Flow-batch analysis of clenbuterol based on analyte extraction on molecularly imprinted polymers coupled to an in-system chromogenic reaction. Application to human urine and milk s...
- Autores
- Gonzalez, Natalia; Grunhut, Marcos; Srámková, Ivana; Lista, Adriana Guillermina; Horstkotte, Burkhard; Solich, Petr; Sklenárová, Hana; Acebal, Carolina Cecilia
- Año de publicación
- 2018
- Idioma
- inglés
- Tipo de recurso
- artículo
- Estado
- versión publicada
- Descripción
- A fully automated spectrophotometric method based on flow-batch analysis has been developed for the determination of clenbuterol including an on-line solid phase extraction using a molecularly imprinted polymer (MIP) as the sorbent. The molecularly imprinted solid phase extraction (MISPE) procedure allowed analyte extraction from complex matrices at low concentration levels and with high selectivity towards the analyte. The MISPE procedure was performed using a commercial MIP cartridge that was introduced into a guard column holder and integrated in the analyzer system. Optimized parameters included the volume of the sample, the type and volume of the conditioning and washing solutions, and the type and volume of the eluent. Quantification of clenbuterol was carried out by spectrophotometry after in-system post-elution analyte derivatization based on azo-coupling using N- (1-Naphthyl) ethylenediamine as the coupling agent to yield a red-colored compound with maximum absorbance at 500 nm. Both the chromogenic reaction and spectrophotometric detection were performed in a lab-made flow-batch mixing chamber that replaced the cuvette holder of the spectrophotometer. The calibration curve was linear in the 0.075–0.500 mg L−1 range with a correlation coefficient of 0.998. The precision of the proposed method was evaluated in terms of the relative standard deviation obtaining 1.1% and 3.0% for intra-day precision and inter-day precision, respectively. The detection limit was 0.021 mg L−1 and the sample throughput for the entire process was 3.4 h−1. The proposed method was applied for the determination of CLB in human urine and milk substitute samples obtaining recoveries values within a range of 94.0–100.0%.
Fil: Gonzalez, Natalia. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - Bahía Blanca. Instituto de Química del Sur. Universidad Nacional del Sur. Departamento de Química. Instituto de Química del Sur; Argentina
Fil: Grunhut, Marcos. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - Bahía Blanca. Instituto de Química del Sur. Universidad Nacional del Sur. Departamento de Química. Instituto de Química del Sur; Argentina
Fil: Srámková, Ivana. Charles University; República Checa
Fil: Lista, Adriana Guillermina. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - Bahía Blanca. Instituto de Química del Sur. Universidad Nacional del Sur. Departamento de Química. Instituto de Química del Sur; Argentina
Fil: Horstkotte, Burkhard. Charles University; República Checa
Fil: Solich, Petr. Charles University; República Checa
Fil: Sklenárová, Hana. Charles University; República Checa
Fil: Acebal, Carolina Cecilia. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - Bahía Blanca. Instituto de Química del Sur. Universidad Nacional del Sur. Departamento de Química. Instituto de Química del Sur; Argentina - Materia
-
CLENBUTEROL
FLOW-BATCH ANALYSIS
MOLECULARLY IMPRINTED POLYMER
SOLID PHASE EXTRACTION
DIAZOTIZATION-COUPLING REACTION - Nivel de accesibilidad
- acceso abierto
- Condiciones de uso
- https://creativecommons.org/licenses/by-nc-sa/2.5/ar/
- Repositorio
- Institución
- Consejo Nacional de Investigaciones Científicas y Técnicas
- OAI Identificador
- oai:ri.conicet.gov.ar:11336/91904
Ver los metadatos del registro completo
id |
CONICETDig_02dda0be70f671bffadf2c16cbdc353d |
---|---|
oai_identifier_str |
oai:ri.conicet.gov.ar:11336/91904 |
network_acronym_str |
CONICETDig |
repository_id_str |
3498 |
network_name_str |
CONICET Digital (CONICET) |
spelling |
