Modeling of FiO2 delivered by self-inflating ventilation device

Autores
Carvalho Filho, José; Molina, Lucas; Bedenik, Graziella; Oliveira Freire, Eduardo; Nunes Carvalho, Elyson Ádan; Feitosa, Larissa A. S.; Bittencourt, Shirley O. S.; Lima, Tamires C.
Año de publicación
2022
Idioma
inglés
Tipo de recurso
artículo
Estado
versión publicada
Descripción
During the COVID-19 pandemic, professionals in the medical field have been using bag resuscitators, also called bag valve masks, to provide respiratory support for patients for long periods. A relevant issue arising because of such use is the little information regarding the FiO2delivered by these devices, leading to toxic effects of O2and its waste. This paper presents a model that relates volume, respiratory frequency, and flow in an O2line with the FiO2delivered by bag resuscitators. The model was obtained based on a formal procedure and data from a 1500 ml Hsiner bag resuscitator collected in the literature, both presented in this paper. Then, we compared through Bland-Altman analysis the FiO2values estimated by our model with the ones from data considering different bag resuscitators to evaluate the amplitude of the errors it can generate. The average error for Hsiner and Laerdal bag resuscitators was less than 0.035 (or 3.5%) and less than 0.063 (or 6.3%) for Ambu devices. The maximum error found for Ambu bag resuscitators was high. The obtained model presented satisfactory errors, compatible with the uncertainty levels associated with FiO2measuring instruments and blenders, meaning the model can be used to estimate the delivered FiO2accurately. Additionally, the procedure presented in this paper can be used to obtain specific models for bag resuscitators of other companies, equipping professionals with tools that can accurately estimate FiO2.
Durante a pandemia de COVID-19, profissionais da área médica têm usado bolsas auto infláveis, também chamadas reanimadores manuais, para fornecer suporte respiratório aos pacientes por longos períodos. Uma questão relevante decorrente desse uso é a pouca informação sobre a FiO2 fornecida por esses dispositivos, levando a efeitos tóxicos do O2 e seus resíduos. Este artigo apresenta um modelo que relaciona volume, frequência respiratória e fluxo em uma linha de O2 com a FiO2 fornecida por ressuscitadores de bolsa. O modelo foi obtido com base em procedimento formal e dados de um reanimador com bolsa Hsiner de 1500 ml coletados na literatura, ambos apresentados neste trabalho. Em seguida, comparamos através da análise de Bland-Altman os valores de FiO2 estimados pelo nosso modelo com os dados de diferentes reanimadores de balão para avaliar a amplitude dos erros que ele pode gerar. O erro médio para ressuscitadores de bolsa Hsiner e Laerdal foi menor que 0,035 (ou 3,5%) e menor que 0,063 (ou 6,3%) para dispositivos Ambu. O erro máximo encontrado para ressuscitadores com bolsa Ambu foi alto. O modelo obtido apresentou erros satisfatórios, compatíveis com os níveis de incerteza associados aos instrumentos de medição de FiO2 e misturadores, o que significa que o modelo pode ser utilizado para estimar com precisão a FiO2 fornecida. Além disso, o procedimento apresentado neste artigo pode ser utilizado para obter modelos específicos para ressuscitadores com bolsa de outras empresas, equipando os profissionais com ferramentas que possam estimar com precisão a FiO2.
