Characterization of a fiberoptic radiotherapy dosimetry probe based on Mg2SiO4:Tb
- Autores
- Molina, Pablo Fernando; Prokic, M.; Marcazzo, Salvador Julian; Santiago, Martin Alejo
- Año de publicación
- 2010
- Idioma
- inglés
- Tipo de recurso
- artículo
- Estado
- versión publicada
- Descripción
- In this work the feasibility of using Mg2SiO4:Tb as a fiberoptic radioluminescent (RL) dosimetry probe for real-time dosimetry has been investigated for the first time. In particular, the stability of the RL signal after repeated use, the spectrum of the RL emission and the dose-rate response curve of a Mg2SiO4:Tb-based fiberoptic probe have been determined. The probe has been also used to obtain a percentage dose depth curve in a water phantom and its performance has been compared to that of a standard ion chamber. Besides, its absolute RL yield has been compared to that of an RL probe based on the commercial Al2O3:C phosphor.
Fil: Molina, Pablo Fernando. Consejo Nacional de Investigaciones Científicas y Técnicas; Argentina. Universidad Nacional del Centro de la Provincia de Buenos Aires. Facultad de Ciencias Exactas. Instituto de Fisica Arroyo Seco; Argentina
Fil: Prokic, M.. Institute of Nuclear Sciences; Serbia
Fil: Marcazzo, Salvador Julian. Universidad Nacional del Centro de la Provincia de Buenos Aires. Facultad de Ciencias Exactas. Instituto de Fisica Arroyo Seco; Argentina. Consejo Nacional de Investigaciones Científicas y Técnicas; Argentina
Fil: Santiago, Martin Alejo. Consejo Nacional de Investigaciones Científicas y Técnicas; Argentina. Universidad Nacional del Centro de la Provincia de Buenos Aires. Facultad de Ciencias Exactas. Instituto de Fisica Arroyo Seco; Argentina - Materia
-
DOSIMETRY
MAGNESIUM ORTHOSILICATE
RADIOLUMINESCENCE - Nivel de accesibilidad
- acceso abierto
- Condiciones de uso
- https://creativecommons.org/licenses/by-nc-nd/2.5/ar/
- Repositorio
- Institución
- Consejo Nacional de Investigaciones Científicas y Técnicas
- OAI Identificador
- oai:ri.conicet.gov.ar:11336/96033
Ver los metadatos del registro completo
id |
CONICETDig_0bd183b0f76e386fe6f001b3348ae2f2 |
---|---|
oai_identifier_str |
oai:ri.conicet.gov.ar:11336/96033 |
network_acronym_str |
CONICETDig |
repository_id_str |
3498 |
network_name_str |
CONICET Digital (CONICET) |
spelling |
Characterization of a fiberoptic radiotherapy dosimetry probe based on Mg2SiO4:TbMolina, Pablo FernandoProkic, M.Marcazzo, Salvador JulianSantiago, Martin AlejoDOSIMETRYMAGNESIUM ORTHOSILICATERADIOLUMINESCENCEhttps://purl.org/becyt/ford/1.3https://purl.org/becyt/ford/1In this work the feasibility of using Mg2SiO4:Tb as a fiberoptic radioluminescent (RL) dosimetry probe for real-time dosimetry has been investigated for the first time. In particular, the stability of the RL signal after repeated use, the spectrum of the RL emission and the dose-rate response curve of a Mg2SiO4:Tb-based fiberoptic probe have been determined. The probe has been also used to obtain a percentage dose depth curve in a water phantom and its performance has been compared to that of a standard ion chamber. Besides, its absolute RL yield has been compared to that of an RL probe based on the commercial Al2O3:C phosphor.Fil: Molina, Pablo Fernando. Consejo Nacional de Investigaciones Científicas y Técnicas; Argentina. Universidad Nacional del Centro de la Provincia de Buenos Aires. Facultad de Ciencias Exactas. Instituto de Fisica Arroyo Seco; ArgentinaFil: Prokic, M.. Institute of Nuclear Sciences; SerbiaFil: Marcazzo, Salvador Julian. Universidad Nacional del Centro de la Provincia de Buenos Aires. Facultad de Ciencias Exactas. Instituto de Fisica Arroyo Seco; Argentina. Consejo Nacional de