Indirect ELISA (iELISA) for routine detection of antibodies against Minute Virus of Mice (MVM) in mice colonies

Autores
Laborde, Juan Martin; Sguazza, Guillermo Hernán; Fuentealba, Nadia Analia; Corva, Santiago Gerardo; Carbone, Cecilia; Galosi, Cecilia Monica
Año de publicación
2017
Idioma
inglés
Tipo de recurso
artículo
Estado
versión publicada
Descripción
In this study we developed an indirect ELISA to detect antibodies against Minute Virus of Mice (MVM) using an antigen produced from BHK-21 cells infected with a prototype strain of the virus. The optimal antigen concentration and serum dilutions were established. In order to analyze variability in the laboratory, reproducibility and repeatability within and between plates were determined. Then, a panel of 460 sera from conventional facilities and previously classified as positive or negative by the indirect fluorescent antibody assay was analyzed. The cutoff value was determined by a receiver operating characteristic (ROC) curve. The results of the indirect ELISA were compared with those of the indirect fluorescent antibody assay. The ELISA assay showed 100% sensitivity and 99% specificity. ELISA is a useful tool to be developed in standard virology laboratories and can be used for screening animals faster than the traditional indirect fluorescent antibody assay.
Se desarrolló un ELISA indirecto para detectar anticuerpos contra el virus diminuto del ratón (Mice minute virus [MVM]), utilizando un antígeno producido a partir de células BHK-21 infectadas con la cepa prototipo del virus. Se establecieron las diluciones óptimas de antígeno y el suero a utilizar. Para analizar la variabilidad en el laboratorio, se determinaron la reproducibilidad y la repetibilidad dentro de una placa y entre placas. Luego se analizaron 460 sueros provenientes de bioterios convencionales y clasificados previamente como positivos o negativos por inmunofluorescencia indirecta. El valor de corte se determinó mediante una curva ROC. Los resultados se compararon con los obtenidos con la prueba de inmunofluorescencia indirecta. El ELISA mostró 100% de sensibilidad y un 99% de especificidad. Esta técnica demostró ser una herramienta útil para desarrollar en laboratorios de virología estándar y puede utilizarse como prueba tamiz para seleccionar animales de manera más rápida que con la tradicional prueba de inmunofluorescencia indirecta
Fil: Laborde, Juan Martin. Universidad Nacional de la Plata. Facultad de Ciencias Veterinarias. Departamento de Microbiología. Cátedra de Animales de Laboratorio; Argentina
Fil: Sguazza, Guillermo Hernán. Universidad Nacional de La Plata. Facultad de Ciencias Veterinarias. Departamento de Microbiología. Cátedra de Virología; Argentina
Fil: Fuentealba, Nadia Analia. Universidad Nacional de La Plata. Facultad de Ciencias Veterinarias. Departamento de Microbiología. Cátedra de Virología; Argentina. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - La Plata; Argentina
Fil: Corva, Santiago Gerardo. Universidad Nacional de La Plata. Facultad de Ciencias Veterinarias. Departamento de Ciencias Básicas; Argentina
Fil: Carbone, Cecilia. Universidad Nacional de la Plata. Facultad de Ciencias Veterinarias. Departamento de Microbiología. Cátedra de Animales de Laboratorio; Argentina
Fil: Galosi, Cecilia Monica. Universidad Nacional de La Plata. Facultad de Ciencias Veterinarias. Departamento de Microbiología. Cátedra de Virología; Argentina. Provincia de Buenos Aires. Gobernación. Comisión de Investigaciones Científicas; Argentina
Materia
DIAGNOSIS
ELISA
MINUTE VIRUS OF MICE
Nivel de accesibilidad
acceso abierto
Condiciones de uso
https://creativecommons.org/licenses/by-nc-nd/2.5/ar/
Repositorio
CONICET Digital (CONICET)
Institución
Consejo Nacional de Investigaciones Científicas y Técnicas
OAI Identificador
oai:ri.conicet.gov.ar:11336/130268

id CONICETDig_759154e2420f794abab0cd3c5d639d11
oai_identifier_str oai:ri.conicet.gov.ar:11336/130268
network_acronym_str CONICETDig
repository_id_str 3498