Fil: Carvalho Filho, José. Universidade Federal de Sergipe; Brasil
Fil: Molina, Lucas. Universidade Federal de Sergipe; Brasil
Fil: Bedenik, Graziella. Universidade Federal de Sergipe; Brasil
Fil: Oliveira Freire, Eduardo. Universidade Federal de Sergipe; Brasil. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - San Juan. Instituto de Automática. Universidad Nacional de San Juan. Facultad de Ingeniería. Instituto de Automática; Argentina
Fil: Nunes Carvalho, Elyson Ádan. Universidade Federal de Sergipe; Brasil
Fil: Feitosa, Larissa A. S.. Universidade Federal de Sergipe; Brasil
Fil: Bittencourt, Shirley O. S.. Universidade Federal de Sergipe; Brasil
Fil: Lima, Tamires C.. Universidade Federal de Sergipe; Brasil
Materia
fraction of inspired oxygen
FiO2 modeling
bag resuscitator
Nivel de accesibilidad
acceso abierto
Condiciones de uso
https://creativecommons.org/licenses/by-nc-sa/2.5/ar/
Repositorio
CONICET Digital (CONICET)
Institución
Consejo Nacional de Investigaciones Científicas y Técnicas
OAI Identificador
oai:ri.conicet.gov.ar:11336/267912

id CONICETDig_f3552ab1f4fcd3c7f20654414713f80c
oai_identifier_str oai:ri.conicet.gov.ar:11336/267912
network_acronym_str CONICETDig
repository_id_str 3498
network_name_str CONICET Digital (CONICET)
spelling Modeling of FiO2 delivered by self-inflating ventilation deviceCarvalho Filho, JoséMolina, LucasBedenik, GraziellaOliveira Freire, EduardoNunes Carvalho, Elyson ÁdanFeitosa, Larissa A. S.Bittencourt, Shirley O. S.Lima, Tamires C.fraction of inspired oxygenFiO2 modelingbag resuscitatorhttps://purl.org/becyt/ford/2.2https://purl.org/becyt/ford/2During the COVID-19 pandemic, professionals in the medical field have been using bag resuscitators, also called bag valve masks, to provide respiratory support for patients for long periods. A relevant issue arising because of such use is the little information regarding the FiO2delivered by these devices, leading to toxic effects of O2and its waste. This paper presents a model that relates volume, respiratory frequency, and flow in an O2line with the FiO2delivered by bag resuscitators. The model was obtained based on a formal procedure and data from a 1500 ml Hsiner bag resuscitator collected in the literature, both presented in this paper. Then, we compared through Bland-Altman analysis the FiO2values estimated by our model with the ones from data considering different bag resuscitators to evaluate the amplitude of the errors it can generate. The average error for Hsiner and Laerdal bag resuscitators was less than 0.035 (or 3.5%) and less than 0.063 (or 6.3%) for Ambu devices. The maximum error found for Ambu bag resuscitators was high. The obtained model presented satisfactory errors, compatible with the uncertainty levels associated with FiO2measuring instruments and blenders, meaning the model can be used to estimate the delivered FiO2accurately. Additionally, the procedure presented in this paper can be used to obtain specific models for bag resuscitators of other companies, equipping professionals with tools that can accurately estimate FiO2.Durante a pandemia de COVID-19, profissionais da área médica têm usado bolsas auto infláveis, também chamadas reanimadores manuais, para fornecer suporte respiratório aos pacientes por longos períodos. Uma questão relevante decorrente desse uso é a pouca informação sobre a FiO2 fornecida por esses dispositivos, levando a efeitos tóxicos do O2 e seus resíduos. Este artigo apresenta um modelo que relaciona volume, frequência respiratória e fluxo em uma linha de O2 com a FiO2 fornecida por ressuscitadores de bolsa. O modelo foi obtido com base em procedimento formal e dados de um reanimador com bolsa Hsiner de 1500 ml coletados na literatura, ambos apresentados neste trabalho. Em seguida, comparamos através da análise de Bland-Altman os valores de FiO2 estimados pelo nosso modelo com os dados de diferentes reanimadores de balão para avaliar a amplitude dos erros que ele pode gerar. O erro médio para ressuscitadores de