Investigaciones Científicas y Técnicas; ArgentinaFil: Santiago, Martin Alejo. Consejo Nacional de Investigaciones Científicas y Técnicas; Argentina. Universidad Nacional del Centro de la Provincia de Buenos Aires. Facultad de Ciencias Exactas. Instituto de Fisica Arroyo Seco; ArgentinaPergamon-Elsevier Science Ltd2010-01info:eu-repo/semantics/articleinfo:eu-repo/semantics/publishedVersionhttp://purl.org/coar/resource_type/c_6501info:ar-repo/semantics/articuloapplication/pdfapplication/pdfapplication/pdfapplication/pdfhttp://hdl.handle.net/11336/96033Molina, Pablo Fernando; Prokic, M.; Marcazzo, Salvador Julian; Santiago, Martin Alejo; Characterization of a fiberoptic radiotherapy dosimetry probe based on Mg2SiO4:Tb; Pergamon-Elsevier Science Ltd; Radiation Measurements; 45; 1; 1-2010; 78-821350-4487CONICET DigitalCONICETenginfo:eu-repo/semantics/altIdentifier/doi/10.1016/j.radmeas.2009.11.040info:eu-repo/semantics/altIdentifier/url/https://www.sciencedirect.com/science/article/pii/S1350448709003588info:eu-repo/semantics/openAccesshttps://creativecommons.org/licenses/by-nc-nd/2.5/ar/reponame:CONICET Digital (CONICET)instname:Consejo Nacional de Investigaciones Científicas y Técnicas2025-09-29T10:09:47Zoai:ri.conicet.gov.ar:11336/96033instacron:CONICETInstitucionalhttp://ri.conicet.gov.ar/Organismo científico-tecnológicoNo correspondehttp://ri.conicet.gov.ar/oai/requestdasensio@conicet.gov.ar; lcarlino@conicet.gov.arArgentinaNo correspondeNo correspondeNo correspondeopendoar:34982025-09-29 10:09:48.197CONICET Digital (CONICET) - Consejo Nacional de Investigaciones Científicas y Técnicasfalse |
dc.title.none.fl_str_mv |
Characterization of a fiberoptic radiotherapy dosimetry probe based on Mg2SiO4:Tb |
title |
Characterization of a fiberoptic radiotherapy dosimetry probe based on Mg2SiO4:Tb |
spellingShingle |
Characterization of a fiberoptic radiotherapy dosimetry probe based on Mg2SiO4:Tb Molina, Pablo Fernando DOSIMETRY MAGNESIUM ORTHOSILICATE RADIOLUMINESCENCE |
title_short |
Characterization of a fiberoptic radiotherapy dosimetry probe based on Mg2SiO4:Tb |
title_full |
Characterization of a fiberoptic radiotherapy dosimetry probe based on Mg2SiO4:Tb |
title_fullStr |
Characterization of a fiberoptic radiotherapy dosimetry probe based on Mg2SiO4:Tb |
title_full_unstemmed |
Characterization of a fiberoptic radiotherapy dosimetry probe based on Mg2SiO4:Tb |
title_sort |
Characterization of a fiberoptic radiotherapy dosimetry probe based on Mg2SiO4:Tb |
dc.creator.none.fl_str_mv |
Molina, Pablo Fernando Prokic, M. Marcazzo, Salvador Julian Santiago, Martin Alejo |
author |
Molina, Pablo Fernando |
author_facet |
Molina, Pablo Fernando Prokic, M. Marcazzo, Salvador Julian Santiago, Martin Alejo |
author_role |
author |
author2 |
Prokic, M. Marcazzo, Salvador Julian Santiago, Martin Alejo |
author2_role |
author author author |
dc.subject.none.fl_str_mv |
DOSIMETRY MAGNESIUM ORTHOSILICATE RADIOLUMINESCENCE |
topic |
DOSIMETRY MAGNESIUM ORTHOSILICATE RADIOLUMINESCENCE |
purl_subject.fl_str_mv |
https://purl.org/becyt/ford/1.3 https://purl.org/becyt/ford/1 |
dc.description.none.fl_txt_mv |
In this work the feasibility of using Mg2SiO4:Tb as a fiberoptic radioluminescent (RL) dosimetry probe for real-time dosimetry has been investigated for the first time. In particular, the stability of the RL signal after repeated use, the spectrum of the RL emission and the dose-rate response curve of a Mg2SiO4:Tb-based fiberoptic probe have been determined. The probe has been also used to obtain a percentage dose depth curve in a water phantom and its performance has been compared to that of a standard ion chamber. Besides, its absolute RL yield has been compared