network_name_str CONICET Digital (CONICET)
spelling Indirect ELISA (iELISA) for routine detection of antibodies against Minute Virus of Mice (MVM) in mice coloniesELISA indirecto (iELISA) para la detección de rutina de anticuerpos contra el virus diminuto del ratón (MVM) en colonias de ratonesLaborde, Juan MartinSguazza, Guillermo HernánFuentealba, Nadia AnaliaCorva, Santiago GerardoCarbone, CeciliaGalosi, Cecilia MonicaDIAGNOSISELISAMINUTE VIRUS OF MICEhttps://purl.org/becyt/ford/4.3https://purl.org/becyt/ford/4In this study we developed an indirect ELISA to detect antibodies against Minute Virus of Mice (MVM) using an antigen produced from BHK-21 cells infected with a prototype strain of the virus. The optimal antigen concentration and serum dilutions were established. In order to analyze variability in the laboratory, reproducibility and repeatability within and between plates were determined. Then, a panel of 460 sera from conventional facilities and previously classified as positive or negative by the indirect fluorescent antibody assay was analyzed. The cutoff value was determined by a receiver operating characteristic (ROC) curve. The results of the indirect ELISA were compared with those of the indirect fluorescent antibody assay. The ELISA assay showed 100% sensitivity and 99% specificity. ELISA is a useful tool to be developed in standard virology laboratories and can be used for screening animals faster than the traditional indirect fluorescent antibody assay.Se desarrolló un ELISA indirecto para detectar anticuerpos contra el virus diminuto del ratón (Mice minute virus [MVM]), utilizando un antígeno producido a partir de células BHK-21 infectadas con la cepa prototipo del virus. Se establecieron las diluciones óptimas de antígeno y el suero a utilizar. Para analizar la variabilidad en el laboratorio, se determinaron la reproducibilidad y la repetibilidad dentro de una placa y entre placas. Luego se analizaron 460 sueros provenientes de bioterios convencionales y clasificados previamente como positivos o negativos por inmunofluorescencia indirecta. El valor de corte se determinó mediante una curva ROC. Los resultados se compararon con los obtenidos con la prueba de inmunofluorescencia indirecta. El ELISA mostró 100% de sensibilidad y un 99% de especificidad. Esta técnica demostró ser una herramienta útil para desarrollar en laboratorios de virología estándar y puede utilizarse como prueba tamiz para seleccionar animales de manera más rápida que con la tradicional prueba de inmunofluorescencia indirectaFil: Laborde, Juan Martin. Universidad Nacional de la Plata. Facultad de Ciencias Veterinarias. Departamento de Microbiología. Cátedra de Animales de Laboratorio; ArgentinaFil: Sguazza, Guillermo Hernán. Universidad Nacional de La Plata. Facultad de Ciencias Veterinarias. Departamento de Microbiología. Cátedra de Virología; ArgentinaFil: Fuentealba, Nadia Analia. Universidad Nacional de La Plata. Facultad de Ciencias Veterinarias. Departamento de Microbiología. Cátedra de Virología; Argentina. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - La Plata; ArgentinaFil: Corva, Santiago Gerardo. Universidad Nacional de La Plata. Facultad de Ciencias Veterinarias. Departamento de Ciencias Básicas; ArgentinaFil: Carbone, Cecilia. Universidad Nacional de la Plata. Facultad de Ciencias Veterinarias. Departamento de Microbiología. Cátedra de Animales de Laboratorio; ArgentinaFil: Galosi, Cecilia Monica. Universidad Nacional de La Plata. Facultad de Ciencias Veterinarias. Departamento de Microbiología. Cátedra de Virología; Argentina. Provincia de Buenos Aires. Gobernación. Comisión de Investigaciones Científicas; ArgentinaAsociación Argentina de Microbiología2017-07info:eu-repo/semantics/articleinfo:eu-repo/semantics/publishedVersionhttp://purl.org/coar/resource_type/c_6501info:ar-repo/semantics/articuloapplication/pdfapplication/pdfapplication/pdfhttp://hdl.handle.net/11336/130268Laborde, Juan Martin; Sguazza, Guillermo Hernán; Fuentealba, Nadia