bolsa Hsiner e Laerdal foi menor que 0,035 (ou 3,5%) e menor que 0,063 (ou 6,3%) para dispositivos Ambu. O erro máximo encontrado para ressuscitadores com bolsa Ambu foi alto. O modelo obtido apresentou erros satisfatórios, compatíveis com os níveis de incerteza associados aos instrumentos de medição de FiO2 e misturadores, o que significa que o modelo pode ser utilizado para estimar com precisão a FiO2 fornecida. Além disso, o procedimento apresentado neste artigo pode ser utilizado para obter modelos específicos para ressuscitadores com bolsa de outras empresas, equipando os profissionais com ferramentas que possam estimar com precisão a FiO2.Fil: Carvalho Filho, José. Universidade Federal de Sergipe; BrasilFil: Molina, Lucas. Universidade Federal de Sergipe; BrasilFil: Bedenik, Graziella. Universidade Federal de Sergipe; BrasilFil: Oliveira Freire, Eduardo. Universidade Federal de Sergipe; Brasil. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - San Juan. Instituto de Automática. Universidad Nacional de San Juan. Facultad de Ingeniería. Instituto de Automática; ArgentinaFil: Nunes Carvalho, Elyson Ádan. Universidade Federal de Sergipe; BrasilFil: Feitosa, Larissa A. S.. Universidade Federal de Sergipe; BrasilFil: Bittencourt, Shirley O. S.. Universidade Federal de Sergipe; BrasilFil: Lima, Tamires C.. Universidade Federal de Sergipe; BrasilBrazilian Journals Publicações de Periódicos e Editora Ltda2022-11info:eu-repo/semantics/articleinfo:eu-repo/semantics/publishedVersionhttp://purl.org/coar/resource_type/c_6501info:ar-repo/semantics/articuloapplication/pdfapplication/pdfhttp://hdl.handle.net/11336/267912Carvalho Filho, José; Molina, Lucas; Bedenik, Graziella; Oliveira Freire, Eduardo; Nunes Carvalho, Elyson Ádan ; et al.; Modeling of FiO2 delivered by self-inflating ventilation device; Brazilian Journals Publicações de Periódicos e Editora Ltda; Brazilian Journal of Development; 8; 11; 11-2022; 70825-708452525-8761CONICET DigitalCONICETenginfo:eu-repo/semantics/altIdentifier/url/https://ojs.brazilianjournals.com.br/ojs/index.php/BRJD/article/view/54499info:eu-repo/semantics/altIdentifier/doi/10.34117/bjdv8n11-003rinfo:eu-repo/semantics/openAccesshttps://creativecommons.org/licenses/by-nc-sa/2.5/ar/reponame:CONICET Digital (CONICET)instname:Consejo Nacional de Investigaciones Científicas y Técnicas2025-09-10T13:14:28Zoai:ri.conicet.gov.ar:11336/267912instacron:CONICETInstitucionalhttp://ri.conicet.gov.ar/Organismo científico-tecnológicoNo correspondehttp://ri.conicet.gov.ar/oai/requestdasensio@conicet.gov.ar; lcarlino@conicet.gov.arArgentinaNo correspondeNo correspondeNo correspondeopendoar:34982025-09-10 13:14:28.478CONICET Digital (CONICET) - Consejo Nacional de Investigaciones Científicas y Técnicasfalse
dc.title.none.fl_str_mv Modeling of FiO2 delivered by self-inflating ventilation device
title Modeling of FiO2 delivered by self-inflating ventilation device
spellingShingle Modeling of FiO2 delivered by self-inflating ventilation device
Carvalho Filho, José
fraction of inspired oxygen
FiO2 modeling
bag resuscitator
title_short Modeling of FiO2 delivered by self-inflating ventilation device
title_full Modeling of FiO2 delivered by self-inflating ventilation device
title_fullStr Modeling of FiO2 delivered by self-inflating ventilation device
title_full_unstemmed Modeling of FiO2 delivered by self-inflating ventilation device
title_sort Modeling of FiO2 delivered by self-inflating ventilation device
dc.creator.none.fl_str_mv Carvalho Filho, José
Molina, Lucas
Bedenik, Graziella
Oliveira Freire, Eduardo
Nunes Carvalho, Elyson Ádan
Feitosa, Larissa A. S.
Bittencourt, Shirley O. S.
Lima, Tamires C.
author Carvalho Filho, José
author_facet Carvalho Filho, José
Molina, Lucas
Bedenik, Graziella
Oliveira Freire, Eduardo
Nunes Carvalho, Elyson Ádan
Feitosa, Larissa A. S.
Bittencourt, Shirley O. S.
Lima, Tamires C.
author_role author
author2 Molina, Lucas
Bedenik, Graziella
Oliveira Freire, Eduardo
Nunes Carvalho, Elyson Ádan
Feitosa, Larissa A. S.