Flow-batch analysis of clenbuterol based on analyte extraction on molecularly imprinted polymers coupled to an in-system chromogenic reaction. Application to human urine and milk substitute samplesGonzalez, NataliaGrunhut, MarcosSrámková, IvanaLista, Adriana GuillerminaHorstkotte, BurkhardSolich, PetrSklenárová, HanaAcebal, Carolina CeciliaCLENBUTEROLFLOW-BATCH ANALYSISMOLECULARLY IMPRINTED POLYMERSOLID PHASE EXTRACTIONDIAZOTIZATION-COUPLING REACTIONhttps://purl.org/becyt/ford/1.4https://purl.org/becyt/ford/1A fully automated spectrophotometric method based on flow-batch analysis has been developed for the determination of clenbuterol including an on-line solid phase extraction using a molecularly imprinted polymer (MIP) as the sorbent. The molecularly imprinted solid phase extraction (MISPE) procedure allowed analyte extraction from complex matrices at low concentration levels and with high selectivity towards the analyte. The MISPE procedure was performed using a commercial MIP cartridge that was introduced into a guard column holder and integrated in the analyzer system. Optimized parameters included the volume of the sample, the type and volume of the conditioning and washing solutions, and the type and volume of the eluent. Quantification of clenbuterol was carried out by spectrophotometry after in-system post-elution analyte derivatization based on azo-coupling using N- (1-Naphthyl) ethylenediamine as the coupling agent to yield a red-colored compound with maximum absorbance at 500 nm. Both the chromogenic reaction and spectrophotometric detection were performed in a lab-made flow-batch mixing chamber that replaced the cuvette holder of the spectrophotometer. The calibration curve was linear in the 0.075–0.500 mg L−1 range with a correlation coefficient of 0.998. The precision of the proposed method was evaluated in terms of the relative standard deviation obtaining 1.1% and 3.0% for intra-day precision and inter-day precision, respectively. The detection limit was 0.021 mg L−1 and the sample throughput for the entire process was 3.4 h−1. The proposed method was applied for the determination of CLB in human urine and milk substitute samples obtaining recoveries values within a range of 94.0–100.0%.Fil: Gonzalez, Natalia. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - Bahía Blanca. Instituto de Química del Sur. Universidad Nacional del Sur. Departamento de Química. Instituto de Química del Sur; ArgentinaFil: Grunhut, Marcos. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - Bahía Blanca. Instituto de Química del Sur. Universidad Nacional del Sur. Departamento de Química. Instituto de Química del Sur; ArgentinaFil: Srámková, Ivana. Charles University; República ChecaFil: Lista, Adriana Guillermina. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - Bahía Blanca. Instituto de Química del Sur. Universidad Nacional del Sur. Departamento de Química. Instituto de Química del Sur; ArgentinaFil: Horstkotte, Burkhard. Charles University; República ChecaFil: Solich, Petr. Charles University; República ChecaFil: Sklenárová, Hana. Charles University; República ChecaFil: Acebal, Carolina Cecilia. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - Bahía Blanca. Instituto de Química del Sur. Universidad Nacional del Sur. Departamento de Química. Instituto de Química del Sur; ArgentinaElsevier Science2018-02-01info:eu-repo/semantics/articleinfo:eu-repo/semantics/publishedVersionhttp://purl.org/coar/resource_type/c_6501info:ar-repo/semantics/articuloapplication/pdfapplication/pdfapplication/pdfapplication/pdfapplication/pdfhttp://hdl.handle.net/11336/91904Gonzalez, Natalia; Grunhut, Marcos; Srámková, Ivana; Lista, Adriana Guillermina; Horstkotte, Burkhard; et al.; Flow-batch analysis of clenbuterol based on analyte extraction on molecularly imprinted polymers coupled to an in-system chromogenic reaction. Application to human urine and milk substitute samples; Elsevier Science; Talanta; 178; 1-2-2018; 934-9420039-9140CONICET DigitalCONICETenginfo:eu-repo/semantics/altIdentifier/url/https://www.sciencedirect.com/science/article/pii/S0039914017310792info:eu-repo/semantics/altIdentifier/doi/10.1016/j.talanta.2017.10.040info:eu-repo/semantics/openAccesshttps://creativecommons.org/licenses/by-nc-sa/2.5/ar/reponame:CONICET Digital (CONICET)instname:Consejo Nacional de Investigaciones Científicas y Técnicas2025-09-29T09:34:43Zoai:ri.conicet.gov.ar:11336/91904instacron:CONICETInstitucionalhttp://ri.conicet.gov.ar/Organismo científico-tecnológicoNo correspondehttp://ri.conicet.gov.ar/oai/requestdasensio@conicet.gov.ar; lcarlino@conicet.gov.arArgentinaNo correspondeNo correspondeNo correspondeopendoar:34982025-09-29 09:34:43.812CONICET Digital (CONICET) - Consejo Nacional de Investigaciones Científicas y Técnicasfalse |
dc.title.none.fl_str_mv |
Flow-batch analysis of clenbuterol based on analyte extraction on molecularly imprinted polymers coupled to an in-system chromogenic reaction. Application to human urine and milk substitute samples |
title |
Flow-batch analysis of clenbuterol based on analyte extraction on molecularly imprinted polymers coupled to an in-system chromogenic reaction. Application to human urine and milk substitute samples |
spellingShingle |