Analia; Corva, Santiago Gerardo; Carbone, Cecilia; et al.; Indirect ELISA (iELISA) for routine detection of antibodies against Minute Virus of Mice (MVM) in mice colonies; Asociación Argentina de Microbiología; Revista Argentina de Microbiología; 49; 3; 7-2017; 210-2151851-76170325-7541CONICET DigitalCONICETenginfo:eu-repo/semantics/altIdentifier/url/http://linkinghub.elsevier.com/retrieve/pii/S0325754117300421info:eu-repo/semantics/altIdentifier/doi/10.1016/j.ram.2017.02.005info:eu-repo/semantics/openAccesshttps://creativecommons.org/licenses/by-nc-nd/2.5/ar/reponame:CONICET Digital (CONICET)instname:Consejo Nacional de Investigaciones Científicas y Técnicas2025-09-03T10:02:35Zoai:ri.conicet.gov.ar:11336/130268instacron:CONICETInstitucionalhttp://ri.conicet.gov.ar/Organismo científico-tecnológicoNo correspondehttp://ri.conicet.gov.ar/oai/requestdasensio@conicet.gov.ar; lcarlino@conicet.gov.arArgentinaNo correspondeNo correspondeNo correspondeopendoar:34982025-09-03 10:02:35.737CONICET Digital (CONICET) - Consejo Nacional de Investigaciones Científicas y Técnicasfalse
dc.title.none.fl_str_mv Indirect ELISA (iELISA) for routine detection of antibodies against Minute Virus of Mice (MVM) in mice colonies
ELISA indirecto (iELISA) para la detección de rutina de anticuerpos contra el virus diminuto del ratón (MVM) en colonias de ratones
title Indirect ELISA (iELISA) for routine detection of antibodies against Minute Virus of Mice (MVM) in mice colonies
spellingShingle Indirect ELISA (iELISA) for routine detection of antibodies against Minute Virus of Mice (MVM) in mice colonies
Laborde, Juan Martin
DIAGNOSIS
ELISA
MINUTE VIRUS OF MICE
title_short Indirect ELISA (iELISA) for routine detection of antibodies against Minute Virus of Mice (MVM) in mice colonies
title_full Indirect ELISA (iELISA) for routine detection of antibodies against Minute Virus of Mice (MVM) in mice colonies
title_fullStr Indirect ELISA (iELISA) for routine detection of antibodies against Minute Virus of Mice (MVM) in mice colonies
title_full_unstemmed Indirect ELISA (iELISA) for routine detection of antibodies against Minute Virus of Mice (MVM) in mice colonies
title_sort Indirect ELISA (iELISA) for routine detection of antibodies against Minute Virus of Mice (MVM) in mice colonies
dc.creator.none.fl_str_mv Laborde, Juan Martin
Sguazza, Guillermo Hernán
Fuentealba, Nadia Analia
Corva, Santiago Gerardo
Carbone, Cecilia
Galosi, Cecilia Monica
author Laborde, Juan Martin
author_facet Laborde, Juan Martin
Sguazza, Guillermo Hernán
Fuentealba, Nadia Analia
Corva, Santiago Gerardo
Carbone, Cecilia
Galosi, Cecilia Monica
author_role author
author2 Sguazza, Guillermo Hernán
Fuentealba, Nadia Analia
Corva, Santiago Gerardo
Carbone, Cecilia
Galosi, Cecilia Monica
author2_role author
author
author
author
author
dc.subject.none.fl_str_mv DIAGNOSIS
ELISA
MINUTE VIRUS OF MICE
topic DIAGNOSIS
ELISA
MINUTE VIRUS OF MICE
purl_subject.fl_str_mv https://purl.org/becyt/ford/4.3
https://purl.org/becyt/ford/4
dc.description.none.fl_txt_mv In this study we developed an indirect ELISA to detect antibodies against Minute Virus of Mice (MVM) using an antigen produced from BHK-21 cells infected with a prototype strain of the virus. The optimal antigen concentration and serum dilutions were established. In order to analyze variability in the laboratory, reproducibility and repeatability within and between plates were determined. Then, a panel of 460 sera from conventional facilities and previously classified as positive or negative by the indirect fluorescent antibody assay was analyzed. The cutoff value was determined by a receiver operating characteristic (ROC) curve. The results of the indirect ELISA were compared with those of the indirect fluorescent antibody assay. The ELISA assay showed 100% sensitivity and 99% specificity. ELISA is a useful tool to be developed in standard virology laboratories and can be used for screening animals faster than the traditional indirect fluorescent antibody assay.