Bittencourt, Shirley O. S.
Lima, Tamires C.
author2_role author
author
author
author
author
author
author
dc.subject.none.fl_str_mv fraction of inspired oxygen
FiO2 modeling
bag resuscitator
topic fraction of inspired oxygen
FiO2 modeling
bag resuscitator
purl_subject.fl_str_mv https://purl.org/becyt/ford/2.2
https://purl.org/becyt/ford/2
dc.description.none.fl_txt_mv During the COVID-19 pandemic, professionals in the medical field have been using bag resuscitators, also called bag valve masks, to provide respiratory support for patients for long periods. A relevant issue arising because of such use is the little information regarding the FiO2delivered by these devices, leading to toxic effects of O2and its waste. This paper presents a model that relates volume, respiratory frequency, and flow in an O2line with the FiO2delivered by bag resuscitators. The model was obtained based on a formal procedure and data from a 1500 ml Hsiner bag resuscitator collected in the literature, both presented in this paper. Then, we compared through Bland-Altman analysis the FiO2values estimated by our model with the ones from data considering different bag resuscitators to evaluate the amplitude of the errors it can generate. The average error for Hsiner and Laerdal bag resuscitators was less than 0.035 (or 3.5%) and less than 0.063 (or 6.3%) for Ambu devices. The maximum error found for Ambu bag resuscitators was high. The obtained model presented satisfactory errors, compatible with the uncertainty levels associated with FiO2measuring instruments and blenders, meaning the model can be used to estimate the delivered FiO2accurately. Additionally, the procedure presented in this paper can be used to obtain specific models for bag resuscitators of other companies, equipping professionals with tools that can accurately estimate FiO2.
Durante a pandemia de COVID-19, profissionais da área médica têm usado bolsas auto infláveis, também chamadas reanimadores manuais, para fornecer suporte respiratório aos pacientes por longos períodos. Uma questão relevante decorrente desse uso é a pouca informação sobre a FiO2 fornecida por esses dispositivos, levando a efeitos tóxicos do O2 e seus resíduos. Este artigo apresenta um modelo que relaciona volume, frequência respiratória e fluxo em uma linha de O2 com a FiO2 fornecida por ressuscitadores de bolsa. O modelo foi obtido com base em procedimento formal e dados de um reanimador com bolsa Hsiner de 1500 ml coletados na literatura, ambos apresentados neste trabalho. Em seguida, comparamos através da análise de Bland-Altman os valores de FiO2 estimados pelo nosso modelo com os dados de diferentes reanimadores de balão para avaliar a amplitude dos erros que ele pode gerar. O erro médio para ressuscitadores de bolsa Hsiner e Laerdal foi menor que 0,035 (ou 3,5%) e menor que 0,063 (ou 6,3%) para dispositivos Ambu. O erro máximo encontrado para ressuscitadores com bolsa Ambu foi alto. O modelo obtido apresentou erros satisfatórios, compatíveis com os níveis de incerteza associados aos instrumentos de medição de FiO2 e misturadores, o que significa que o modelo pode ser utilizado para estimar com precisão a FiO2 fornecida. Além disso, o procedimento apresentado neste artigo pode ser utilizado para obter modelos específicos para ressuscitadores com bolsa de outras empresas, equipando os profissionais com ferramentas que possam estimar com precisão a FiO2.