Flow-batch analysis of clenbuterol based on analyte extraction on molecularly imprinted polymers coupled to an in-system chromogenic reaction. Application to human urine and milk substitute samples Gonzalez, Natalia CLENBUTEROL FLOW-BATCH ANALYSIS MOLECULARLY IMPRINTED POLYMER SOLID PHASE EXTRACTION DIAZOTIZATION-COUPLING REACTION |
title_short |
Flow-batch analysis of clenbuterol based on analyte extraction on molecularly imprinted polymers coupled to an in-system chromogenic reaction. Application to human urine and milk substitute samples |
title_full |
Flow-batch analysis of clenbuterol based on analyte extraction on molecularly imprinted polymers coupled to an in-system chromogenic reaction. Application to human urine and milk substitute samples |
title_fullStr |
Flow-batch analysis of clenbuterol based on analyte extraction on molecularly imprinted polymers coupled to an in-system chromogenic reaction. Application to human urine and milk substitute samples |
title_full_unstemmed |
Flow-batch analysis of clenbuterol based on analyte extraction on molecularly imprinted polymers coupled to an in-system chromogenic reaction. Application to human urine and milk substitute samples |
title_sort |
Flow-batch analysis of clenbuterol based on analyte extraction on molecularly imprinted polymers coupled to an in-system chromogenic reaction. Application to human urine and milk substitute samples |
dc.creator.none.fl_str_mv |
Gonzalez, Natalia Grunhut, Marcos Srámková, Ivana Lista, Adriana Guillermina Horstkotte, Burkhard Solich, Petr Sklenárová, Hana Acebal, Carolina Cecilia |
author |
Gonzalez, Natalia |
author_facet |
Gonzalez, Natalia Grunhut, Marcos Srámková, Ivana Lista, Adriana Guillermina Horstkotte, Burkhard Solich, Petr Sklenárová, Hana Acebal, Carolina Cecilia |
author_role |
author |
author2 |
Grunhut, Marcos Srámková, Ivana Lista, Adriana Guillermina Horstkotte, Burkhard Solich, Petr Sklenárová, Hana Acebal, Carolina Cecilia |
author2_role |
author author author author author author author |
dc.subject.none.fl_str_mv |
CLENBUTEROL FLOW-BATCH ANALYSIS MOLECULARLY IMPRINTED POLYMER SOLID PHASE EXTRACTION DIAZOTIZATION-COUPLING REACTION |
topic |
CLENBUTEROL FLOW-BATCH ANALYSIS MOLECULARLY IMPRINTED POLYMER SOLID PHASE EXTRACTION DIAZOTIZATION-COUPLING REACTION |
purl_subject.fl_str_mv |
https://purl.org/becyt/ford/1.4 https://purl.org/becyt/ford/1 |
dc.description.none.fl_txt_mv |
A fully automated spectrophotometric method based on flow-batch analysis has been developed for the determination of clenbuterol including an on-line solid phase extraction using a molecularly imprinted polymer (MIP) as the sorbent. The molecularly imprinted solid phase extraction (MISPE) procedure allowed analyte extraction from complex matrices at low concentration levels and with high selectivity towards the analyte. The MISPE procedure was performed using a commercial MIP cartridge that was introduced into a guard column holder and integrated in the analyzer system. Optimized parameters included the volume of the sample, the type and volume of the conditioning and washing solutions, and the type and volume of the eluent. Quantification of clenbuterol was carried out by spectrophotometry after in-system post-elution analyte derivatization based on azo-coupling using N- (1-Naphthyl) ethylenediamine as the coupling agent to yield a red-colored compound with maximum absorbance at 500 nm. Both the chromogenic reaction and spectrophotometric detection were performed in a lab-made flow-batch mixing chamber that replaced the cuvette holder of the spectrophotometer. The calibration curve was linear in the 0.075–0.500 mg L−1 range with a correlation coefficient of 0.998. The precision of the proposed method was evaluated in terms of the relative standard deviation obtaining 1.1% and 3.0% for intra-day precision and inter-day precision, respectively. The detection limit was 0.021 mg L−1 and the sample throughput for the entire process was 3.4 h−1. The proposed method was applied for the determination of CLB in human urine and milk substitute samples obtaining recoveries values within a range of 94.0–100.0%. Fil: Gonzalez, Natalia. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - Bahía Blanca. Instituto de Química del Sur. Universidad Nacional del Sur. Departamento de Química. Instituto de Química del Sur; Argentina Fil: Grunhut, Marcos. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - Bahía Blanca. Instituto de Química del Sur. Universidad Nacional del Sur. Departamento de Química. Instituto de Química del Sur; Argentina Fil: Srámková, Ivana. Charles University; República Checa Fil: Lista, Adriana Guillermina. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - Bahía Blanca. Instituto de Química del Sur. Universidad Nacional del Sur. Departamento de Química. Instituto de Química del Sur; Argentina Fil: Horstkotte, Burkhard. Charles University; República Checa Fil: Solich, Petr. Charles University; República Checa Fil: Sklenárová, Hana. Charles University; República Checa Fil: Acebal, Carolina Cecilia. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - Bahía Blanca. Instituto de Química del Sur. Universidad Nacional del Sur. Departamento de Química. Instituto de Química del Sur; Argentina |