to that of an RL probe based on the commercial Al2O3:C phosphor. Fil: Molina, Pablo Fernando. Consejo Nacional de Investigaciones Científicas y Técnicas; Argentina. Universidad Nacional del Centro de la Provincia de Buenos Aires. Facultad de Ciencias Exactas. Instituto de Fisica Arroyo Seco; Argentina Fil: Prokic, M.. Institute of Nuclear Sciences; Serbia Fil: Marcazzo, Salvador Julian. Universidad Nacional del Centro de la Provincia de Buenos Aires. Facultad de Ciencias Exactas. Instituto de Fisica Arroyo Seco; Argentina. Consejo Nacional de Investigaciones Científicas y Técnicas; Argentina Fil: Santiago, Martin Alejo. Consejo Nacional de Investigaciones Científicas y Técnicas; Argentina. Universidad Nacional del Centro de la Provincia de Buenos Aires. Facultad de Ciencias Exactas. Instituto de Fisica Arroyo Seco; Argentina |
description |
In this work the feasibility of using Mg2SiO4:Tb as a fiberoptic radioluminescent (RL) dosimetry probe for real-time dosimetry has been investigated for the first time. In particular, the stability of the RL signal after repeated use, the spectrum of the RL emission and the dose-rate response curve of a Mg2SiO4:Tb-based fiberoptic probe have been determined. The probe has been also used to obtain a percentage dose depth curve in a water phantom and its performance has been compared to that of a standard ion chamber. Besides, its absolute RL yield has been compared to that of an RL probe based on the commercial Al2O3:C phosphor. |
publishDate |
2010 |
dc.date.none.fl_str_mv |
2010-01 |
dc.type.none.fl_str_mv |
info:eu-repo/semantics/article info:eu-repo/semantics/publishedVersion http://purl.org/coar/resource_type/c_6501 info:ar-repo/semantics/articulo |
format |
article |
status_str |
publishedVersion |
dc.identifier.none.fl_str_mv |
http://hdl.handle.net/11336/96033 Molina, Pablo Fernando; Prokic, M.; Marcazzo, Salvador Julian; Santiago, Martin Alejo; Characterization of a fiberoptic radiotherapy dosimetry probe based on Mg2SiO4:Tb; Pergamon-Elsevier Science Ltd; Radiation Measurements; 45; 1; 1-2010; 78-82 1350-4487 CONICET Digital CONICET |
url |
http://hdl.handle.net/11336/96033 |
identifier_str_mv |
Molina, Pablo Fernando; Prokic, M.; Marcazzo, Salvador Julian; Santiago, Martin Alejo; Characterization of a fiberoptic radiotherapy dosimetry probe based on Mg2SiO4:Tb; Pergamon-Elsevier Science Ltd; Radiation Measurements; 45; 1; 1-2010; 78-82 1350-4487 CONICET Digital CONICET |
dc.language.none.fl_str_mv |
eng |
language |
eng |
dc.relation.none.fl_str_mv |
info:eu-repo/semantics/altIdentifier/doi/10.1016/j.radmeas.2009.11.040 info:eu-repo/semantics/altIdentifier/url/https://www.sciencedirect.com/science/article/pii/S1350448709003588 |
dc.rights.none.fl_str_mv |
info:eu-repo/semantics/openAccess https://creativecommons.org/licenses/by-nc-nd/2.5/ar/ |
eu_rights_str_mv |
openAccess |
rights_invalid_str_mv |
https://creativecommons.org/licenses/by-nc-nd/2.5/ar/ |
dc.format.none.fl_str_mv |
application/pdf application/pdf application/pdf application/pdf |
dc.publisher.none.fl_str_mv |
Pergamon-Elsevier Science Ltd |
publisher.none.fl_str_mv |
Pergamon-Elsevier Science Ltd |
dc.source.none.fl_str_mv |
reponame:CONICET Digital (CONICET) instname:Consejo Nacional de Investigaciones Científicas y Técnicas |
reponame_str |
CONICET Digital (CONICET) |
collection |
CONICET Digital (CONICET) |
instname_str |
Consejo Nacional de Investigaciones Científicas y Técnicas |
repository.name.fl_str_mv |
CONICET Digital (CONICET) - Consejo Nacional de Investigaciones Científicas y Técnicas |
repository.mail.fl_str_mv |
dasensio@conicet.gov.ar; lcarlino@conicet.gov.ar |
_version_ |
1844613980070346752 |
score |
13.070432 |