Fil: Carvalho Filho, José. Universidade Federal de Sergipe; Brasil
Fil: Molina, Lucas. Universidade Federal de Sergipe; Brasil
Fil: Bedenik, Graziella. Universidade Federal de Sergipe; Brasil
Fil: Oliveira Freire, Eduardo. Universidade Federal de Sergipe; Brasil. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - San Juan. Instituto de Automática. Universidad Nacional de San Juan. Facultad de Ingeniería. Instituto de Automática; Argentina
Fil: Nunes Carvalho, Elyson Ádan. Universidade Federal de Sergipe; Brasil
Fil: Feitosa, Larissa A. S.. Universidade Federal de Sergipe; Brasil
Fil: Bittencourt, Shirley O. S.. Universidade Federal de Sergipe; Brasil
Fil: Lima, Tamires C.. Universidade Federal de Sergipe; Brasil
description During the COVID-19 pandemic, professionals in the medical field have been using bag resuscitators, also called bag valve masks, to provide respiratory support for patients for long periods. A relevant issue arising because of such use is the little information regarding the FiO2delivered by these devices, leading to toxic effects of O2and its waste. This paper presents a model that relates volume, respiratory frequency, and flow in an O2line with the FiO2delivered by bag resuscitators. The model was obtained based on a formal procedure and data from a 1500 ml Hsiner bag resuscitator collected in the literature, both presented in this paper. Then, we compared through Bland-Altman analysis the FiO2values estimated by our model with the ones from data considering different bag resuscitators to evaluate the amplitude of the errors it can generate. The average error for Hsiner and Laerdal bag resuscitators was less than 0.035 (or 3.5%) and less than 0.063 (or 6.3%) for Ambu devices. The maximum error found for Ambu bag resuscitators was high. The obtained model presented satisfactory errors, compatible with the uncertainty levels associated with FiO2measuring instruments and blenders, meaning the model can be used to estimate the delivered FiO2accurately. Additionally, the procedure presented in this paper can be used to obtain specific models for bag resuscitators of other companies, equipping professionals with tools that can accurately estimate FiO2.
publishDate 2022
dc.date.none.fl_str_mv 2022-11
dc.type.none.fl_str_mv info:eu-repo/semantics/article
info:eu-repo/semantics/publishedVersion
http://purl.org/coar/resource_type/c_6501
info:ar-repo/semantics/articulo
format article
status_str publishedVersion
dc.identifier.none.fl_str_mv http://hdl.handle.net/11336/267912
Carvalho Filho, José; Molina, Lucas; Bedenik, Graziella; Oliveira Freire, Eduardo; Nunes Carvalho, Elyson Ádan ; et al.; Modeling of FiO2 delivered by self-inflating ventilation device; Brazilian Journals Publicações de Periódicos e Editora Ltda; Brazilian Journal of Development; 8; 11; 11-2022; 70825-70845
2525-8761
CONICET Digital
CONICET
url http://hdl.handle.net/11336/267912
identifier_str_mv Carvalho Filho, José; Molina, Lucas; Bedenik, Graziella; Oliveira Freire, Eduardo; Nunes Carvalho, Elyson Ádan ; et al.; Modeling of FiO2 delivered by self-inflating ventilation device; Brazilian Journals Publicações de Periódicos e Editora Ltda; Brazilian Journal of Development; 8; 11; 11-2022; 70825-70845
2525-8761
CONICET Digital
CONICET
dc.language.none.fl_str_mv eng
language eng
dc.relation.none.fl_str_mv info:eu-repo/semantics/altIdentifier/url/https://ojs.brazilianjournals.com.br/ojs/index.php/BRJD/article/view/54499
info:eu-repo/semantics/altIdentifier/doi/10.34117/bjdv8n11-003r
dc.rights.none.fl_str_mv info:eu-repo/semantics/openAccess
https://creativecommons.org/licenses/by-nc-sa/2.5/ar/
eu_rights_str_mv openAccess
rights_invalid_str_mv https://creativecommons.org/licenses/by-nc-sa/2.5/ar/
dc.format.none.fl_str_mv application/pdf
application/pdf
dc.publisher.none.fl_str_mv Brazilian Journals Publicações de Periódicos e Editora Ltda
publisher.none.fl_str_mv Brazilian Journals Publicações de Periódicos e Editora Ltda
dc.source.none.fl_str_mv reponame:CONICET Digital (CONICET)
instname:Consejo Nacional de Investigaciones Científicas y Técnicas
reponame_str CONICET Digital (CONICET)
collection CONICET Digital (CONICET)
instname_str Consejo Nacional de Investigaciones Científicas y Técnicas
repository.name.fl_str_mv CONICET Digital (CONICET) - Consejo Nacional de Investigaciones Científicas y Técnicas
repository.mail.fl_str_mv dasensio@conicet.gov.ar; lcarlino@conicet.gov.ar
_version_ 1842980773325438976
score 12.993085