Se desarrolló un ELISA indirecto para detectar anticuerpos contra el virus diminuto del ratón (Mice minute virus [MVM]), utilizando un antígeno producido a partir de células BHK-21 infectadas con la cepa prototipo del virus. Se establecieron las diluciones óptimas de antígeno y el suero a utilizar. Para analizar la variabilidad en el laboratorio, se determinaron la reproducibilidad y la repetibilidad dentro de una placa y entre placas. Luego se analizaron 460 sueros provenientes de bioterios convencionales y clasificados previamente como positivos o negativos por inmunofluorescencia indirecta. El valor de corte se determinó mediante una curva ROC. Los resultados se compararon con los obtenidos con la prueba de inmunofluorescencia indirecta. El ELISA mostró 100% de sensibilidad y un 99% de especificidad. Esta técnica demostró ser una herramienta útil para desarrollar en laboratorios de virología estándar y puede utilizarse como prueba tamiz para seleccionar animales de manera más rápida que con la tradicional prueba de inmunofluorescencia indirecta
Fil: Laborde, Juan Martin. Universidad Nacional de la Plata. Facultad de Ciencias Veterinarias. Departamento de Microbiología. Cátedra de Animales de Laboratorio; Argentina
Fil: Sguazza, Guillermo Hernán. Universidad Nacional de La Plata. Facultad de Ciencias Veterinarias. Departamento de Microbiología. Cátedra de Virología; Argentina
Fil: Fuentealba, Nadia Analia. Universidad Nacional de La Plata. Facultad de Ciencias Veterinarias. Departamento de Microbiología. Cátedra de Virología; Argentina. Consejo Nacional de Investigaciones Científicas y Técnicas. Centro Científico Tecnológico Conicet - La Plata; Argentina
Fil: Corva, Santiago Gerardo. Universidad Nacional de La Plata. Facultad de Ciencias Veterinarias. Departamento de Ciencias Básicas; Argentina
Fil: Carbone, Cecilia. Universidad Nacional de la Plata. Facultad de Ciencias Veterinarias. Departamento de Microbiología. Cátedra de Animales de Laboratorio; Argentina
Fil: Galosi, Cecilia Monica. Universidad Nacional de La Plata. Facultad de Ciencias Veterinarias. Departamento de Microbiología. Cátedra de Virología; Argentina. Provincia de Buenos Aires. Gobernación. Comisión de Investigaciones Científicas; Argentina
description In this study we developed an indirect ELISA to detect antibodies against Minute Virus of Mice (MVM) using an antigen produced from BHK-21 cells infected with a prototype strain of the virus. The optimal antigen concentration and serum dilutions were established. In order to analyze variability in the laboratory, reproducibility and repeatability within and between plates were determined. Then, a panel of 460 sera from conventional facilities and previously classified as positive or negative by the indirect fluorescent antibody assay was analyzed. The cutoff value was determined by a receiver operating characteristic (ROC) curve. The results of the indirect ELISA were compared with those of the indirect fluorescent antibody assay. The ELISA assay showed 100% sensitivity and 99% specificity. ELISA is a useful tool to be developed in standard virology laboratories and can be used for screening animals faster than the traditional indirect fluorescent antibody assay.
publishDate 2017
dc.date.none.fl_str_mv 2017-07
dc.type.none.fl_str_mv info:eu-repo/semantics/article
info:eu-repo/semantics/publishedVersion
http://purl.org/coar/resource_type/c_6501
info:ar-repo/semantics/articulo
format article
status_str publishedVersion
dc.identifier.none.fl_str_mv http://hdl.handle.net/11336/130268
Laborde, Juan Martin; Sguazza, Guillermo Hernán; Fuentealba, Nadia Analia; Corva, Santiago Gerardo; Carbone, Cecilia; et al.; Indirect ELISA (iELISA) for routine detection of antibodies against Minute Virus of Mice (MVM) in mice colonies; Asociación Argentina de Microbiología; Revista Argentina de Microbiología; 49; 3; 7-2017; 210-215
1851-7617
0325-7541
CONICET Digital
CONICET
url http://hdl.handle.net/11336/130268
identifier_str_mv Laborde, Juan Martin; Sguazza, Guillermo Hernán; Fuentealba, Nadia Analia; Corva, Santiago Gerardo; Carbone, Cecilia; et al.; Indirect ELISA (iELISA) for routine detection of antibodies against Minute Virus of Mice (MVM) in mice colonies; Asociación Argentina de Microbiología; Revista Argentina de Microbiología; 49; 3; 7-2017; 210-215
1851-7617
0325-7541
CONICET Digital
CONICET
dc.language.none.fl_str_mv eng
language eng
dc.relation.none.fl_str_mv info:eu-repo/semantics/altIdentifier/url/http://linkinghub.elsevier.com/retrieve/pii/S0325754117300421
info:eu-repo/semantics/altIdentifier/doi/10.1016/j.ram.2017.02.005
dc.rights.none.fl_str_mv info:eu-repo/semantics/openAccess
https://creativecommons.org/licenses/by-nc-nd/2.5/ar/
eu_rights_str_mv openAccess
rights_invalid_str_mv https://creativecommons.org/licenses/by-nc-nd/2.5/ar/
dc.format.none.fl_str_mv application/pdf
application/pdf
application/pdf
dc.publisher.none.fl_str_mv Asociación Argentina de Microbiología
publisher.none.fl_str_mv Asociación Argentina de Microbiología
dc.source.none.fl_str_mv reponame:CONICET Digital (CONICET)
instname:Consejo Nacional de Investigaciones Científicas y Técnicas
reponame_str CONICET Digital (CONICET)
collection CONICET Digital (CONICET)
instname_str Consejo Nacional de Investigaciones Científicas y Técnicas
repository.name.fl_str_mv CONICET Digital (CONICET) - Consejo Nacional de Investigaciones Científicas y Técnicas
repository.mail.fl_str_mv dasensio@conicet.gov.ar; lcarlino@conicet.gov.ar
_version_ 1842269764517363712
score 13.13397