repo/semantics/articleinfo:eu-repo/semantics/publishedVersionhttp://purl.org/coar/resource_type/c_6501info:ar-repo/semantics/articuloapplication/pdfapplication/pdfhttp://hdl.handle.net/11336/49266Martínez Matos, Ó.; Rickenstorff, C.; Zamora, S.; Izquierdo, J. G.; Vaveliuk, Pablo; Characterization of digital dispersive spectrometers by low coherence interferometry; Optical Society of America; Optics Express; 25; 4; 2-2017; 3222-32331094-4087CONICET DigitalCONICETenginfo:eu-repo/semantics/altIdentifier/doi/10.1364/OE.25.003222info:eu-repo/semantics/altIdentifier/url/https://www.osapublishing.org/oe/abstract.cfm?uri=oe-25-4-3222info:eu-repo/semantics/openAccesshttps://creativecommons.org/licenses/by/2.5/ar/reponame:CONICET Digital (CONICET)instname:Consejo Nacional de Investigaciones Científicas y Técnicas2025-09-03T10:02:52Zoai:ri.conicet.gov.ar:11336/49266instacron:CONICETInstitucionalhttp://ri.conicet.gov.ar/Organismo científico-tecnológicoNo correspondehttp://ri.conicet.gov.ar/oai/requestdasensio@conicet.gov.ar; lcarlino@conicet.gov.arArgentinaNo correspondeNo correspondeNo correspondeopendoar:34982025-09-03 10:02:52.958CONICET Digital (CONICET) - Consejo Nacional de Investigaciones Científicas y Técnicasfalse |
dc.title.none.fl_str_mv |
Characterization of digital dispersive spectrometers by low coherence interferometry |
title |
Characterization of digital dispersive spectrometers by low coherence interferometry |
spellingShingle |
Characterization of digital dispersive spectrometers by low coherence interferometry Martínez Matos, Ó. Spectrometers Interferometry |
title_short |
Characterization of digital dispersive spectrometers by low coherence interferometry |
title_full |
Characterization of digital dispersive spectrometers by low coherence interferometry |
title_fullStr |
Characterization of digital dispersive spectrometers by low coherence interferometry |
title_full_unstemmed |
Characterization of digital dispersive spectrometers by low coherence interferometry |
title_sort |
Characterization of digital dispersive spectrometers by low coherence interferometry |
dc.creator.none.fl_str_mv |
Martínez Matos, Ó. Rickenstorff, C. Zamora, S. Izquierdo, J. G. Vaveliuk, Pablo |
author |
Martínez Matos, Ó. |
author_facet |
Martínez Matos, Ó. Rickenstorff, C. Zamora, S. Izquierdo, J. G. Vaveliuk, Pablo |
author_role |
author |
author2 |
Rickenstorff, C. Zamora, S. Izquierdo, J. G. Vaveliuk, Pablo |
author2_role |
author author author author |
dc.subject.none.fl_str_mv |
Spectrometers Interferometry |
topic |
Spectrometers Interferometry |
purl_subject.fl_str_mv |
https://purl.org/becyt/ford/1.3 https://purl.org/becyt/ford/1 |
dc.description.none.fl_txt_mv |
We propose a procedure to determine the spectral response of digital dispersive spectrometers without previous knowledge of any parameter of the system. The method consists of applying the Fourier transform spectroscopy technique to each pixel of the detection plane, a CCD camera, to obtain its individual spectral response. From this simple procedure, the system-point spread function and the effect of the finite pixel width are taken into account giving rise to a response matrix that fully characterizes the spectrometer. Using the response matrix information we find the resolving power of a given spectrometer, predict in advance its response to any virtual input spectrum and improve numerically the spectrometer's resolution. We consider that the presented approach could be useful in most spectroscopic branches such as in computational spectroscopy, optical coherence tomography, hyperspectral imaging, spectral interferometry and analytical chemistry, among others. Fil: Martínez Matos, Ó.. Universidad Complutense de Madrid; España Fil: Rickenstorff, C.. Universidad Complutense de Madrid; España Fil: Zamora, S.. Universidad Complutense de Madrid; España Fil: Izquierdo, J. G.. Universidad Complutense de Madrid; España Fil: Vaveliuk, Pablo. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - La Plata. Centro de Investigaciones Ópticas. Provincia de Buenos Aires. Gobernación. Comisión de Investigaciones Científicas. Centro de Investigaciones Ópticas. Universidad Nacional de La Plata. Centro de Investigaciones Ópticas; Argentina |
description |
We propose a procedure to determine the spectral response of digital dispersive spectrometers without previous knowledge of any parameter of the system. The method consists of applying the Fourier transform spectroscopy technique to each pixel of the detection plane, a CCD camera, to obtain its individual spectral response. From this simple procedure, the system-point spread function and the effect of the finite pixel width are taken into account giving rise to a response matrix that fully characterizes the spectrometer. Using the response matrix information we find the resolving power of a given spectrometer, predict in advance its response to any virtual input spectrum and improve numerically the spectrometer's resolution. We consider that the presented approach could be useful in most spectroscopic branches such as in computational spectroscopy, optical coherence tomography, hyperspectral imaging, spectral interferometry and analytical chemistry, among others. |
publishDate |
2017 |
dc.date.none.fl_str_mv |
2017-02 |
dc.type.none.fl_str_mv |
info:eu-repo/semantics/article info:eu-repo/semantics/publishedVersion http://purl.org/coar/resource_type/c_6501 info:ar-repo/semantics/articulo |
format |
article |
status_str |
publishedVersion |
dc.identifier.none.fl_str_mv |
http://hdl.handle.net/11336/49266 Martínez Matos, Ó.; Rickenstorff, C.; Zamora, S.; Izquierdo, J. G.; Vaveliuk, Pablo; Characterization of digital dispersive spectrometers by low coherence interferometry; Optical Society of America; Optics Express; 25; 4; 2-2017; 3222-3233 1094-4087 CONICET Digital CONICET |
url |
http://hdl.handle.net/11336/49266 |
identifier_str_mv |
Martínez Matos, Ó.; Rickenstorff, C.; Zamora, S.; Izquierdo, J. G.; Vaveliuk, Pablo; Characterization of digital dispersive spectrometers by low coherence interferometry; Optical Society of America; Optics Express; 25; 4; 2-2017; 3222-3233 1094-4087 CONICET Digital CONICET |
dc.language.none.fl_str_mv |
eng |
language |
eng |
dc.relation.none.fl_str_mv |
info:eu-repo/semantics/altIdentifier/doi/10.1364/OE.25.003222 info:eu-repo/semantics/altIdentifier/url/https://www.osapublishing.org/oe/abstract.cfm?uri=oe-25-4-3222 |
dc.rights.none.fl_str_mv |
info:eu-repo/semantics/openAccess https://creativecommons.org/licenses/by/2.5/ar/ |
eu_rights_str_mv |
openAccess |
rights_invalid_str_mv |
https://creativecommons.org/licenses/by/2.5/ar/ |
dc.format.none.fl_str_mv |
application/pdf application/pdf |
dc.publisher.none.fl_str_mv |
Optical Society of America |
publisher.none.fl_str_mv |
Optical Society of America |
dc.source.none.fl_str_mv |
reponame:CONICET Digital (CONICET) instname:Consejo Nacional de Investigaciones Científicas y Técnicas |
reponame_str |
CONICET Digital (CONICET) |
collection |
CONICET Digital (CONICET) |
instname_str |
Consejo Nacional de Investigaciones Científicas y Técnicas |
repository.name.fl_str_mv |
CONICET Digital (CONICET) - Consejo Nacional de Investigaciones Científicas y Técnicas |
repository.mail.fl_str_mv |
dasensio@conicet.gov.ar; lcarlino@conicet.gov.ar |
_version_ |
1842269775175090176 |
score |
13.13397 |