description |
A fully automated spectrophotometric method based on flow-batch analysis has been developed for the determination of clenbuterol including an on-line solid phase extraction using a molecularly imprinted polymer (MIP) as the sorbent. The molecularly imprinted solid phase extraction (MISPE) procedure allowed analyte extraction from complex matrices at low concentration levels and with high selectivity towards the analyte. The MISPE procedure was performed using a commercial MIP cartridge that was introduced into a guard column holder and integrated in the analyzer system. Optimized parameters included the volume of the sample, the type and volume of the conditioning and washing solutions, and the type and volume of the eluent. Quantification of clenbuterol was carried out by spectrophotometry after in-system post-elution analyte derivatization based on azo-coupling using N- (1-Naphthyl) ethylenediamine as the coupling agent to yield a red-colored compound with maximum absorbance at 500 nm. Both the chromogenic reaction and spectrophotometric detection were performed in a lab-made flow-batch mixing chamber that replaced the cuvette holder of the spectrophotometer. The calibration curve was linear in the 0.075–0.500 mg L−1 range with a correlation coefficient of 0.998. The precision of the proposed method was evaluated in terms of the relative standard deviation obtaining 1.1% and 3.0% for intra-day precision and inter-day precision, respectively. The detection limit was 0.021 mg L−1 and the sample throughput for the entire process was 3.4 h−1. The proposed method was applied for the determination of CLB in human urine and milk substitute samples obtaining recoveries values within a range of 94.0–100.0%. |
publishDate |
2018 |
dc.date.none.fl_str_mv |
2018-02-01 |
dc.type.none.fl_str_mv |
info:eu-repo/semantics/article info:eu-repo/semantics/publishedVersion http://purl.org/coar/resource_type/c_6501 info:ar-repo/semantics/articulo |
format |
article |
status_str |
publishedVersion |
dc.identifier.none.fl_str_mv |
http://hdl.handle.net/11336/91904 Gonzalez, Natalia; Grunhut, Marcos; Srámková, Ivana; Lista, Adriana Guillermina; Horstkotte, Burkhard; et al.; Flow-batch analysis of clenbuterol based on analyte extraction on molecularly imprinted polymers coupled to an in-system chromogenic reaction. Application to human urine and milk substitute samples; Elsevier Science; Talanta; 178; 1-2-2018; 934-942 0039-9140 CONICET Digital CONICET |
url |
http://hdl.handle.net/11336/91904 |
identifier_str_mv |
Gonzalez, Natalia; Grunhut, Marcos; Srámková, Ivana; Lista, Adriana Guillermina; Horstkotte, Burkhard; et al.; Flow-batch analysis of clenbuterol based on analyte extraction on molecularly imprinted polymers coupled to an in-system chromogenic reaction. Application to human urine and milk substitute samples; Elsevier Science; Talanta; 178; 1-2-2018; 934-942 0039-9140 CONICET Digital CONICET |
dc.language.none.fl_str_mv |
eng |
language |
eng |
dc.relation.none.fl_str_mv |
info:eu-repo/semantics/altIdentifier/url/https://www.sciencedirect.com/science/article/pii/S0039914017310792 info:eu-repo/semantics/altIdentifier/doi/10.1016/j.talanta.2017.10.040 |
dc.rights.none.fl_str_mv |
info:eu-repo/semantics/openAccess https://creativecommons.org/licenses/by-nc-sa/2.5/ar/ |
eu_rights_str_mv |
openAccess |
rights_invalid_str_mv |
https://creativecommons.org/licenses/by-nc-sa/2.5/ar/ |
dc.format.none.fl_str_mv |
application/pdf application/pdf application/pdf application/pdf application/pdf |
dc.publisher.none.fl_str_mv |
Elsevier Science |
publisher.none.fl_str_mv |
Elsevier Science |
dc.source.none.fl_str_mv |
reponame:CONICET Digital (CONICET) instname:Consejo Nacional de Investigaciones Científicas y Técnicas |
reponame_str |
CONICET Digital (CONICET) |
collection |
CONICET Digital (CONICET) |
instname_str |
Consejo Nacional de Investigaciones Científicas y Técnicas |
repository.name.fl_str_mv |
CONICET Digital (CONICET) - Consejo Nacional de Investigaciones Científicas y Técnicas |
repository.mail.fl_str_mv |
dasensio@conicet.gov.ar; lcarlino@conicet.gov.ar |
_version_ |
1844613076123385856 |
score |
13.070432 |